PET Performance Evaluation of an MR-Compatible PET Insert

被引:32
|
作者
Wu, Yibao [1 ]
Catana, Ciprian [2 ]
Farrell, Richard [3 ]
Dokhale, Purushottam A. [3 ]
Shah, Kanai S. [3 ]
Qi, Jinyi [1 ]
Cherry, Simon R. [1 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Massachusetts Gen Hosp, Dept Radiol, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA 02129 USA
[3] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
基金
美国国家卫生研究院;
关键词
Avalanche photodiodes; positron emission tomography (PET); magnetic resonance imaging (MRI); POSITRON-EMISSION-TOMOGRAPHY; SENSITIVE AVALANCHE PHOTODIODES; SMALL ANIMAL PET; SIMULTANEOUS PET/MRI; SCANNER; MICROPET; DETECTOR; DESIGN; ACQUISITION; RESOLUTION;
D O I
10.1109/TNS.2009.2015448
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A magnetic resonance (MR) compatible positron emission tomography (PET) insert has been developed in our laboratory for simultaneous small animal PET/MR imaging. This system is based on lutetium oxyorthosilicate (LSO) scintillator arrays with position-sensitive avalanche photodiode (PSAPD) photodetectors. The PET performance of this insert has been measured. The average reconstructed image spatial resolution was 1.51 mm. The sensitivity at the center of the field of view (CFOV) was 0.35%, which is comparable to the simulation predictions of 0.40%. The average photopeak energy resolution was 25%. The scatter fraction inside the MRI scanner with a line source was 12% (with a mouse-sized phantom and standard 35 mm Bruker 1H RF coil), 7% (with RF coil only) and 5% (without phantom or RF coil) for an energy window of 350-650 keV. The front-end electronics had a dead time of 390 ns, and a trigger extension dead time of 7.32 mu s that degraded counting rate performance for injected doses above similar to 0.75 mCi (28 MBq). The peak noise-equivalent count rate (NECR) of 1.27 kcps was achieved at 290 mu Ci (10.7 MBq). The system showed good imaging performance inside a 7-T animal MRI system; however improvements in data acquisition electronics and reduction of the coincidence timing window are needed to realize improved NECR performance.
引用
收藏
页码:574 / 580
页数:7
相关论文
共 50 条
  • [1] Performance evaluation of an MR-compatible silicon photomultiplier (SiPM) small animal PET insert
    Kwon, Sun Il
    Ito, Mikiko
    Lee, Chan Mi
    Ko, Guen Bae
    Yoon, Hyun Suk
    Kim, Joong Hyun
    Song, In Chan
    Lee, Dong Soo
    Hong, Seong Jong
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [2] Evaluation of an MR-compatible blood sampler for PET
    Breuer, J.
    Grazioso, R.
    Zhang, N.
    Schmand, M.
    Wienhard, K.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (19): : 5883 - 5893
  • [3] PET Performance Evaluation of a pre-clinical SiPM based MR-compatible PET scanner
    Mackewn, Jane E.
    Lerche, Christoph W.
    Sunassee, Kavitha
    de Rosales, Rafael T. M.
    Phinikaridou, Alkystis
    Salomon, Andre
    Ayres, Richard
    Tsoumpas, Charalampos
    Soultanidis, Georgios M.
    Schaeffter, Tobias
    Marsden, Paul K.
    Schulz, Volkmar
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2776 - 2779
  • [4] MR-compatible blood sampler for PET
    Breuer, J.
    Grazioso, R.
    Corbeil, J.
    Zhang, N.
    Schmand, M.
    Wienhard, K.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3426 - +
  • [5] Simultaneous PET/MR imaging with a MR-compatible SiPM PET scanner
    Yoon, Hyun Suk
    Ko, Guen Bae
    Kwon, Sun Il
    Lee, Chan Mi
    Ito, Mikiko
    Song, In Chan
    Lee, Dong Soo
    Hong, Seong Jong
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [6] PET Performance Evaluation of a Pre-Clinical SiPM-Based MR-Compatible PET Scanner
    Mackewn, Jane E.
    Lerche, Christoph W.
    Weissler, Bjoern
    Sunassee, Kavitha
    de Rosales, Rafael T. M.
    Phinikaridou, Alkystis
    Salomon, Andre
    Ayres, Richard
    Tsoumpas, Charalampos
    Soultanidis, Georgios M.
    Gebhardt, Pierre
    Schaeffter, Tobias
    Marsden, Paul K.
    Schulz, Volkmar
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (03) : 784 - 790
  • [7] Gamma shielding materials for MR-compatible PET
    Strul, D
    Cash, D
    Keevil, SF
    Halsted, P
    Williams, SCR
    Marsden, PK
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (01) : 60 - 69
  • [8] Comparison of PET/CT with Sequential PET/MRI Using an MR-Compatible Mobile PET System
    Nakamoto, Ryusuke
    Nakamoto, Yuji
    Ishimori, Takayoshi
    Fushimi, Yasutaka
    Kido, Aki
    Togashi, Kaori
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (05) : 846 - 851
  • [9] Simultaneous acquisition of multislice PET and MR images: Initial results with a MR-compatible PET scanner
    Catana, Ciprian
    Wu, Yibao
    Judenhofer, Martin S.
    Qi, Jinyi
    Pichler, Berrid J.
    Cherry, Simon R.
    JOURNAL OF NUCLEAR MEDICINE, 2006, 47 (12) : 1968 - 1976
  • [10] PET Performance Evaluation of a RF-Penetrable PET Insert for Simultaneous PET/MR Imaging
    Chang, Chen-Ming
    Grant, Alexander M.
    Lee, Brian J.
    Levin, Craig S.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,