A Novel electronic chain for high level neutron coincidence counters

被引:3
|
作者
Fazzi, A
Pirovano, C
Varoli, V
机构
[1] Politecnico di Milano, Dip. Ingegneria Nucleare, Milano 20133
关键词
D O I
10.1109/23.603710
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes a counting chain (charge preamplifier, shaping amplifier and threshold discriminator) devoted to He-3 neutron detectors. The preamplifier adopts an innovative circuit with the gate of the input JFET floating and a DC feedback loop that stabilizes the output voltage acting on the input cascade second transistor. Static and dynamic analysis, including the effects of the detector bias network, is developed and reported. The shaping amplifier is a fifth-order quasi-gaussian OPAMP active filter that minimizes the number of components. Shaping time constants ranging from 1 mu s down to 100 ns are explored in order to maximize the count rate capability, maintaining nevertheless neutron discrimination. Experimental signals and spectra show that 100 ns shaping time is a lower limit for the usual He-3 proportional detector. Since this counting chain is characterized by single 5V supply operation, high counting rate, small size and low cost, it is well suited for high efficiency neutron well detectors where a large number (10 - 100) of counting tubes are used (High Level Neutron Coincidence Counter HLNCC).
引用
收藏
页码:557 / 562
页数:6
相关论文
共 50 条
  • [31] A HIGH LEVEL ELECTRONIC NOISE SOURCE
    COBINE, JD
    GALLAGHER, CJ
    PHYSICAL REVIEW, 1946, 70 (1-2): : 119 - 119
  • [32] DETERMINATION OF THE BACKGROUND OF 3HE-FILLED PROPORTIONAL COUNTERS USED FOR LOW-LEVEL NEUTRON MEASUREMENTS
    Reginatto, Marcel
    RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 403 - 406
  • [34] Front-end electronics for high rate neutron counters: Its performance and radiation hardness improvement
    Lee, TH
    Kim, HD
    Park, SW
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 1423 - 1427
  • [35] A STUDY OF NEUTRON SENSITIVITY AND ALPHA-BACKGROUND IN URANIUM 233 FISSION COUNTERS FOR LOW-LEVEL FLUX MONITORING
    LEE, SM
    ATOMKERNENERGIE-KERNTECHNIK, 1985, 46 (01): : 52 - 53
  • [36] COINCIDENCE OF 2 CARCINOMATA WITH EXCESSIVE HIGH-LEVEL OF SERUM-ESTROGEN
    STELDINGER, R
    SCHMIDT, L
    SCHMIDTGOLLWITZER, M
    EBER, H
    IGLESIAS, JR
    FRISIUS, H
    MEDIZINISCHE KLINIK, 1976, 71 (41) : 1717 - 1719
  • [37] Measurements with low pressure tissue equivalent proportional counters in well defined high energy neutron beams
    Menzel, H-G.
    Schuhmacher, H.
    Dietze, G.
    Proceedings of the International Congress of Radiation Research, 1984, : 1 - 12
  • [38] A Novel Detector for High Neutron Flux Measurements
    Singo, T. D.
    Papka, P.
    Wyngaardt, S. M.
    Dobson, R. T.
    NUCLEAR PHYSICS METHODS AND ACCELERATORS IN BIOLOGY AND MEDICINE, 2009, 1204 : 168 - +
  • [39] HIGH-EFFICIENCY SCINTILLATOR COINCIDENCE SYSTEM FOR ABSOLUTE DETERMINATION OF NEUTRON-FLUX DENSITY
    COLLEN, J
    NUCLEAR INSTRUMENTS & METHODS, 1973, 108 (02): : 269 - 270
  • [40] NOVEL ELECTRONIC-PROPERTIES OF SOLIDS REVEALED BY AUGER-PHOTOELECTRON COINCIDENCE SPECTROSCOPY (APECS)
    BARTYNSKI, RA
    JENSEN, E
    HULBERT, SL
    PHYSICA SCRIPTA, 1992, T41 : 168 - 174