Patient-specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT)

被引:11
|
作者
Chow, Vivian U. Y. [1 ]
Kan, Monica W. K. [1 ,2 ]
Chan, Anthony T. C. [1 ,2 ]
机构
[1] Prince Wales Hosp, Dept Clin Oncol, Hong Kong, Peoples R China
[2] Chinese Univ Hong Kong, Dept Clin Oncol, Hong Kong, Peoples R China
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 11期
关键词
patient-specific QA; log file analysis; stereotactic body radiation therapy; volumetric-modulated arc therapy; plan delivery accuracy; TRAJECTORY LOG; DELIVERY; IMRT; ERRORS; IMPLEMENTATION;
D O I
10.1002/acm2.13053
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An in-house trajectory log analysis program (LOGQA) was developed to evaluate the delivery accuracy of volumetric-modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT). Methods have been established in LOGQA to provide analysis on dose indices, gantry angles, and multi-leaf collimator (MLC) positions. Between March 2019 and May 2020, 120 VMAT SBRT plans of various treatment sites using flattening filter-free (FFF) mode were evaluated using both LOGQA and phantom measurements. Gantry angles, dose indices, and MLC positions were extracted from log and compared with each plan. Integrated transient fluence map (ITFM) was reconstructed from log to examine the deviation of delivered fluence against the planned one. Average correlation coefficient of dose index versus gantry angle and ITFM for all patients were 1.0000, indicating that the delivered beam parameters were in good agreement with planned values. Maximum deviation of gantry angles and monitor units (MU) of all patients were less than 0.2 degree and 0.03 % respectively. Regarding MLC positions, maximum and root-mean-square (RMS) deviations from planned values were less than 0.6 mm and 0.3 mm respectively, indicating that MLC positions during delivery followed planned values in precise manner. Results of LOGQA were consistent with measurement, where all gamma-index passing rates were larger than 95 %, with 2 %/2 mm criteria. Three types of intentional errors were introduced to patient plan for software validation. LOGQA was found to recognize the introduced errors of MLC positions, gantry angles, and dose indices with magnitudes of 1 mm, 1 degree, and 5 %, respectively, which were masked in phantom measurement. LOGQA was demonstrated to have the potential to reduce or even replace patient-specific QA measurements for SBRT plan delivery provided that the frequency and amount of measurement-based machine-specific QA can be increased to ensure the log files record real values of machine parameters.
引用
收藏
页码:179 / 187
页数:9
相关论文
共 50 条
  • [21] Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance
    Stathakis, Sotirios
    Mavroidis, Panayiotis
    Narayanasamy, Ganesh
    Markovic, Miljenko
    Myers, Pamela
    Papanikolaou, Niko
    JOURNAL OF BUON, 2015, 20 (04): : 1154 - 1163
  • [22] Quality Assurance for Stereotactic Body Radiation Therapy for Gynecologic Malignancies
    Andersen, Samuel N.
    Bonnen, Mark D.
    Ludwig, Michelle S.
    Dalwadi, Shraddha M.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (02)
  • [23] Analysis of Characteristics of Phantomless Patient-Specific Quality Assurance Using PerFRACTIONTM
    Shin, J.
    Kim, S. J.
    Kim, H.
    Lee, H.
    Ahn, Y. C.
    Pyo, H. R.
    Park, W.
    Kim, H.
    Park, H. C.
    MEDICAL PHYSICS, 2024, 51 (10) : 7958 - 7958
  • [24] Monte Carlo Dose Calculations for Patient-Specific QA Based On Machine Log Files
    Parenica, H.
    Kabat, C.
    Papaconstadopoulos, P.
    Kirby, N.
    Myers, P.
    Wagner, T.
    Papanikolaou, N.
    Stathakis, S.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [25] Applications of artificial intelligence for machine- and patient-specific quality assurance in radiation therapy: current status and future directions
    Ono, Tomohiro
    Iramina, Hiraku
    Hirashima, Hideaki
    Adachi, Takanori
    Nakamura, Mitsuhiro
    Mizowaki, Takashi
    JOURNAL OF RADIATION RESEARCH, 2024, 65 (04) : 421 - 432
  • [26] A comparison of three-film analysis software for stereotactic radiotherapy patient-specific quality assurance
    Buddhavarapu, Athreya
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (03):
  • [27] Patient-specific quality assurance in SBRT treatments using 3D polymer gel dosimetry
    Guadarrama-Huerta, P. J.
    Arzaga-Barajas, E.
    Rodriguez-Laguna, A.
    Jimenez-Acosta, J. A.
    Poitevin-Chacon, M. A.
    Massillon-Jl, G.
    RADIATION MEASUREMENTS, 2024, 175
  • [28] Quick, efficient and effective patient-specific intensity-modulated radiation therapy quality assurance using log file and electronic portal imaging device
    Kumar, Rajesh
    Amols, H. I.
    Lovelock, M.
    Sharma, S. D.
    Datta, D.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (02) : 297 - 303
  • [29] Comparing log file to measurement-based patient-specific quality assurance
    Li Ting Chan
    Yun Inn Tan
    Poh Wee Tan
    Yuh Fun Leong
    Jong Shin Khor
    Mun Woan Teh
    Joan Faith Loria Cruz
    Sháun Peter Baggarley
    Kiat Huat Ooi
    Yiat Horng Leong
    Physical and Engineering Sciences in Medicine, 2023, 46 : 303 - 311
  • [30] Evaluation of digital linac log data for patient-specific VMAT quality assurance
    U'wais, F.
    Zin, H. M.
    Radzi, Y.
    Rizan, N. N.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1346 - S1346