In vitro and in vivo characterization of SU-8 flexible neuroprobe: From mechanical properties to electrophysiological recording

被引:28
|
作者
Huang, Shun-Ho [1 ]
Lin, Shu-Ping [2 ,3 ]
Chen, Jia-Jin Jason [1 ,4 ,5 ]
机构
[1] Natl Cheng Kung Univ, Dept Biomed Engn, Tainan 70101, Taiwan
[2] Natl Chung Hsing Univ, Grad Inst Biomed Engn, Taichung 40227, Taiwan
[3] Kaohsiung Med Univ, Biomed Engn & Syst Bioinformat Res Ctr, Kaohsiung, Taiwan
[4] Natl Cheng Kung Univ, Med Device Innovat Ctr, Tainan 70101, Taiwan
[5] Natl Appl Res Labs, Taipei 106, Taiwan
关键词
Neural implant; Flexible neuroprobe; SU-8; Biocompatibility; Tissue reaction; MICROELECTRODE ARRAYS; DRUG-DELIVERY; BRAIN-TISSUE; MICROPROBES; CELL; MICROENVIRONMENT; FABRICATION; SUBSTRATE; STIFFNESS; RESPONSES;
D O I
10.1016/j.sna.2014.06.005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Flexible neuroprobe with better tissue compliance exhibits unique mechanical characteristics in maintaining stability of neural implant in vivo. In this study, a flexible neuroprobe using SU-8 was designed and fabricated for in vitro and in vivo electrical sensing to show the improved tissue compatibility compared to that of the traditional rigid neuroprobe. The validation of neuroprobe was achieved by in vitro mechanical and cytotoxicity tests as well as in vivo neural recording and immunohistological staining. The fabrication process consisted of the creation of a backbone structure using photolithography, photopatterning of evaporated metal, and insulating of the electrode trace. The results of mechanical test of our fabricated SU-8 neuroprobe showed four times of physical stress (18.77 mN) than the insertion force (4.69 mN) to sustain resistance from brain tissue during implantation. The in vitro cytotoxicity assay showed well neuronal survival and proved the sufficient surface biocompatibility of the SU-8 neuroprobe. Further in vivo immunohistological staining showed no obvious glia aggregation around the implantation site indicating suitable biocompatibility compared with that of a rigid neuroprobe. Our in vitro and in vivo studies showed SU-8 neuroprobe possessed enough stress to complete the implantation in brain tissue and remained flexibility to comply micromovement of soft tissue with minor immune responses to achieve in vivo electrophysiological recordings at a signal-noise-ratio of greater than 7. (C) 2014 Elsevier BM. All rights reserved.
引用
收藏
页码:257 / 265
页数:9
相关论文
共 49 条
  • [31] Structural characterization and antioxidant properties of a novel polysaccharide isolated from Jiuzao in vitro and in vivo
    Li, Qing
    Geng, Xiaojie
    Zhu, Lin
    Zheng, Fuping
    Chen, Erbao
    Wang, Guangnan
    Li, Xing
    FOOD RESEARCH INTERNATIONAL, 2022, 162
  • [32] On modeling the in vivo ventricular passive mechanical behavior from in vitro experimental properties in porcine hearts
    Laita, Nicolas
    Rosales, Ricardo M.
    Wu, Ming
    Claus, Piet
    Janssens, Stefan
    Martinez, Miguel Angel
    Doblare, Manuel
    Pena, Estefania
    COMPUTERS & STRUCTURES, 2024, 292
  • [33] Improved dissolution and micromeritic properties of naproxen from spherical agglomerates: preparation, in vitro and in vivo characterization
    Saritha, Damineni
    Bose, Penjuri Subhash Chandra
    Reddy, Poreddy Srikanth
    Madhuri, Grandhi
    Nagaraju, Ravouru
    BRAZILIAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2012, 48 (04) : 667 - 676
  • [34] Chemical characterization and pharmacological properties of polysaccharides from Allium roseum leaves: In vitro and in vivo assays
    Kherroubi, Sara
    Morjen, Maram
    Teka, Nesrine
    Mraihi, Farouk
    Srairi-Abid, Najet
    Le Cerf, Didier
    Marrakchi, Naziha
    Majdoub, Hatem
    Cherif, Jamila Kalthoum
    Jebali, Jed
    Ternane, Riadh
    INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2024, 277
  • [35] In vivo mechanical properties of thoracic aortic aneurysmal wall estimated from in vitro biaxial tensile test
    Fukui, T
    Matsumoto, T
    Tanaka, T
    Ohashi, T
    Kumagai, K
    Akimoto, H
    Tabayashi, K
    Sato, M
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2005, 15 (04) : 295 - 305
  • [37] Fabrication and Characterization of 3D Multi-Electrode Array on Flexible Substrate for In Vivo EMG Recording from Expiratory Muscle of Songbird
    Zia, Muneeb
    Chung, Bryce
    Sober, Samuel J.
    Bakir, Muhannad S.
    2018 IEEE INTERNATIONAL ELECTRON DEVICES MEETING (IEDM), 2018,
  • [38] In vitro and in vivo characterization of the interaction, proinflammatory, immunomodulatory and antigenic properties of capsular polysaccharide from Streptococcus pneumoniae serotype 1
    Kaushal, Nitika
    Kumari, Sujata
    Jhelum, Hina
    Sehgal, Devinder
    INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2020, 143 : 521 - 532
  • [39] Characterization of properties that allow lymphomyeloid cells to be detected in vitro (as LTC-ICML) and their distinction from in vivo repopulating ability.
    Lemieux, ME
    Eaves, CJ
    BLOOD, 1995, 86 (10) : 1671 - 1671
  • [40] Superparamagnetic iron oxide nanoparticles (SPIONs):: From synthesis to in vivo studies -: A summary of the synthesis, characterization, in vitro, and in vivo investigations of SPIONs with particular focus on surface and colloidal properties
    Petri-Fink, Alke
    Hofmann, Heinrich
    IEEE TRANSACTIONS ON NANOBIOSCIENCE, 2007, 6 (04) : 289 - 297