摘要
This paper describes how to calibrate a four-terminal-pair (4TP) LCR meter at the right (90 degrees) angle. The LCR meter uses the right angle signal to process a synchronous detector as a vector signal receiver. The right-angle calibration method uses, at a minimum, one variable-phase signal generator (SG) and a phase shift circuit. The SG does not require a precision right angle, but it does need good phase stability. The described phase shifter provides the required differential phase of nearly the right angle. As a result, the right-angle phase deviation of the LCR meter is better than 65 mu rad (0.223') with 37-mu rad (0.127') uncertainty at 10 MHz. This method is also applicable to the phase calibration of a vector network analyzer.
引用
收藏
页码:993 / 996
页数:4
相关论文
共 50 条
  • [31] A self-calibration method for robotic measurement system
    Gong, CH
    Yuan, JX
    Ni, J
    JOURNAL OF MANUFACTURING SCIENCE AND ENGINEERING-TRANSACTIONS OF THE ASME, 2000, 122 (01): : 174 - 181
  • [32] Digital self-calibration method for MEMS sensors
    Colinet, E
    Juillard, J
    Nicu, L
    Bergaud, C
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2005, 54 (04) : 1438 - 1443
  • [33] Self-calibration method for robotic measurement system
    Gong, Chunhe
    Yuan, Jingxia
    Ni, Jun
    American Society of Mechanical Engineers, Manufacturing Engineering Division, MED, 1999, 10 : 879 - 888
  • [34] A linear method for camera pair self-calibration
    Melanitis, Nikos
    Maragos, Petros
    COMPUTER VISION AND IMAGE UNDERSTANDING, 2021, 210
  • [35] The self-calibration method of IR mixer with low IF
    Szymanski, Andrzej
    Kurjata-Pfitzner, Ewa
    Lesinski, Jan
    Wasowski, Jerzy
    ANALOG INTEGRATED CIRCUITS AND SIGNAL PROCESSING, 2008, 55 (02) : 115 - 124
  • [36] A Self-Calibration Method of Microbolometer With Vacuum Package
    Liu, Chao
    Fu, Jianyu
    Hou, Ying
    Zhou, Qiong
    Chen, Dapeng
    IEEE SENSORS JOURNAL, 2020, 20 (15) : 8570 - 8575
  • [37] A Self-Calibration Method for an Implantable Displacement Sensor
    Hao, Shiying
    2013 IEEE INTERNATIONAL CONFERENCE ON MICROWAVES, COMMUNICATIONS, ANTENNAS AND ELECTRONICS SYSTEMS (IEEE COMCAS 2013), 2013,
  • [38] A WEIGHTED ITERATIVE METHOD FOR ROBUST SELF-CALIBRATION
    Liu Shigang Wu Chengke Tang Li Jia Jing (National Key Laboratory of ISN
    Journal of Electronics(China), 2006, (05) : 713 - 717
  • [39] A method of self-calibration for integrated capacitive sensors
    Johnson, PM
    Thiel, DV
    James, D
    Cornelius, M
    SMART STRUCTURES, DEVICES, AND SYSTEMS II, PT 1 AND 2, 2005, 5649 : 486 - 491
  • [40] Stage Cartesian self-calibration: a second method
    Takac, MT
    Whittey, J
    18TH ANNUAL SYMPOSIUM ON PHOTOMASK TECHNOLOGY AND MANAGEMENT, 1998, 3546 : 498 - 504