An extended radon chamber model

被引:0
|
作者
ButterweckDempewolf, G
Schuler, C
机构
关键词
D O I
暂无
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The radon chamber model developed at the Paul Scherrer Institut (PSI) is covering differences in size between the different short-lived radon daughters by taking into account nuclide specific deposition and attachment rates. The resulting set of differential equations was analytically solved and the solutions were installed into a computer code. Different modes of operation of a radon chamber (continuous flow, pulse, and feedback) were analysed with the model. The influence of aerosol sampling on the chamber atmosphere was studied. Increasing the diffusion constant of the radon daughter clusters increases both attachment and deposition rates. As a consequence, the unattached concentration of the radon daughter with the shortest half-life, Po-218, is reduced relative to the radon progeny Pb-214 and Bi-214/Po-214, compared with calculations with equal mobilities of the radon progeny. The amount of Po-218 activity deposited on the chamber walls also is increased relative to the other radon progeny if the size of Po-218 is considered to be smaller. Copyright 1996 Elsevier Science Ltd.
引用
收藏
页码:S891 / S898
页数:8
相关论文
共 50 条
  • [41] LARGE VOLUME SCINTILLATION CHAMBER FOR RADON COUNTING
    KRANER, HW
    EVANS, RD
    LEWIS, AR
    SCHROEDER, GL
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1964, 35 (10): : 1259 - &
  • [42] Development of a sparging chamber for field radon analysis
    Richard N. Peterson
    John A. Breier
    Lindsay R. Harmon
    Jamie Brusa
    Patrick R. Hutchins
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 298 : 1347 - 1357
  • [43] Design, construct and test of a calibration radon chamber
    Shweikani, R
    Raja, G
    RADIATION MEASUREMENTS, 2005, 40 (2-6) : 316 - 319
  • [44] Methodology of integrated measurements with radon diffusion chamber
    Cimbaljevic, Z.
    Dragnic, R.
    Radonjic, S.
    Markovic, V. M.
    Stevanovic, N.
    Krstic, D.
    Ruixue, H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1068
  • [45] Simulated equilibrium factor studies in radon chamber
    Chu, Tieh-Chi
    Liu, Ho-Ling
    Applied Radiation and Isotopes, 47 (5-6): : 543 - 550
  • [46] Prototype exposure chamber of radon for animal experiments
    Yamada, Y
    Koizumi, A
    Yonehara, H
    Shimo, M
    Inaba, J
    RADON AND THORON IN THE HUMAN ENVIRONMENT, 1998, : 67 - 72
  • [47] Realization of radioactive equilibrium in the KRISS radon chamber
    Lee, Mo Sung
    Park, Tae Soon
    Lee, Jong Man
    APPLIED RADIATION AND ISOTOPES, 2013, 81 : 226 - 229
  • [48] Methods of kernel parameter calculation for stabilization technology of radon concentration in a closed radon chamber
    Zhang Xiongjie
    Zhang Ye
    Tang Bin
    Wang Renbo
    Qu Jinhui
    APPLIED RADIATION AND ISOTOPES, 2018, 140 : 267 - 271
  • [49] EXTENDED LIFETIME OF A WIRE CHAMBER
    SIPILA, H
    JARVINEN, ML
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 217 (1-2): : 298 - 300
  • [50] Measuring Atmospheric Radon Levels Using a Cloud Chamber
    Geske, Matthew
    Murray-Weston, Crystal
    Lelack, Graham
    PHYSICS TEACHER, 2022, 60 (06): : 498 - 500