On the efficiency calibration of Si(Li) detector in the low-energy region using thick-target bremsstrahlung

被引:14
|
作者
An, Z [1 ]
Liu, MT [1 ]
机构
[1] Sichuan Univ, Inst Nucl Sci & Technol, Minist Educ, Key Lab Radiat Phys & Technol, Chengdu 610064, Peoples R China
关键词
efficiency calibration; bremsstrahlung; Si(Li) detector;
D O I
10.1016/S0168-583X(02)01033-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, the efficiency calibration of a Si(Li) detector in the low-energy region down to 0.58 keV has been performed using thick-carbon-target bremsstrahlung by 19 keV electron impact. The shape of the efficiency calibration curve was determined from the thick-carbon-target bremsstrahlung spectrum, and the absolute value for the efficiency calibration was obtained from the use of(241) Am radioactive standard source. The modified Wentzel's formula for thick-target bremsstrahlung was employed and it was also compared with the most recently developed theoretical model based upon the doubly differential cross-sections for bremsstrahlung of Kissel, Quarles and Pratt. In the present calculation of theoretical bremsstrahlung, the self-absorption correction and the convolution of detector's response function with the bremsstrahlung spectrum have simultaneously been taken into account. The accuracy for the efficiency calibration in the low-energy region with the method described here was estimated to be about 6%. Moreover, the self-absorption correction calculation based upon the prescription of Wolters et al. has also been presented as an analytical factor with the accuracy of similar to1%. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:513 / 518
页数:6
相关论文
共 50 条
  • [1] Low-energy intensity calibration of a Si(Li) detector using thick-target bremsstrahlung
    Wolters, U
    Meyer, D
    Wiesemann, K
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1998, 31 (17) : 2112 - 2116
  • [2] Efficiency calibration of a Si(Li) detector in the low-energy region using electron bremsstrahlung
    Tschischgale, J
    Kuchler, D
    Lehnert, U
    Zschornack, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 400 (2-3): : 387 - 391
  • [3] Efficiency calibration of Si(Li) detector in low-energy region
    An, Zhu
    Liu, Man-Tian
    Wu, Ying
    Duan, Yan-Min
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2006, 40 (01): : 83 - 87
  • [4] ABSOLUTE CALIBRATION OF LOW-ENERGY, THICK TARGET BREMSSTRAHLUNG
    VONNEUMANNCOSEL, P
    HUXEL, N
    RICHTER, A
    SPIELER, C
    CARROLL, JJ
    COLLINS, CB
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 338 (2-3): : 425 - 431
  • [5] RELATIVE EFFICIENCY CALIBRATION OF GE(LI) DETECTOR IN LOW-ENERGY REGION
    BABA, S
    SUZUKI, T
    JOURNAL OF RADIOANALYTICAL CHEMISTRY, 1976, 29 (02): : 301 - 308
  • [6] A model for low-energy thick-target bremsstrahlung produced in a scanning electron microscope
    Semaan, M
    Quarles, CA
    X-RAY SPECTROMETRY, 2001, 30 (01) : 37 - 43
  • [7] Z-dependence of thick-target bremsstrahlung produced by monoenergetic low-energy electrons
    Czarnecki, S.
    Short, A.
    Williams, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2016, 378 : 54 - 58
  • [8] Si(Li) detector efficiency and peak shape calibration in the low-energy range using synchrotron radiation
    Lepy, MC
    Plagnard, J
    Stemmler, P
    Ban, G
    Beck, L
    Dhez, P
    X-RAY SPECTROMETRY, 1997, 26 (04) : 195 - 202
  • [9] EFFICIENCY CALIBRATION OF A GE(IN) SEMICONDUCTOR DETECTOR BY THIN-TARGET AND THICK-TARGET PIXE
    KALLITHRAKASKONTOS, N
    KATSANOS, AA
    X-RAY SPECTROMETRY, 1994, 23 (02) : 96 - 99
  • [10] EFFICIENCY CALIBRATION OF LOW-ENERGY SI(LI) PHOTON SPECTROMETER
    GEHRKE, RJ
    LOKKEN, RA
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1971, 14 (01): : 126 - &