Metallurgical aspects of failure in a broken femoral HIP prosthesis

被引:20
|
作者
Khouzani, M. Kiani [1 ]
Bahrami, A. [1 ]
Eslami, A. [1 ]
机构
[1] Isfahan Univ Technol, Dept Mat Engn, Esfahan 8415683111, Iran
关键词
Failure analysis; AISI 316L stainless steels; HIP prosthesis; Fatigue fracture; FATIGUE FAILURE; STEM; FRACTURE;
D O I
10.1016/j.engfailanal.2018.03.018
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
In this research failure of femoral hip prosthesis, made of AISI 316 L stainless steel, in a 69 years old patient with a weight of 70 kg was investigated. Replacement of the damaged cartilage and bone of the hip joint with prosthesis component through Total HIP Arthroplasty (THA) is a widely used surgery. Patients with this type of implants are anticipated to have a normal daily life with capability of performing routine physical activities for a period of 15 years. However, in investigated case, the failure had occurred in< 10 years. Optical and electron microscopy as well as hardness measurements were conducted to investigate possible causes of the failure. Results showed that high cycle fatigue was the failure mechanism. Microstructure of the investigated hip consisted of surprisingly high density of inclusions which should have caused the initiation and growth of fatigue cracks. Possible failure scenarios and mitigation strategies have been discussed in this regard.
引用
收藏
页码:168 / 178
页数:11
相关论文
共 50 条
  • [1] Failure of the uncoated titanium ProxiLock femoral hip prosthesis
    Luites, J. W. H.
    Spruit, M.
    van Hellemondt, G. G.
    Horstmann, W. G.
    Valstar, E. R.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (448) : 79 - 86
  • [2] Failure of the uncoated titanium ProxiLock femoral hip prosthesis
    Stoller, Craig
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (455) : 283 - 283
  • [3] RADIOLOGIC ASPECTS OF NEW UNIVERSAL PROXIMAL FEMORAL HIP PROSTHESIS
    DRINKER, H
    MALL, JC
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1977, 129 (03) : 531 - 533
  • [4] Design of a composite hip femoral prosthesis
    Simoes, JA
    Marques, AT
    MATERIALS & DESIGN, 2005, 26 (05) : 391 - 401
  • [5] Failure of a proximally porous-coated femoral prosthesis in revision total hip arthroplasty
    Woolson, ST
    Delaney, TJ
    JOURNAL OF ARTHROPLASTY, 1995, 10 : S22 - S28
  • [6] CEMENTLESS MODULAR CERAMIC HIP AND FEMORAL PROSTHESIS
    HUCKSTEP, RL
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1984, 66 (05): : 787 - 787
  • [7] USE OF HIP PROSTHESIS IN FEMORAL NECK FRACTURES
    FUREY, JG
    PIERCE, DJ
    SPENCER, GE
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1961, 177 (02): : 100 - &
  • [8] ECCENTRIC FEMORAL HEADS IN TOTAL HIP PROSTHESIS
    GOODMAN, L
    MCGEE, JW
    RADIOLOGY, 1974, 111 (01) : 235 - 235
  • [9] FEMORAL HEAD PROSTHESIS FOR HYPERTROPHIC ARTHRITIS OF HIP
    NORDBY, EJ
    SACHTJEN, KM
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1968, (57) : 191 - &
  • [10] Behavior of the femoral stem in the Bihapro hip prosthesis
    Hernández-Vaquero, D
    Murcia-Mazón, A
    Paz-Jimenez, J
    Alegre-Mateo, R
    Martínez-García, J
    Pena-Vazquez, J
    ORTHOPEDICS, 1999, 22 (11) : 1049 - 1053