Analysis of non-linear ion drift in spectrometers of ion mobility increment with cylindrical drift chamber

被引:4
|
作者
Elistratov, A. A. [1 ]
Shibkov, S. V.
Nikolaev, E. N.
机构
[1] Inst Cryptosyst Commun & Informat, Moscow, Russia
[2] Russian Acad Sci, Inst Chem Phys Energy Problems, Moscow V71, Russia
关键词
ion mobility; theory; FAIMS; differential ion mobility;
D O I
10.1255/ejms.796
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A model for the drift of ions under a non-uniform, high-frequency electric field in the drift chamber of a spectrometer of ion mobility increment is developed. For the general dependence of the ion mobility on the electric field strength and the general time-dependence of the separating voltage, we suggest a procedure for calculating of the ion peak form. The shape of the peak for the ion focusing and defocusing conditions has been obtained.
引用
收藏
页码:153 / 160
页数:8
相关论文
共 50 条
  • [11] Analytical model for the signal-to-noise-ratio of drift tube ion mobility spectrometers
    Kirk, Ansgar T.
    Bohnhorst, Alexander
    Zimmermann, Stefan
    TM-TECHNISCHES MESSEN, 2021, 88 (05) : 262 - 273
  • [12] NON-LINEAR ION-CYCLOTRON AND DRIFT-CYCLOTRON WAVES IN NONTHERMAL PLASMAS
    MYRA, JR
    LIU, CS
    CHEN, HH
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1978, 23 (07): : 794 - 794
  • [13] A NON-LINEAR MODEL FOR TOROIDAL DRIFT MODES DRIVEN BY ION PRESSURE-GRADIENTS
    STEY, GC
    HORTON, W
    PHYSICS LETTERS A, 1981, 81 (05) : 268 - 270
  • [14] Numerical simulation of ion drift within ion mobility spectrometers in high peclet conditions using FEM techniques
    Kalms, A.
    Salleras, M.
    Liu, Z.
    Korvink, J. G.
    Goebel, J.
    Mueller, G.
    Samitier, J.
    Marco, S.
    EUROSIME 2007: THERMAL, MECHANICAL AND MULTI-PHYSICS SIMULATION AND EXPERIMENTS IN MICRO-ELECTRONICS AND MICRO-SYSTEMS, PROCEEDINGS, 2007, : 594 - +
  • [15] DRIFT TIMES IN ION-CYCLOTRON RESONANCE SPECTROMETERS
    VANDERHART, WJ
    VANSPRANG, HA
    CHEMICAL PHYSICS LETTERS, 1975, 36 (02) : 215 - 217
  • [16] Performance evaluation of a miniature ion mobility spectrometer drift cell for application in hand-held explosives detection ion mobility spectrometers
    J. S. Babis
    R. P. Sperline
    A. K. Knight
    D. A. Jones
    C. A. Gresham
    M. B. Denton
    Analytical and Bioanalytical Chemistry, 2009, 395 : 411 - 419
  • [17] Performance evaluation of a miniature ion mobility spectrometer drift cell for application in hand-held explosives detection ion mobility spectrometers
    Babis, J. S.
    Sperline, R. P.
    Knight, A. K.
    Jones, D. A.
    Gresham, C. A.
    Denton, M. B.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 395 (02) : 411 - 419
  • [18] ION DISTRIBUTION PROFILES IN THE DRIFT REGION OF AN ION MOBILITY SPECTROMETER
    KARPAS, Z
    EICEMAN, GA
    EWING, RG
    ALGOM, A
    AVIDA, R
    FRIEDMAN, M
    MATMOR, A
    SHAHAL, O
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1993, 127 : 95 - 104
  • [19] A model of nonlinear ion drift spectrometry for gas detectors with separating chamber of cylindrical geometry
    A. A. Elistratov
    S. V. Shibkov
    Technical Physics Letters, 2004, 30 : 183 - 185
  • [20] A model of nonlinear ion drift spectrometry for gas detectors with separating chamber of cylindrical geometry
    Elistratov, AA
    Shibkov, SV
    TECHNICAL PHYSICS LETTERS, 2004, 30 (03) : 183 - 185