The effect of coronal preflaring on the accuracy of two electronic apex locators

被引:2
|
作者
Suryantoro, R. [1 ]
Meidyawati, R. [1 ]
Suprastiwi, E. [1 ]
机构
[1] Univ Indonesia, Dept Conservat Dent, Fac Dent, Jakarta, Indonesia
关键词
IN-VIVO EVALUATION; ROOT ZX; LENGTH; FREQUENCY;
D O I
10.1088/1742-6596/884/1/012059
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
One of the main factors that determines the success of root canal treatment is the instrumentation confined to the root canal; specific methods are available for measuring its working length. This study aims to compare the accuracy of two types of apex locators and evaluate the influence of coronal preflaring on their use. The working lengths of sixteen first mandibular bicuspids were measured visually, using two types of apex locators before and after coronal preflaring. The result showed that there is a statistically significant difference between the two types of apex locators before coronal preflaring, but no statistically significant difference after coronal preflaring. The study concludes that coronal preflaring increases the accuracy of apex locators in locating the apical foramen.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] Use of electronic apex locators may improve determination of working lengthQuestion: Are electronic apex locators (EALs) accurate?
    Isabel Mello
    Evidence-Based Dentistry, 2014, 15 (4) : 120 - 120
  • [42] In vivo comparison of working length determination with two electronic apex locators
    Wrbas, K. T.
    Ziegler, A. A.
    Altenburger, M. J.
    Schirrmeister, J. F.
    INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (02) : 133 - 138
  • [43] The influence of periapical lesions on the repeatability of two electronic apex locators in vivo
    Carlos G. Adorno
    Sandra M. Solaeche
    Ileana E. Ferreira
    Andrea Pedrozo
    Patricia M. Escobar
    Vicente R. Fretes
    Clinical Oral Investigations, 2021, 25 : 5239 - 5245
  • [44] A comparison between two electronic apex locators:: an ex vivo investigation
    Venturi, M.
    Breschi, L.
    INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (05) : 362 - 373
  • [45] Comparison of working length determination with radiographs and two electronic apex locators
    Vieyra, J. P.
    Acosta, J.
    Mondaca, J. M.
    INTERNATIONAL ENDODONTIC JOURNAL, 2010, 43 (01) : 16 - 20
  • [46] The influence of periapical lesions on the repeatability of two electronic apex locators in vivo
    Adorno, Carlos G.
    Solaeche, Sandra M.
    Ferreira, Ileana E.
    Pedrozo, Andrea
    Escobar, Patricia M.
    Fretes, Vicente R.
    CLINICAL ORAL INVESTIGATIONS, 2021, 25 (09) : 5239 - 5245
  • [47] Occurrence of no-function of two electronic apex locators: an in vivo study
    Chevalier, Valerie
    Arbab-Chirani, Reza
    Nicolas, Marie
    Morin, Vincent
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 108 (06): : E61 - E65
  • [48] An in vivo comparison of two frequency-based electronic apex locators
    Welk, AR
    Baumgartner, JC
    Marshall, JG
    JOURNAL OF ENDODONTICS, 2003, 29 (08) : 497 - 500
  • [49] Accuracy of two different apex locators in primary teeth with and without root resorption
    Haluk Bodur
    Mesut Odabaş
    Özlem Tulunoğlu
    Ali Cemal Tinaz
    Clinical Oral Investigations, 2008, 12
  • [50] Accuracy of Three Electronic Apex Locators in Anterior and Posterior Teeth: An Ex Vivo Study
    Mancini, Manuele
    Felici, Roberto
    Conte, Gabriele
    Costantini, Marco
    Cianconi, Luigi
    JOURNAL OF ENDODONTICS, 2011, 37 (05) : 684 - 687