Automating the correction of flow integration drift during whole-body plethysmography

被引:0
|
作者
Stucky, Frederic [1 ]
Cazzaniga, Giacomo [2 ]
Aliverti, Andrea [2 ]
Kayser, Bengt [1 ]
Uva, Barbara [1 ]
机构
[1] Univ Lausanne, Inst Sport Sci, CH-1015 Lausanne, Switzerland
[2] Politecn Milan, Dipartimento Bioingn, I-20133 Milan, Italy
来源
42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20 | 2020年
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Prolonged measurement of total body volume variations (deltaVb) with whole-body, flow-based plethysmography (WBP) results in a drift of the signal due to changes in temperature and humidity inside the plethysmograph and to numerical integration of the flow to obtain deltaVb. This drift has been previously corrected with the application of a wavelet-based filter using visual inspection of the signal to select the optimal filter level (Uva et al. Front. Physiol. 6:411, 2016), thus introducing potential operator bias. To exclude the latter we compared this approach with a newly developed automatic method based on (1) correction for actual changes in temperature and humidity inside the plethysmograph (algorithm TH) and (2) automatic selection of the wavelet filter level based on comparison between deltaVb and intra-thoracic and abdominal pressure variations measured simultaneously (algorithm WAV). The Pearson's correlation coefficient between deltaVb and the changes in volume of the chest wall (deltaVcw) simultaneously obtained by optoelectronic plethysmography (OEP) was calculated after correction of deltaVb with TH and WAV applied separately, TH andWAV applied consecutively (TH+WAV), manual selection of a wavelet filter based on visual inspection (MAN) or no correction (CTRL). The correlation between deltaVb and deltaVcw increased marginally with WAV, TH+WAV and MAN compared to CTRL ( P < 0.01). Conversely, TH alone yielded a lower correlation (P < 0.01). It follows that while the automated wavelet filter level selection method (WAV) represents an effective, operator-independent method for the correction of deltaVb, whether or not it is combined with specific correction for changes in thermodynamic conditions inside the plethysmograph, the manual method (MAN) yields satisfactory results without the constraints of intra-thoracic and abdominal pressure measurement.
引用
收藏
页码:5 / 8
页数:4
相关论文
共 50 条
  • [11] ERGOMETRIC STUDIES BY MEANS OF WHOLE-BODY PLETHYSMOGRAPHY
    MEYERERKELENZ, JD
    SIEVERTS, H
    PRAXIS UND KLINIK DER PNEUMOLOGIE, 1981, 35 : 724 - 727
  • [12] CLOSING VOLUMES MEASURED BY WHOLE-BODY PLETHYSMOGRAPHY
    MATTHYS, H
    KONIETZKO, N
    BULLETIN DE PHYSIO-PATHOLOGIE RESPIRATOIRE, 1973, 9 (05): : 1263 - 1263
  • [13] Accuracy of Whole-Body Plethysmography Requires Biological Calibration
    Poorisrisak, Porntiva
    Vrang, Carsten
    Henriksen, Jorn Molgaard.
    Klug, Bent
    Hanel, Birgitte
    Bisgaard, Hans
    CHEST, 2009, 135 (06) : 1476 - 1480
  • [14] NEW USES OF WHOLE-BODY PLETHYSMOGRAPHY APPLIED TO CHILDREN
    DAB, I
    PEDIATRIC RESEARCH, 1977, 11 (09) : 1029 - 1029
  • [15] Evaluation Of Body Composition Measurements Obtained Using Whole-body Plethysmography
    Macaulay, Timothy R.
    Uhlir, Conor D.
    Schroeder, E. Todd
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2018, 50 (05): : 161 - 161
  • [16] Impulse oscillometry in COPD: Correlation to spirometry and whole-body plethysmography
    Vukoja, Marija
    Milicic, Dragana
    Kopitovic, Ivan
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [17] Impulse oscillometry in COPD: Correlation to spirometry and whole-body plethysmography
    Vukoja, Marija
    Milicic, Dragana
    Kopitovic, Ivan
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [18] COMPUTER-AIDED INTERPRETATION OF THE RESULTS OF WHOLE-BODY PLETHYSMOGRAPHY
    ZAISS, AW
    WILLIM, G
    MATTHYS, H
    ATEMWEGS-UND LUNGENKRANKHEITEN, 1987, 13 (08) : 380 - 381
  • [19] INHALATIONAL ANTIGEN INDUCTION TRIALS USING WHOLE-BODY PLETHYSMOGRAPHY
    GONSIOR, E
    KRUGER, M
    MEIERSYDOW, J
    ACTA ALLERGOLOGICA, 1976, 31 (04): : 283 - 296
  • [20] Measuring Breathing Patterns in Mice Using Whole-body Plethysmography
    Prada-Dacasa, Patricia
    Urpi, Andrea
    Sanchez-Benito, Laura
    Bianchi, Patrizia
    Quintana, Albert
    BIO-PROTOCOL, 2020, 10 (17):