Metallurgical aspects of failure in a broken femoral HIP prosthesis

被引:20
|
作者
Khouzani, M. Kiani [1 ]
Bahrami, A. [1 ]
Eslami, A. [1 ]
机构
[1] Isfahan Univ Technol, Dept Mat Engn, Esfahan 8415683111, Iran
关键词
Failure analysis; AISI 316L stainless steels; HIP prosthesis; Fatigue fracture; FATIGUE FAILURE; STEM; FRACTURE;
D O I
10.1016/j.engfailanal.2018.03.018
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
In this research failure of femoral hip prosthesis, made of AISI 316 L stainless steel, in a 69 years old patient with a weight of 70 kg was investigated. Replacement of the damaged cartilage and bone of the hip joint with prosthesis component through Total HIP Arthroplasty (THA) is a widely used surgery. Patients with this type of implants are anticipated to have a normal daily life with capability of performing routine physical activities for a period of 15 years. However, in investigated case, the failure had occurred in< 10 years. Optical and electron microscopy as well as hardness measurements were conducted to investigate possible causes of the failure. Results showed that high cycle fatigue was the failure mechanism. Microstructure of the investigated hip consisted of surprisingly high density of inclusions which should have caused the initiation and growth of fatigue cracks. Possible failure scenarios and mitigation strategies have been discussed in this regard.
引用
收藏
页码:168 / 178
页数:11
相关论文
共 50 条
  • [31] Total hip arthroplasty with the cementless spiron femoral neck prosthesis
    Lugeder, A.
    Haering, E.
    Mueller, A.
    Droste, P.
    Zeichen, J.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2013, 25 (04): : 388 - 397
  • [32] FRACTURE OF THE FEMORAL PROSTHESIS AFTER TOTAL HIP-REPLACEMENT
    MCNEUR, JC
    MEDICAL JOURNAL OF AUSTRALIA, 1984, 141 (06) : 341 - 344
  • [33] Osteolysis of the femoral shaft after implantation of a cemented hip prosthesis
    Katzer, A.
    Behnke, B.
    von, Foerster, G.
    Journal of Biomechanics, 2001, 34 (SUPPL. 1)
  • [34] FRACTURE OF THE FEMORAL STEM OF THE RING TCH HIP-PROSTHESIS
    WILSON, LF
    NOLAN, JF
    HEYWOODWADDINGTON, MB
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (05): : 725 - 728
  • [35] FEMORAL FRACTURES FOLLOWING IMPLANTATION OF TOTAL HIP-PROSTHESIS
    HAHLEN, B
    STAUBLI, HU
    NOESBERGER, B
    HELVETICA CHIRURGICA ACTA, 1982, 49 (1-2) : 61 - 66
  • [36] TOTAL HIP-PROSTHESIS AFTER HIGH FEMORAL OSTEOTOMY
    CARRET, JP
    DEJOUR, H
    BIANCARELLI, P
    BONNIN, M
    GALLAND, O
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1991, 77 (02): : 83 - 91
  • [37] METALLURGICAL ASPECTS OF FAILURE OF POINT PLEASANT BRIDGE
    BENNETT, JA
    MINDLIN, H
    JOURNAL OF TESTING AND EVALUATION, 1973, 1 (02) : 152 - 161
  • [38] FACTORS IN MECHANICAL FAILURE OF FEMORAL COMPONENT IN TOTAL HIP PROSTHESIS - REPORT OF SIX FATIGUE FRACTURES OF FEMORAL STEM AND RESULTS OF EXPERIMENTAL LOADING TESTS
    MARTENS, M
    AERNOUDT, E
    MEESTER, PD
    DUCHEYNE, P
    MULIER, JC
    LANGH, RD
    KESTELIJN, P
    ACTA ORTHOPAEDICA SCANDINAVICA, 1974, 45 (05): : 693 - 710
  • [39] Current and future biocompatibility aspects of biomaterials for hip prosthesis
    Aherwar, Amit
    Singh, Amit K.
    Patnaik, Amar
    AIMS BIOENGINEERING, 2016, 3 (01): : 23 - 43
  • [40] Treatment of femoral bony defects with a modular hip prosthesis and distal interlocking
    Andress, H. -J.
    Junghans, K.
    Schinkel, C.
    Lob, G.
    ZENTRALBLATT FUR CHIRURGIE, 2007, 132 (06): : 547 - 553