Transthoracic electrical impedance during external defibrillation: comparison of measured and modelled waveforms

被引:10
|
作者
Al Hatib, F
Trendafilova, E
Daskalov, I
机构
[1] Bulgarian Acad Sci, Ctr Biomed Engn, BU-1113 Sofia, Bulgaria
[2] Natl Ctr Cardiovasc Dis & Rehabil, Sofia 1309, Bulgaria
关键词
defibrillation; transthoracic impedance; modelling;
D O I
10.1088/0967-3334/21/1/318
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The transthoracic electrical impedance is an important defibrillation parameter, affecting the defibrillating current amplitude and energy, and therefore the defibrillation efficiency. A close relationship between transthoracic impedance and defibrillation success rate was observed. Pre-shock measurements (using low amplitude high frequency current) of the impedance were considered a solution for selection of adequate shock voltages or for current-based defibrillation dosage. A recent approach, called 'impedance-compensating defibrillation' was implemented, where the pulse duration was controlled with respect to the impedance measured during the initial phase of the shock. These considerations raised our interest in reassessment of the transthoracic impedance characteristics and the corresponding measurement methods. The purpose of this work is to study the variations of the transthoracic impedance by a continuous measurement technique during the defibrillation shock and comparing the data with results obtained by modelling. Voltage and current impulse waveforms were acquired during cardioversion of patients with atrial fibrillation or flutter. The same type of defibrillation pulse was taken from dogs after induction of fibrillation. The electrodes were located in the anterior position, for both the patients and animals.
引用
收藏
页码:145 / 153
页数:9
相关论文
共 50 条
  • [1] Differential contribution of skin impedance and thoracic volume to transthoracic impedance during external defibrillation
    Deakin, CD
    Sado, DM
    Petley, GW
    Clewlow, F
    RESUSCITATION, 2004, 60 (02) : 171 - 174
  • [2] Quadriphasic waveforms are superior to triphasic waveforms for transthoracic defibrillation in a cardiac arrest swine model with high impedance
    Zhang, Y
    Rhee, B
    Davies, LR
    Zimmerman, MB
    Snyder, D
    Jones, JL
    Kerber, RE
    RESUSCITATION, 2006, 68 (02) : 251 - 258
  • [3] Comparison of the effects of removal of chest hair with not doing so before external defibrillation on transthoracic impedance
    Sado, DM
    Deakin, CD
    Petley, GW
    Clewlow, F
    AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (01): : 98 - 100
  • [4] Transthoracic impedance during external defibrillation is related to cardiac arrest site and electrode pad size
    Mathew, TP
    McKay, YR
    Harbinson, MT
    Adgey, AAJ
    Dalzell, GWN
    CIRCULATION, 1998, 98 (17) : 410 - 410
  • [5] Changes in transthoracic impedance during sequential biphasic defibrillation
    Deakin, Charles D.
    Ambler, Jonathan J. S.
    Shaw, Steven
    RESUSCITATION, 2008, 78 (02) : 141 - 145
  • [6] A Systematic Review of the Transthoracic Impedance during Cardiac Defibrillation
    Heyer, Yasmine
    Baumgartner, Daniela
    Baumgartner, Christian
    SENSORS, 2022, 22 (07)
  • [7] ELECTRODE-SUBJECT IMPEDANCE DURING TRANSTHORACIC VENTRICULAR DEFIBRILLATION
    GEDDES, LA
    TACKER, WA
    CABLER, PS
    GOTHARD, R
    FEDERATION PROCEEDINGS, 1974, 33 (03) : 332 - 332
  • [8] Ventricular pacing threshold after transthoracic external defibrillation with two different waveforms: an experimental study
    Assumpcao, Antonio Carlos
    Martins de Oliveira, Pedro Paulo
    de Souza Vilarinho, Karlos Alexandre
    Eghtesady, Pirooz
    Silveira Filho, Lindemberg Mota
    Ramos Lavagnoli, Carlos Fernando
    Barbosa de Oliveira Severino, Elaine Soraya
    Petrucci, Orlando
    EUROPACE, 2013, 15 (02): : 297 - 302
  • [9] A comparison of transthoracic impedance using standard defibrillation paddles and self-adhesive defibrillation pads
    Deakin, CD
    McLaren, RM
    Petley, GW
    Clewlow, F
    Dalrymple-Hay, MJR
    RESUSCITATION, 1998, 39 (1-2) : 43 - 46
  • [10] Electrical Cardioversion for Atrial Fibrillation; Initial and Subsequent Change in Transthoracic Impedance Predicts the Likelihood to Require Charge Escalation to High Energies During External Defibrillation
    Mohammedzein, Assad
    Walsh, Katie
    Abubakr, Abushama
    Owens, Patrick
    CIRCULATION, 2019, 140