A hydrogen peroxide biosensor with high stability based on gelatin-multiwalled carbon nanotubes modified glassy carbon electrode

被引:20
|
作者
Wang, Yulin [1 ]
Li, Tianyu [2 ]
Zhang, Weijie [1 ]
Huang, Yaqin [1 ]
机构
[1] Beijing Univ Chem Technol, State Key Lab Chem Resource Engn, Key Lab Carbon Fiber & Funct Polymers, Minist Educ, Beijing 100029, Peoples R China
[2] Tsinghua Univ, Dept Chem, Key Lab Organ Optoelect & Mol Engn, Beijing 100084, Peoples R China
关键词
Hydrogen peroxide biosensor; Multiwalled carbon nanotube; Gelatin; Catalase; Direct electrochemistry; GLUCOSE BIOSENSOR; HEMOGLOBIN; CATALASE; BIOELECTROCHEMISTRY; IMMOBILIZATION; NANOPARTICLES; SURFACES; FILM;
D O I
10.1007/s10008-014-2441-z
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
A biosensor with high stability was prepared to determine hydrogen peroxide (H2O2). This hydrogen peroxide biosensor was obtained by modifying glassy carbon electrode (GCE) with a composite film composed of gelatin-multiwalled carbon nanotubes. Catalase (Cat) was covalently immobilized into gelatin-multiwalled carbon nanotubes modified GCE through the well-known glutaraldehyde (GAD) chemistry in order to enhance the stability of electrodes. The enzyme sensor can achieve direct electrochemical response of hydrogen peroxide. The cyclic voltammograms at different scan rates, electrochemical impedance spectroscopy (EIS), and scanning electron microscope (SEM) tests indicate that the enzyme sensor performs positively on increasing permeability, reducing the electron transfer resistance, and improving the electrode performance. The linear response of standard curve for H2O2 is in the range of 0.2 to 5.0 mM with a correlation coefficient of 0.9972, and the detection limit of 0.001 mM. A high operational and storage stability is demonstrated for the biosensor. The peak potential at room temperature in two consecutive weeks stays almost consistent, and the enzyme activity is kept stable even after 30 days in further study.
引用
收藏
页码:1981 / 1987
页数:7
相关论文
共 50 条
  • [31] A non-enzymatic hydrogen peroxide sensor based on poly(vinyl alcohol)-multiwalled carbon nanotubes-platinum nanoparticles hybrids modified glassy carbon electrode
    Fang, Yuxin
    Zhang, Di
    Qin, Xia
    Miao, Zhiying
    Takahashi, Shigehiro
    Anzai, Jun-ichi
    Chen, Qiang
    ELECTROCHIMICA ACTA, 2012, 70 : 266 - 271
  • [32] Amperometric biosensor for hydrogen peroxide based on electrodeposited sub-micrometer gold modified glassy carbon electrode
    Wang, SQ
    Chen, J
    Lin, XQ
    CHINESE JOURNAL OF CHEMISTRY, 2004, 22 (04) : 360 - 364
  • [33] Amperometric Biosensor for Hydrogen Peroxide Based on Electrodeposited Sub-micrometer Gold Modified Glassy Carbon Electrode
    王树青
    陈峻
    林祥钦
    Chinese Journal of Chemistry, 2004, (04) : 360 - 364
  • [34] A New Hydrogen Peroxide Biosensor Based on Hrp/Ag Nanowires/Glassy Carbon Electrode
    Zhang, Min
    Cheng, Faliang
    Cai, Zhiquan
    Chen, Meiqiong
    MECHANICAL AND AEROSPACE ENGINEERING, PTS 1-7, 2012, 110-116 : 577 - 584
  • [35] An Amperometric Hydrogen Peroxide Sensor Based on Reduced Graphene Oxide/Carbon Nanotubes/Pt NPs Modified Glassy Carbon Electrode
    Zhang, Yuanyuan
    Cao, Qi
    Zhu, Fengmei
    Xu, Hedan
    Zhang, Yang
    Xu, Wenfeng
    Liao, Xiaoling
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2020, 15 (09): : 8771 - 8785
  • [36] Picomolar detection of hydrogen peroxide at glassy carbon electrode modified with NAD+ and single walled carbon nanotubes
    Salimi, Abdollah
    Miranzadeh, Leyla
    Hallaj, Rahman
    Mamkhezri, Hussein
    ELECTROANALYSIS, 2008, 20 (16) : 1760 - 1768
  • [37] A hydrogen peroxide amperometric biosensor based on carbon nanotubes
    Mai Zhibin
    Tan Xuecai
    Zou Xiaoyong
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2006, 34 (06) : 801 - 804
  • [38] Voltammetric Determination of Sulfonamides with a Modified Glassy Carbon Electrode Using Carboxyl Multiwalled Carbon Nanotubes
    He, Bao-Shan
    Chen, Wen-Bo
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2016, 27 (12) : 2216 - 2225
  • [39] Electroanalysis of Bisphenol A at a Multiwalled Carbon Nanotubes-gold Nanoparticles Modified Glassy Carbon Electrode
    Tu, Xinman
    Yan, Liushui
    Luo, Xubiao
    Luo, Shenglian
    Xie, Qingji
    ELECTROANALYSIS, 2009, 21 (22) : 2491 - 2494
  • [40] Amperometric Lactate Biosensor Based on Carbon Paste Electrode Modified with Benzo[c]cinnoline and Multiwalled Carbon Nanotubes
    Celik, Ali Cihan
    Ozturk, Funda
    Erden, Pinar Esra
    Kacar, Ceren
    Kilic, Esma
    ELECTROANALYSIS, 2015, 27 (12) : 2820 - 2828