An uncommon cause of cemented unicompartmental knee arthroplasty failure: fracture of metallic components

被引:10
|
作者
Manzotti, Alfonso [1 ]
Chemello, Cesare [2 ]
Pullen, Chris [3 ]
Cerveri, Pietro [4 ]
Confalonieri, Norberto [1 ]
机构
[1] CTO Hosp, Orthoped Dept 1, I-20100 Milan, Italy
[2] Azienda Osped Padova, Clin Ortoped, I-35123 Padua, Italy
[3] Royal Melbourne Hosp, Dept Orthopaed, Parkville, Vic 3050, Australia
[4] Politecn Milan, Bioengn Dept, I-20100 Milan, Italy
关键词
Unicompartmental Knee Arthroplasty; Metallic component breakage; FEMORAL COMPONENT; OXFORD KNEE; REPLACEMENT; REVISION; PROSTHESIS; THICKNESS; ALIGNMENT; WEAR;
D O I
10.1007/s00167-012-2062-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Despite good overall clinical results, unicompartmental knee replacements (UKR) are not without their problems and failures have been reported. The most common causes of UKR failure are component loosening, poor patient selection, poor surgical technique, polyethylene wear and progression of arthritis in other compartments. The purpose of this study is to present a series of atraumatic fractures of metallic components in a UKR treated in a single orthopaedic centre. Since 1999, 121 failed unicompartmental knee arthroplasties have been referred to our centre. In six of these, atraumatic breakage of a metal component in the cemented UKR was seen and included in this study. Pre-operative alignment, BMI and implant longevity were documented. The femoral implant failed in 4 patients and the tibial implant in a further 2. All the femoral implant fractures occurred within 3 years of UKR surgery (mean: 22.2 months, SD: 10.6 months). Tibial implant breakage occurred at a mean of 8.5 years (SD: 2.4 months) following UKR. All patients were treated with conversion to a navigated total knee replacement. A primary total knee arthroplasty was used in all cases with one patient requiring a tibial component incorporating a wedge and stem following breakage of the original UKR tibial implant. Fracture of the metallic components is a potential cause of failure of unicompartmental knee arthroplasty. In our experience, the incidence of this complication was 4.9 % of all UKR failures. Patients with a BMI greater than 30 and a progressive deterioration in limb alignment were at greater risk. IV.
引用
收藏
页码:2518 / 2522
页数:5
相关论文
共 50 条
  • [1] An uncommon cause of cemented unicompartmental knee arthroplasty failure: fracture of metallic components
    Alfonso Manzotti
    Cesare Chemello
    Chris Pullen
    Pietro Cerveri
    Norberto Confalonieri
    Knee Surgery, Sports Traumatology, Arthroscopy, 2013, 21 : 2518 - 2522
  • [2] Cemented versus cementless unicompartmental knee arthroplasty
    Iniguez, C. Magaly
    Anastasiadis, Zoy
    Nazer, Maria Ignacia
    Sandoval, Rodrigo
    JOURNAL OF ISAKOS JOINT DISORDERS & ORTHOPAEDIC SPORTS MEDICINE, 2024, 9 (06)
  • [3] Functional analysis of cemented medial unicompartmental knee arthroplasty
    Chassin, EP
    Mikosz, RP
    Andriacchi, TP
    Rosenberg, AG
    JOURNAL OF ARTHROPLASTY, 1996, 11 (05): : 553 - 559
  • [4] Early failure of unicompartmental knee arthroplasty
    Mariani, E. Marc
    Bourne, Michael H.
    Jackson, Richard T.
    Jackson, Scott T.
    Jones, Portia
    JOURNAL OF ARTHROPLASTY, 2007, 22 (06): : 81 - 84
  • [5] Early failure of the porous coated anatomic cemented unicompartmental knee arthroplasty - Aids to diagnosis and revision
    Knight, JL
    Atwater, RD
    Guo, J
    JOURNAL OF ARTHROPLASTY, 1997, 12 (01): : 11 - 20
  • [6] Cemented versus Uncemented Oxford Unicompartmental Knee Arthroplasty: Is There a Difference?
    Akan, Burak
    Karaguven, Dogac
    Guclu, Berk
    Yildirim, Tugrul
    Kaya, Alper
    Armangil, Mehmet
    Cetin, Ilker
    ADVANCES IN ORTHOPEDICS, 2013, 2013
  • [7] Distinct age-related modes of failure in cemented and cementless Oxford medial unicompartmental knee arthroplasty
    de Brauw, G. V. ten Noever
    Vossen, R. J. M.
    Bayoumi, T.
    Sierevelt, I. N.
    Burger, J. A.
    Pearle, A. D.
    Kerkhoffs, G. M. M. J.
    Spekenbrink- Spooren, A.
    Zuiderbaan, A.
    BONE & JOINT JOURNAL, 2025, 107B (03): : 329 - 336
  • [8] Bearing failure in a mobile bearing unicompartmental knee arthroplasty: an uncommon presentation of an implant-specific complication
    Sravya P. Vajapey
    Paul M. Alvarez
    Douglas Chonko
    Arthroplasty, 3
  • [9] Bearing failure in a mobile bearing unicompartmental knee arthroplasty: an uncommon presentation of an implant-specific complication
    Vajapey, Sravya P.
    Alvarez, Paul M.
    Chonko, Douglas
    ARTHROPLASTY, 2021, 3 (01)
  • [10] Delayed tibial-platform periprosthetic stress fracture after unicompartmental knee arthroplasty: Uncommon and devastating complication
    Yuk-Wah, Hung
    Jason, Chi-Ho Fan
    Carson, Ka-Bon Kwok
    Ellis, Lok-Fai Wong
    Lo Carman, Ka-Man