Limited significance of screening computed tomography after cementless total hip arthroplasty with highly cross-linked polyethylene at 7-10 years of follow-up

被引:1
|
作者
Takada, Ryohei [1 ]
Jinno, Tetsuya [1 ]
Koga, Daisuke [1 ]
Yamauchi, Yuki [1 ]
Asou, Yoshinori [1 ]
Muneta, Takeshi [1 ]
Okawa, Atsushi [1 ]
机构
[1] Tokyo Med & Dent Univ, Hosp Med, Dept Orthopaed Surg, Tokyo, Japan
关键词
Annealed; Fit and fill stem; Modular acetabular cup; Radiographic analysis; Remelted; PELVIC OSTEOLYSIS; WEAR; OUTCOMES; CT;
D O I
10.3109/14397595.2015.1131136
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: The purpose of this retrospective study is to report the incidence of osteolysis and evaluate the significance of screening computed tomography (CT) compared to plain radiography in detecting osteolysis after total hip arthroplasty with metal-on-highly crosslinked polyethylene bearings. Methods: We retrospectively reviewed 264 primary cementless total hip arthroplasties of 211 patients, 24 males, 187 females, who received postoperative screening CT scan in addition to radiography at postoperative 7-10 years (average 8.2 years). First-generation highly crosslinked polyethylene was used in all cases. Results: On the plain radiographs, no acetabular osteolysis (0%) and two cases of femoral osteolysis (0.8%) were found in the follow-up period. No osteolysis was newly found by screening CT scan. Conclusions: Very low incidence of osteolysis after total hip arthroplasty with highly cross-linked polyethylene at postoperative 7-10 years was confirmed, and routine screening CT scan for detecting osteolysis in this setting was not supported from this study.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [41] Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years
    Dorr, LD
    Wan, ZN
    Shahrdar, C
    Sirianni, L
    Boutary, M
    Yun, A
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08): : 1816 - 1821
  • [42] Primary cemented total hip arthroplasty: 10 years follow-up
    Rajendra Nath
    Anil Kumar Gupta
    Unmesh Chakravarty
    Rohit Nath
    Indian Journal of Orthopaedics, 2010, 44 : 283 - 288
  • [43] No higher tibial post fracture rate with highly cross-linked polyethylene inserts in total knee arthroplasty at a minimum 10-year follow-up
    Baek, Ji-Hoon
    Lee, Su Chan
    Ryu, Suengryol
    Ahn, Hye Sun
    Nam, Chang Hyun
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (09) : 5807 - 5813
  • [44] Primary cemented total hip arthroplasty: 10 years follow-up
    Nath, Rajendra
    Gupta, Anil Kumar
    Chakravarty, Unmesh
    Nath, Rohit
    INDIAN JOURNAL OF ORTHOPAEDICS, 2010, 44 (03) : 283 - 288
  • [45] Comparison of cementless total hip arthroplasty survivorship between metal-on-highly cross-linked polyethylene and ceramic on ceramic bearings: A case control study with a 5-9-year follow-up
    Higuchi, Yoshitoshi
    Seki, Taisuke
    Hasegawa, Yukiharu
    Morita, Daigo
    Komatsu, Daigo
    Takegami, Yasuhiko
    Ishiguro, Naoki
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2018, 104 (05) : 663 - 669
  • [46] No higher tibial post fracture rate with highly cross-linked polyethylene inserts in total knee arthroplasty at a minimum 10-year follow-up
    Ji-Hoon Baek
    Su Chan Lee
    Suengryol Ryu
    Hye Sun Ahn
    Chang Hyun Nam
    Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 5807 - 5813
  • [47] Minimum 10-Year Follow-Up of Cementless Total Hip Arthroplasty With a 32-mm Cobalt-Chromium Head on Highly Cross-Linked Polyethylene and a Tapered, Fiber Metal Proximally Coated Femoral Stem
    Naito, Yohei
    Hasegawa, Masahiro
    Tone, Shine
    Wakabayashi, Hiroki
    Sudo, Akihiro
    JOURNAL OF ARTHROPLASTY, 2021, 36 (02): : 647 - 652
  • [48] Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty?
    Glyn-Jones, Sion
    Saac, Sherif
    Hauptfleisch, Jennifer
    McLardy-Smith, Peter
    Murray, David W.
    Singh, Harinderjit
    JOURNAL OF ARTHROPLASTY, 2008, 23 (03): : 337 - 343
  • [49] Cementless Mobile-Bearing Total Knee Arthroplasty: 10 Years Follow-Up
    Abdallatif, Ahmad G.
    Winkworth, Corrina
    Aslam, Nadim
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (04)
  • [50] Large Femoral Heads in Total Hip Arthroplasty With Vitamin E Highly Cross-Linked Polyethylene: Head Penetration Rates Compared to Highly Cross-Linked Polyethylene
    Nielson, Trent
    Owens, Garrett
    Miller, Brandon
    Meneghini, Ethan
    Deckard, Evan R.
    Meneghini, R. Michael
    JOURNAL OF ARTHROPLASTY, 2022, 37 (07): : S685 - S691