Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring

被引:103
|
作者
Kim, Chang-Sei [1 ]
Carek, Andrew M. [2 ]
Mukkamala, Ramakrishna [3 ]
Inan, Omer T. [2 ]
Hahn, Jin-Oh [1 ]
机构
[1] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
[2] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[3] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
基金
美国国家卫生研究院;
关键词
Ballistocardiogram (BCG); blood pressure; electrocardiogram (ECG); pulse arrival time (PAT); pulse transit time (PTT); ARRIVAL-TIME; PREEJECTION PERIOD; HEART-RATE; CONTRACTILITY; CALIBRATION; MARKER; ECG;
D O I
10.1109/TBME.2015.2440291
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Goal: We tested the hypothesis that the ballistocardiogram (BCG) waveform could yield a viable proximal timing reference for measuring pulse transit time (PTT). Methods: From 15 healthy volunteers, we measured PTT as the time interval between BCG and a noninvasively measured finger blood pressure (BP) waveform. To evaluate the efficacy of the BCG-based PTT in estimating BP, we likewise measured pulse arrival time (PAT) using the electrocardiogram (ECG) as proximal timing reference and compared their correlations to BP. Results: BCG-based PTT was correlated with BP reasonably well: the mean correlation coefficient (r) was 0.62 for diastolic (DP), 0.65 for mean (MP), and 0.66 for systolic (SP) pressures when the intersecting tangent method was used as distal timing reference. Comparing four distal timing references (intersecting tangent, maximum second derivative, diastolic minimum, and systolic maximum), PTT exhibited the best correlation with BP when the systolic maximum method was used (mean r value was 0.66 for DP, 0.67 for MP, and 0.70 for SP). PTT was more strongly correlated with DP than PAT regardless of the distal timing reference: mean r value was 0.62 versus 0.51 (p = 0.07) for intersecting tangent, 0.54 versus 0.49 (p = 0.17) for maximum second derivative, 0.58 versus 0.52 (p = 0.37) for diastolic minimum, and 0.66 versus 0.60 (p = 0.10) for systolic maximum methods. The difference between PTT and PAT in estimating DP was significant (p = 0.01) when the r values associated with all the distal timing references were compared altogether. However, PAT appeared to outperform PTT in estimating SP (p = 0.31 when the r values associated with all the distal timing references were compared altogether). Conclusion: We conclude that BCG is an adequate proximal timing reference in deriving PTT, and that BCG-based PTT may be superior to ECG-based PAT in estimating DP. Significance: PTT with BCG as proximal timing reference has potential to enable convenient and ubiquitous cuffless BP monitoring.
引用
收藏
页码:2657 / 2664
页数:8
相关论文
共 50 条
  • [41] Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation
    Schmalgemeier, Heidi
    Bitter, Thomas
    Bartsch, Stephan
    Bullert, Kevin
    Fischbach, Thomas
    Eckert, Siegfried
    Horstkotte, Dieter
    Oldenburg, Olaf
    SLEEP AND BREATHING, 2012, 16 (04) : 1105 - 1112
  • [42] Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation
    Heidi Schmalgemeier
    Thomas Bitter
    Stephan Bartsch
    Kevin Bullert
    Thomas Fischbach
    Siegfried Eckert
    Dieter Horstkotte
    Olaf Oldenburg
    Sleep and Breathing, 2012, 16 : 1105 - 1112
  • [43] CONTINUOUS AND NON-INVASIVE BLOOD PRESSURE MEASUREMENT BASED ON PULSE TRANSIT TIME COMPARED TO AMBULATORY BLOOD PRESSURE MEASUREMENT
    Haberl, R.
    Wesely, M.
    Thimel, K.
    Gehring, J.
    Kuechler, G.
    JOURNAL OF HYPERTENSION, 2016, 34 : E287 - E287
  • [44] A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
    Buxi, Dilpreet
    Redoute, Jean-Michel
    Yuce, Mehmet Rasit
    PHYSIOLOGICAL MEASUREMENT, 2015, 36 (03) : R1 - R26
  • [45] Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice
    Mukkamala, Ramakrishna
    Hahn, Jin-Oh
    Inan, Omer T.
    Mestha, Lalit K.
    Kim, Chang-Sei
    Toreyin, Hakan
    Kyal, Survi
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (08) : 1879 - 1901
  • [46] The Wavelet Transform of Pulse Wave and Electrocardiogram Improves Accuracy of Blood Pressure Estimation in Cuffless Blood Pressure Measurement
    Yamanaka, Syunsuke
    Morikawa, Koji
    Yamamura, Osamu
    Morita, Hiroshi
    Huh, Ji Young
    CIRCULATION, 2016, 134
  • [47] Validation of Two Pulse Transit Time-Based Blood Pressure Measurement Devices
    Bard, Dylan M.
    Hahnen, Christina
    Joseph, Jeffrey I.
    van Helmond, Noud
    FASEB JOURNAL, 2019, 33
  • [48] Restrictive factors of the measurement accuracy to estimate Blood Pressure with Pulse wave Transit Time
    Chen, X. M.
    Li, Y. B.
    Zhang, Y.
    Deng, N.
    BIOINFORMATICS AND BIOMEDICAL ENGINEERING: NEW ADVANCES, 2016, : 435 - 441
  • [49] An Investigation of Pulse Transit Time as a Non-Invasive Blood Pressure Measurement Method
    McCarthy, B. M.
    O'Flynn, B.
    Mathewson, A.
    SENSORS & THEIR APPLICATIONS XVI, 2011, 307
  • [50] Smartphone-based Continuous Blood Pressure Measurement Using Pulse Transit Time
    Gholamhosseini, Hamid
    Meintjes, Andries
    Baig, Mirza
    Linden, Maria
    PHEALTH 2016, 2016, 224 : 84 - 89