The effect of ionising radiation on poly(methyl methacrylate) used in intraocular lenses

被引:8
|
作者
Pratt, C. M. [1 ]
Barton, S. [1 ]
McGonigle, E. [1 ]
Kishi, M. [1 ]
Foot, P. J. S. [1 ]
机构
[1] Kingston Univ, Sch Chem & Pharmaceut Sci, Mat Res Grp, Kingston upon Thames KT1 2EE, Surrey, England
关键词
PMMA; intraocular; sterilisation; irradiation; SEC; FT-IR;
D O I
10.1016/j.polymdegradstab.2006.04.006
中图分类号
O63 [高分子化学(高聚物)];
学科分类号
070305 ; 080501 ; 081704 ;
摘要
Gamma radiation is increasingly being used to sterilise intraocular lenses (IOLs) made from poly(methyl methacrylate) (PNIMA). In this study, samples of PMMA used in the fabrication of IOLs were exposed to irradiation doses typically used for their sterilisation. The effect of this treatment on the polymer was analysed by size-exclusion chromatography (SEC), UV-visible and infrared spectroscopy and scanning electron microscopy (SEM). The PMMA was found to have undergone chain scission, decarboxylation and colour change following the irradiation. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2315 / 2317
页数:3
相关论文
共 50 条
  • [31] Evaluation of cellular adhesions on silicone and poly(methyl methacrylate) intraocular lenses in monkey eyes - An electron microscopic study
    Ohnishi, Y
    Yoshitomi, T
    Sakamoto, T
    Fujisawa, K
    Ishibashi, T
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (12): : 2036 - 2040
  • [32] Low-vacuum, low-voltage scanning electron microscopy of poly(methyl methacrylate) intraocular lenses
    Hayashi, H
    Oshima, K
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (03): : 385 - 389
  • [33] Changes in ESR spectra of poly(methyl methacrylate) used in retrospective dosimetry of radiation accidents
    Szocs, F
    RADIATION PHYSICS AND CHEMISTRY, 1998, 51 (02) : 181 - 184
  • [34] Effect of 3-piece AcrySof and downsized heparin-surface-modified poly(methyl methacrylate) intraocular lenses in infant rabbit eyes
    Lundvall, A
    Zetterström, C
    Lundgren, B
    Kugelberg, U
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01): : 159 - 163
  • [35] Principles of radiation protection of poly(methyl methacrylate)
    Salimgareeva, V.N.
    Kolesov, S.V.
    Russian Journal of Applied Chemistry, 2006, 79 (09): : 1393 - 1402
  • [36] MECHANISM OF THE RADIATION AGING OF POLY(METHYL METHACRYLATE
    RAFIKOV, SR
    IBRAGIMOV, SS
    LEPLYANIN, GV
    SALIMGAREEVA, VN
    PIVOVAROV, SP
    POLYAKOV, AI
    HIGH ENERGY CHEMISTRY, 1981, 15 (04) : 253 - 257
  • [37] Principles of radiation protection of poly(methyl methacrylate)
    V. N. Salimgareeva
    S. V. Kolesov
    Russian Journal of Applied Chemistry, 2006, 79 : 1393 - 1402
  • [38] EFFECTIVENESS OF RADIATION PROTECTION IN POLY(METHYL METHACRYLATE
    WANG, LJ
    ZHANG, ZP
    RADIATION PHYSICS AND CHEMISTRY, 1983, 22 (06): : 1062 - 1062
  • [39] Principles of radiation protection of poly(methyl methacrylate)
    Salimgareeva, V. N.
    Kolesov, S. V.
    RUSSIAN JOURNAL OF APPLIED CHEMISTRY, 2006, 79 (09) : 1393 - 1402
  • [40] VISUAL DISTURBANCES ASSOCIATED WITH OVAL-OPTIC POLY(METHYL METHACRYLATE) AND ROUND-OPTIC SILICONE INTRAOCULAR LENSES
    ANDERSON, CJ
    STURM, RJ
    SHAPIRO, MB
    BALLEW, C
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1994, 20 (03): : 295 - 298