Off-pump transapical mitral valve replacement

被引:26
|
作者
Lutter, Georg [1 ]
Quaden, Rene [1 ]
Osaki, Satoru [2 ]
Hu, Jian [2 ]
Renner, Jochen [2 ]
Edwards, Niloo M. [3 ]
Cremer, Jochen [1 ]
Lozonschi, Lucian [3 ]
机构
[1] Univ Hosp Schleswig Holstein, Dept Cardiovasc Surg, Kiel, Germany
[2] Univ Hosp Schleswig Holstein, Dept Anaesthesia, Kiel, Germany
[3] Univ Wisconsin, Sch Med & Publ Hlth, Dept Cardiovasc Surg, Madison, WI USA
关键词
Valve replacement; Mitral; Off-pump; Catheter; Transapical; Transcatheter; Percutaneous; Minimally invasive; Valved stent; STENT;
D O I
10.1016/j.ejcts.2009.02.037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Percutaneous valve replacement was recently introduced, and reports of early clinical experience have already been published. To date, this technique is limited to the replacement of pulmonary and aortic valves in a strictly selected group of patients. The aim of this study was to analyse a self-expanding valved stent for minimally invasive replacement of the mitral valve in animals. Methods: A newly designed nitinol stent was specially designed for this experimental acute study. It comprised of a left ventricular tubular stent with star shaped left atrial anchoring springs and carried a trileaflet bovine pericardial valve. A polytetrafluoroethylene membrane was sutured to envelop the atrial springs and the outside of the ventricular stent. The ventricular anchoring system was the same as in our previously reported results with a similar mitral valved stent. Seven pigs underwent minimally invasive off-pump mitral valved stent implantation. This was per-formed through a tower ministernotomy and a standard transapical approach under transoesophageal echocardiographic (TEE) guidance was used. Results: The valved stent is fully retrievable and precise deployment and accurate adjustment of its intra-annular position is achievable to eliminate paravalvular leakage. The deployment time ranged from 127 to 255 s and the blood loss from 70 to 220 cc. One animal died of intractable ventricular fibrillation. Mitral regurgitation in all surviving animals was minimal (trace in 5/6 and mild in 1/6 during echo examination; on the contrast ventriculogram no mitral insufficiency was observed except in one documented as mild paravalvular regurgitation). These animals remained haemodynamically stable (6/6) and without TEE or ventriculographic changes for 1 h. Conclusion: Implantation of a tricuspid bovine pericardial valved stent in the mitral position is feasible in pigs through a transcatheter approach. This was possible through a smaller delivery system than previously reported. Additional studies are required to demonstrate long-term feasibility, durability, and heart function. (C) 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. Ail rights reserved.
引用
收藏
页码:124 / 128
页数:5
相关论文
共 50 条
  • [1] Off-pump transapical mitral valve replacement: evaluation after one month
    Iino, Kenji
    Boldt, Jessica
    Lozonschi, Lucian
    Metzner, Anja
    Schoettler, Jan
    Petzina, Rainer
    Cremer, Jochen
    Lutter, Georg
    Maisano, F.
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2012, 41 (03) : 512 - 517
  • [2] One-Stage Off-Pump Transapical Mitral Valve Repair and Aortic Valve Replacement
    Gerosa, Gino
    D'Onofrio, Augusto
    Manzan, Erica
    Besola, Laura
    Bizzotto, Eleonora
    Zucchetta, Fabio
    Tarantini, Giuseppe
    Pittarello, Demetrio
    Colli, Andrea
    CIRCULATION, 2015, 131 (19) : E430 - E434
  • [3] Off-pump transapical mitral valve-in-ring implantation
    Zou, Yu
    Ferrari, Enrico
    von Segesser, Ludwig K.
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (04) : 849 - 855
  • [4] Transapical off-pump neochord implantation procedure for mitral valve insufficiency
    Colli, A.
    Zucchetta, F.
    Manzan, E.
    Sarais, C.
    Pittarello, D.
    Speziali, G.
    Gerosa, G. I. N. O.
    EUROPEAN HEART JOURNAL, 2014, 35 : 36 - 36
  • [5] Review of transapical off-pump mitral valve intervention with NeoChord implantation
    Hassanabad, Ali Fatehi
    Kent, William D. T.
    Adams, Corey
    CURRENT OPINION IN CARDIOLOGY, 2021, 36 (02) : 130 - 140
  • [6] Surgical Phantom for Off-Pump Mitral Valve Replacement
    McLeod, A. Jonathan
    Moore, John
    Guiraudon, Gerard M.
    Jones, Doug L.
    Campbell, Gordon
    Peters, Terry M.
    MEDICAL IMAGING 2011: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, 2011, 7964
  • [7] Transapical off-pump mitral valve repair following prior mitral valve surgery A case report
    Blessberger, Hermann
    Kellermair, Joerg
    Kammler, Juergen
    Steinwender, Clemens
    Zierer, Andreas F.
    MEDICINE, 2021, 100 (21) : E26148
  • [8] Artificial papillary muscle device for off-pump transapical mitral valve repair
    Imbrie-Moore, Annabel M.
    Zhu, Yuanjia
    Park, Matthew H.
    Paulsen, Michael J.
    Wang, Hanjay
    Woo, Y. Joseph
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2022, 164 (04): : E133 - E141
  • [9] Numerical simulation of transapical off-pump mitral valve repair with neochordae implantation
    Gaidulis, Gediminas
    Votta, Emiliano
    Selmi, Matteo
    Aidietiene, Sigita
    Aidietis, Audrius
    Kacianauskas, Rimantas
    TECHNOLOGY AND HEALTH CARE, 2018, 26 : S635 - S645
  • [10] CT for the Transapical Off-Pump Mitral Valve Repair With Neochord Implantation Procedure
    Colli, Andrea
    Zucchetta, Fabio
    Kliger, Chad
    Bellu, Roberto
    Francone, Marco
    Sedati, Pietro
    Jelnin, Vladimir
    Ruiz, Carlos E.
    Manzan, Erica
    Besola, Laura
    Bizzotto, Eleonora
    Gerosa, Gino
    JACC-CARDIOVASCULAR IMAGING, 2017, 10 (11) : 1397 - 1400