Calibration Tools for Scanning Thermal Microscopy Probes Used in Temperature Measurement Mode

被引:6
|
作者
Nguyen, T. P. [1 ]
Thiery, L. [1 ]
Euphrasie, S. [1 ]
Lemaire, E. [2 ]
Khan, S. [2 ]
Briand, D. [2 ]
Aigouy, L. [3 ,4 ]
Gomes, S. [5 ]
Vairac, P. [1 ]
机构
[1] Univ Franche Comte, CNRS, ENSMM, UTBM,FEMTO ST Inst,UMR 6174, F-25030 Besancon, France
[2] Ecole Polytech Fed Lausanne, Soft Transducers Lab, CH-2002 Neuchatel, Switzerland
[3] ESPCI Paris, LPEM, F-75005 Paris, France
[4] Sorbonne Univ, PSL Res Univ, CNRS, F-75006 Paris, France
[5] Univ Claude Bernard Lyon 1, INSA Lyon, CNRS, Univ Lyon,CETHIL,UMR5008, F-69621 Villeurbanne, France
来源
关键词
scanning thermal microscopy; temperature measurement; calibration; microhotplate; NANOSCALE THERMOMETRY; HEAT-TRANSFER; CONTACT;
D O I
10.1115/1.4043381
中图分类号
O414.1 [热力学];
学科分类号
摘要
We demonstrate the functionality of a new active thermal microchip dedicated to the temperature calibration of scanning thermal microscopy (SThM) probes. The silicon micro-machined device consists in a suspended thin dielectric membrane in which a heating resistor with a circular area of 50 mu m in diameter was embedded. A circular calibration target of 10 mu m in diameter was patterned at the center and on top of the membrane on which the SThM probe can land. This target is a resistive temperature detector (RTD) that measures the surface temperature of the sample at the level of the contact area. This allows evaluating the ability of any SThM probe to measure a surface temperature in ambient air conditions. Furthermore, by looking at the thermal balance of the device, the heat dissipated through the probe and the different thermal resistances involved at the contact can be estimated. A comparison of the results obtained for two different SThM probes, microthermocouples and probes with a fluorescent particle is presented to validate the functionality of the micromachined device. Based on experiments and simulations, an analysis of the behavior of probes allows pointing out their performances and limits depending on the sample characteristics whose role is always preponderant. Finally, we also show that a smaller area of the temperature sensor would be required to assess the local disturbance at the contact point.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Microscale temperature measurement by scanning thermal microscopy
    Nakabeppu, O
    Suzuki, T
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2002, 69 (03): : 727 - 737
  • [2] Microscale temperature measurement by scanning thermal microscopy
    O. Nakabeppu
    T. Suzuki
    Journal of Thermal Analysis and Calorimetry, 2002, 69 : 727 - 737
  • [3] Micromachined temperature calibration tool for contact Scanning Thermal Microscope probes
    Bontempi, A.
    Nguyen, T. P.
    Lemaire, E.
    Thiery, L.
    Teyssieux, D.
    Euphrasie, S.
    Briand, D.
    Vairac, P.
    2015 21ST INTERNATIONAL WORKSHOP ON THERMAL INVESTIGATIONS OF ICS AND SYSTEMS (THERMINIC), 2015,
  • [4] Calibration methodologies for scanning thermal microscopy
    Guen, Eloise
    Renahy, David
    Massoud, Mouhannad
    Bluet, Jean-Marie
    Chapuis, Pierre-Olivier
    Gomes, Severine
    2016 22ND INTERNATIONAL WORKSHOP ON THERMAL INVESTIGATIONS OF ICS AND SYSTEMS (THERMINIC), 2016, : 110 - 114
  • [5] CALIBRATION AND EVALUATION OF SCANNING-FORCE-MICROSCOPY PROBES
    SHEIKO, SS
    MOLLER, M
    REUVEKAMP, EMCM
    ZANDBERGEN, HW
    PHYSICAL REVIEW B, 1993, 48 (08): : 5675 - 5678
  • [6] Micromachined probes for high frequency scanning force microscopy and scanning thermal microscopy
    Stopka, M
    Munster, S
    Leinhos, T
    Mihalcea, C
    Scholz, W
    Leyk, A
    Mertin, W
    Oesterschulze, E
    MICROMACHINING AND IMAGING, 1997, 3009 : 92 - 100
  • [7] A dark mode in scanning thermal microscopy
    Ramiandrisoa, Liana
    Allard, Alexandre
    Joumani, Youssef
    Hay, Bruno
    Gomes, Severine
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (12):
  • [8] A General Method for Calibration of Active Scanning Thermal Probes
    Tselev, Alexander
    ADVANCED ENGINEERING MATERIALS, 2024, 26 (21)
  • [9] The material dependence of temperature measurement resolution in thermal scanning electron microscopy
    Wu, Xiaowei
    Hull, Robert
    APPLIED PHYSICS LETTERS, 2013, 102 (11)
  • [10] Calibration of conductance channels and heat flux sharing in scanning thermal microscopy combining resistive thermal probes and pyroelectric sensors
    Chirtoc, M.
    Bodzenta, J.
    Kazmierczak-Balata, A.
    INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2020, 156