Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery

被引:5
|
作者
Maddison, Benjamin [1 ]
Wolff, Christopher [1 ]
Findlay, George [2 ]
Radermacher, Peter [3 ]
Hinds, Charles [1 ]
Pearse, Rupert M. [1 ]
机构
[1] Queen Marys Univ London, Royal London Hosp, Barts & London Sch Med & Dent, London E1 1BB, England
[2] Univ Wales Hosp, Intens Care Unit, Cardiff CF14 4XW, S Glam, Wales
[3] Univ Ulm Klinikum, Sekt Anasthesiol Pathophysiol & Verfahrensentwick, D-89081 Ulm, Germany
来源
CRITICAL CARE | 2009年 / 13卷 / 04期
关键词
CRITICALLY ILL PATIENTS; POSTMORTEM GRAVIMETRY; OUTPUT MEASUREMENT; THERMAL DILUTION; INDICATOR; SINGLE; THERMODILUTION; BLOOD; CIRCULATION; MANAGEMENT;
D O I
10.1186/cc7948
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction Measurement of extravascular lung water (EVLW) by using the lithium-thermal (Li-thermal) and single-thermal indicator dilution methods was compared with the indocyanine green-thermal (ICG-thermal) method in humans. Methods Single-center observational study involving patients undergoing cardiac surgery with cardiopulmonary bypass. Paired measurements were taken 1, 2, 4, and 6 hours after surgery. Bland-Altman analysis was used to calculate bias and limits of agreement. Data are presented as mean (SD) or median (IQR). Results Seventeen patients were recruited (age, 69 years ( 54 to 87 years); Parsonnet score 10 (0 to 29)). Sixteen ICG-thermal measurements were excluded after blinded assessment because of poor-quality indicator dilution curves. EVLW volume as measured by the ICG-thermal technique was 4.6 (1.9) ml/kg, compared with 5.3 (1.4) ml/kg for the single-thermal method. Measurements taken with the Li-thermal method were clearly erroneous (-7.6 (7.4) ml/kg). In comparison with simultaneous measurements with the ICG-thermal method, single-thermal measurements had an acceptable degree of bias, but limits of agreement were poor (bias, -0.3 ml/kg (2.3)). Li-thermal measurements compared poorly with the ICG-thermal reference method (bias, 13.2 ml/kg (14.4)). Conclusions The principal finding of this study was that the prototype Li-thermal method did not provide reliable measurements of EVLW volume when compared with the ICG-thermal reference technique. Although minimal bias was associated with the single-thermal method, limits of agreement were approximately 45% of the normal value of EVLW volume. The Li-thermal method performed very poorly because of the overestimation of mean indicator transit time by using an external lithium ion electrode. These findings suggest that the assessment of lung water content by lithium-indicator dilution is not sufficiently reliable for clinical use in individual patients.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery
    Benjamin Maddison
    Christopher Wolff
    George Findlay
    Peter Radermacher
    Charles Hinds
    Rupert M Pearse
    Critical Care, 13
  • [2] MEASUREMENT OF EXTRAVASCULAR LUNG WATER IN INFANTS AND CHILDREN AFTER CARDIAC-SURGERY
    VINCENT, RN
    LANG, P
    ELIXSON, EM
    GAMBLE, WJ
    FULTON, DR
    FELLOWS, KE
    NORWOOD, WI
    CASTANEDA, AR
    AMERICAN JOURNAL OF CARDIOLOGY, 1984, 54 (01): : 161 - 165
  • [3] Assessment of Extravascular Lung Water by Ultrasound After Congenital Cardiac Surgery
    Kaskinen, Anu K.
    Martelius, Laura
    Kirjavainen, Turkka
    Rautiainen, Paula
    Andersson, Sture
    Pitkanen, Olli M.
    PEDIATRIC PULMONOLOGY, 2017, 52 (03) : 345 - 352
  • [4] MEASUREMENT OF EXTRAVASCULAR LUNG WATER - METHODS AND CLINICAL IMPLICATIONS
    LEKSELL, LG
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1991, 35 : 81 - 86
  • [5] CHANGES IN EXTRAVASCULAR LUNG WATER AFTER VALVULAR SURGERY
    SHEH, JM
    OCONNOR, N
    GAZZANIG.A
    DEWATER, JV
    CIRCULATION, 1969, 40 (4S3) : I186 - &
  • [6] AGE AND CARDIAC-SURGERY - INFLUENCE ON EXTRAVASCULAR LUNG WATER
    BOLDT, J
    VONBORMANN, B
    KLING, D
    MULCH, J
    HEMPELMANN, G
    CHEST, 1987, 91 (02) : 185 - 189
  • [7] Measurement of extravascular lung water in patients with pulmonary edema
    Matthay, Michael A.
    AMERICAN JOURNAL OF PHYSIOLOGY-LUNG CELLULAR AND MOLECULAR PHYSIOLOGY, 2008, 294 (06) : L1021 - L1022
  • [8] Extravascular lung water measurement in critically ill patients
    Khosravi, Parisa Moll
    Reuter, Daniel
    Kassiri, Negin
    Hashemian, Seyed Mohammadreza
    BIOMEDICAL AND BIOTECHNOLOGY RESEARCH JOURNAL, 2018, 2 (04): : 237 - 241
  • [9] Measurement of Extravascular Lung Water in Critically Ill Patients
    Sakka, Samir G.
    CURRENT RESPIRATORY MEDICINE REVIEWS, 2007, 3 (03) : 206 - 213
  • [10] Ultrasound assessment of extravascular lung water after thoracic surgery
    Cagini, Lucio
    Becattini, Cecilia
    Ranalli, Maria Giovanna
    Andolfi, Marco
    Mancuso, Alessandra
    Vannucci, Jacopo
    Agnelli, Giancarlo
    Puma, Francesco
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46