DIMA-magnification mammography: use of phantoms for the investigation of image quality and radiation exposure.

被引:4
|
作者
Hermann, KP
Funke, M
Bruggemeyer, H
Sperner, W
Gruhl, T
Koch, R
机构
[1] UNIV GOTTINGEN, ZENTRUM RADIOL, RONTGENDIAGNOST ABT, D-37073 GOTTINGEN, GERMANY
[2] NIEDERSACHS LANDESAMT OKOL, HANNOVER, GERMANY
[3] FEINFOCUS MED TECH GMBH, GARBSEN, GERMANY
关键词
mammography; direct radiographic magnification; resolution; image contrast; radiation exposure;
D O I
10.1055/s-2007-1015535
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A mammographic system has been developed to provide direct radiographic magnification; investigation of image quality and radiation exposure for surveys and aimed exposures were carried out on phantoms and the results compared with two modern screen methods. Material and method: Image quality was defined by contrast transfer function, resolution of low contrast, contrast of large areas and recognition of low contrast detail, using suitable phantoms. Measurements of entrance surface dose were made with a 1 cm(3) flat chamber. Results: For survey examinations, the new mammographic system provided image quality equal to a screen system but using 50% radiation dose. For localised views, the new system used similar radiation dose but provided significantly better demonstration of diagnostic detail. Conclusion: Using a focal spot of 40-120 mu m, the new mammographic system provides markedly improved direct radiographic magnification. With reduced radiation exposure at times, there is a significant gain in diagnostic information compared with current standard systems.
引用
收藏
页码:304 / 310
页数:7
相关论文
共 50 条
  • [41] Phantom-based analysis of variations in automatic exposure control across three mammography systems: implications for radiation dose and image quality in mammography, DBT, and CEM
    Gennaro, Gisella
    Del Genio, Sara
    Manco, Giuseppe
    Caumo, Francesca
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [42] Direct radiographical high magnification mammography using a new microfocus tube. Experimental analysis of the resolution and radiation exposure
    Post, K
    Hermann, KP
    Funke, M
    Hundertmark, C
    Breiter, N
    Grabbe, E
    RADIOLOGE, 1997, 37 (08): : 604 - &
  • [43] Evaluation of the quality of image for various breast composition and exposure conditions in digital mammography
    Yamada, Maki
    Kato, Yuri
    Fujita, Naotoshi
    Kodera, Yoshie
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [44] Magnification survey and spot view mammography with a new microfocus X-ray unit: detail resolution and radiation exposure
    Funke, M
    Breiter, N
    Hermann, KP
    Oestmann, JW
    Grabbe, E
    EUROPEAN RADIOLOGY, 1998, 8 (03) : 386 - 390
  • [45] Magnification survey and spot view mammography with a new microfocus X-ray unit: detail resolution and radiation exposure
    M. Funke
    N. Breiter
    K.-P. Hermann
    J.-W. Oestmann
    E. Grabbe
    European Radiology, 1998, 8 : 386 - 390
  • [46] Radiation exposure and image quality in chest CT examinations
    Ravenel, J
    Scalzetti, EM
    Huda, W
    Garrisi, W
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (02) : 279 - 284
  • [47] Evaluation of exposure factors of dual-energy contrast-enhanced mammography to optimize radiation dose with improved image quality
    Niroshani, Sachila
    Nakamura, Tokiko
    Michiru, Nikaidou
    Negishi, Toru
    ACTA RADIOLOGICA OPEN, 2022, 11 (08)
  • [48] Survey of Radiation Dose, Image Quality and Equipment Performance of Mammography Units in Taiwan
    Hwang, Y. S.
    Tsai, H. Y.
    Chen, C. C.
    Chia, S. H.
    Lin, J. H.
    Wan, Y. L.
    Hsu, G. C.
    Chen, C. P.
    Liu, H. L.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 417 - 420
  • [49] Objective comparison of image quality and dose between conventional and synchrotron radiation mammography
    Moeckli, R
    Verdun, FR
    Fiedler, S
    Pachoud, M
    Schnyder, P
    Valley, JF
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12): : 3509 - 3523
  • [50] A SURVEY OF RADIATION-DOSE AND IMAGE QUALITY IN MAMMOGRAPHY - AN ONGOING PROGRAM IN ITALY
    RIMONDI, O
    GAMBACCINI, M
    CANDINI, GC
    INDOVINA, PL
    ROSATI, A
    HEALTH PHYSICS, 1987, 52 (04): : 437 - 441