DIMA-magnification mammography: use of phantoms for the investigation of image quality and radiation exposure.

被引:4
|
作者
Hermann, KP
Funke, M
Bruggemeyer, H
Sperner, W
Gruhl, T
Koch, R
机构
[1] UNIV GOTTINGEN, ZENTRUM RADIOL, RONTGENDIAGNOST ABT, D-37073 GOTTINGEN, GERMANY
[2] NIEDERSACHS LANDESAMT OKOL, HANNOVER, GERMANY
[3] FEINFOCUS MED TECH GMBH, GARBSEN, GERMANY
关键词
mammography; direct radiographic magnification; resolution; image contrast; radiation exposure;
D O I
10.1055/s-2007-1015535
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A mammographic system has been developed to provide direct radiographic magnification; investigation of image quality and radiation exposure for surveys and aimed exposures were carried out on phantoms and the results compared with two modern screen methods. Material and method: Image quality was defined by contrast transfer function, resolution of low contrast, contrast of large areas and recognition of low contrast detail, using suitable phantoms. Measurements of entrance surface dose were made with a 1 cm(3) flat chamber. Results: For survey examinations, the new mammographic system provided image quality equal to a screen system but using 50% radiation dose. For localised views, the new system used similar radiation dose but provided significantly better demonstration of diagnostic detail. Conclusion: Using a focal spot of 40-120 mu m, the new mammographic system provides markedly improved direct radiographic magnification. With reduced radiation exposure at times, there is a significant gain in diagnostic information compared with current standard systems.
引用
收藏
页码:304 / 310
页数:7
相关论文
共 50 条
  • [31] Effect of Collimation on Radiation Exposure and Image Quality
    Ghobadifar, Mohamed Amin
    Zarei, Safar
    KOREAN JOURNAL OF PAIN, 2013, 26 (03): : 307 - 308
  • [32] Importance of Collimation for Image Quality and Radiation Exposure
    Toussaint, Kaolane
    Khan, Tommy
    Velez, Kerly Cedeno
    Alicea, Christopher
    RADIOLOGIC TECHNOLOGY, 2023, 94 (03) : 206 - 207
  • [33] A population of pediatric phantoms to investigate the tradeoff between radiation dose and image quality
    O'Reilly, Shannon
    Li, Wenxiao
    Frey, Eric
    Treves, S. Ted
    Sgouros, George
    Fahey, Frederic
    Bolch, Wesley
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [34] Relationship between quality of reconstructed 3-dimensional CT image and patient exposure.
    Honda, E
    Hisamitsu, Y
    Morita, Y
    Nguyen, VL
    Nagaike, S
    Noikura, T
    Sasaki, T
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : A431 - A431
  • [35] RADIATION-LEVELS OF MACRORADIOGRAPHY - RADIATION EXPOSURE AND IMAGE QUALITY
    TAKAHASH.S
    SAKUMA, S
    AYAKAWA, Y
    MAEKOSHI, H
    OHARA, K
    RADIOLOGY, 1974, 112 (03) : 709 - 713
  • [36] Can radiation dose in mammography be further reduced by increasing the image quality?
    O'Leary, D.
    Rainford, L.
    BREAST CANCER RESEARCH, 2011, 13
  • [37] Can radiation dose in mammography be further reduced by increasing the image quality?
    D O'Leary
    L Rainford
    Breast Cancer Research, 13
  • [38] The relationship between compression force, image quality and radiation dose in mammography
    Korf, A.
    Herbst, C. P.
    Rae, W. I. D.
    SA JOURNAL OF RADIOLOGY, 2009, 13 (04): : 86 - 92
  • [39] BIMETAL ANODE WITH TUNGSTEN OR RHODIUM - COMPARATIVE-STUDIES ON IMAGE QUALITY AND RADIATION EXPOSURE IN SCREEN-FILM MAMMOGRAPHY
    FUNKE, M
    HERMANN, KP
    BREITER, N
    MORITZ, J
    MULLER, D
    GRABBE, E
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 163 (05): : 388 - 394
  • [40] Image quality in conventional film screen system, digital phosphor storage plate mammography in magnification technique and digital mammography in CCD-technique
    Schulz-Wendtland, R
    Aichinger, U
    Säbel, M
    Böhner, C
    Dobritz, M
    Bautz, W
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (12): : 965 - 968