A mathematical model for patient skin dose assessment in cardiac catheterization procedures

被引:7
|
作者
Morrell, R. E. [1 ]
Rogers, A. T. [1 ]
机构
[1] Nottingham Univ Hosp NHS Trust, Nottingham NG5 1PB, England
来源
BRITISH JOURNAL OF RADIOLOGY | 2006年 / 79卷 / 945期
关键词
D O I
10.1259/bjr/57359387
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A mathematical model has been developed for the assessment of patient skin doses from cardiac catheterization procedures. This uses exposure and projection data stored in the DICOM image files. Since these contain only information about the acquisition runs, a correction is needed to estimate and include the contribution from fluoroscopy. Maximum skin doses calculated by the model were found to correlate well with those measured on Kodak EDR2 film. Three methods for including the contribution from fluoroscopy were investigated, and all successfully identified patients receiving skin doses in excess of 1 Gy. It is hoped to automate this tool for routine assessment of skin doses in our cardiac catheterization laboratories.
引用
收藏
页码:756 / 761
页数:6
相关论文
共 50 条
  • [21] EVALUATING THE MAXIMUM PATIENT RADIATION DOSE IN CARDIAC INTERVENTIONAL PROCEDURES
    Kato, Mamoru
    Chida, Koichi
    Sato, Tadaya
    Oosaka, Hajime
    Tosa, Tetsuo
    Kadowaki, Ken
    RADIATION PROTECTION DOSIMETRY, 2011, 143 (01) : 69 - 73
  • [22] Entrance skin dose using dose-area product measurements: Application to a cardiac catheterization laboratory
    Adnani, N
    MEDICAL PHYSICS, 2003, 30 (06) : 1381 - 1381
  • [23] Assessment of Patient Dose in Selected Non-Cardiac Interventional Fluoroscopy Procedures Using OSL Dosimeters
    Elona, I. A.
    Morales, A. A., Jr.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 783 - 786
  • [24] An assessment of cardiac nurses' knowledge and practices regarding patient safety after cardiac catheterization
    Alshammari, Thuwaini
    Alshammari, Mohammad
    Alshammari, Najah
    Alshammari, Tahani
    Alshammari, Amjad
    INTERNATIONAL JOURNAL OF ADVANCED AND APPLIED SCIENCES, 2024, 11 (04): : 1 - 8
  • [25] Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization
    Bacher, K
    Bogaert, E
    Lapere, R
    De Wolf, D
    Thierens, H
    CIRCULATION, 2005, 111 (01) : 83 - 89
  • [26] Characteristics Associated With Reduced Patient Radiation Dose in an Academic Cardiac Catheterization Lab
    Mann, Steve
    Green, Cynthia
    Russ, Megan
    Wei, Trent
    Patel, Manesh
    Tcheng, James
    Rao, Sunil
    Jones, Schuyler
    Wegermann, Zachary
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 76 (17) : B80 - B80
  • [27] Implementing Patient Safety Radiation Dose Alerts for Pediatric Cardiac Catheterization Examinations
    Somasundaram, E.
    Brady, S. L.
    Strauss, K. J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7846 - 7846
  • [28] Characterization of Patient Dose Distribution and Probabilistic Risk Analysis in Cardiac Catheterization Laboratories
    Weir, V.
    Zhang, J.
    Khan, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E665 - E665
  • [29] Assessment of Angular Dose Profile for Cardiac and Noncardiac Interventional Procedures
    Ai, H.
    Supanich, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E315 - E315
  • [30] A skin lesion after cardiac catheterization
    Arias-Santiago, Salvador
    Aneiros-Fernandez, Jose
    Carriel, Victor
    Orgaz-Molina, Jacinto
    Gonzalez-Andrades, Miguel
    Buendia-Eisman, Agustin
    Alaminos, Miguel
    CLEVELAND CLINIC JOURNAL OF MEDICINE, 2012, 79 (06) : 424 - 426