Development and validation of a measurement system for continuously monitoring postoperative reservoir levels

被引:1
|
作者
Eichhorn, S. [1 ,2 ]
Reisinger, T. [1 ,2 ]
Boehm, J. [1 ,2 ]
Voss, S. [1 ,2 ]
Doppler, S. [1 ,2 ]
Lange, R. [1 ,2 ,3 ]
Krane, M. [1 ,2 ,3 ]
机构
[1] Tech Univ Munich, German Heart Ctr Munich, Dept Cardiovasc Surg, Lazarettstr 36, D-80636 Munich, Germany
[2] Tech Univ Munich, INSURE Inst Translat Cardiac Surg, Dept Cardiovasc Surg, Lothstr 11, D-80636 Munich, Germany
[3] DZHK German Ctr Cardiovasc Res, Partner Site Munich Heart Alliance, D-80636 Munich, Germany
关键词
Postoperative monitoring; Surgical reservoir; Measurement system; CARDIAC-SURGERY; BLOOD; RISK; REEXPLORATION; TRANSFUSION;
D O I
10.1007/s13246-019-00746-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Following cardiac surgical procedures, multiple drainage systems remain in place inside the patient's chest to prevent the development of pericardial effusion or pneumothorax. Therefore, postoperative bleeding must be diligently observed. Currently, observation of the exudate rate is performed through periodical visual inspection of the reservoir. To improve postoperative monitoring, a measurement system based on load cells was developed to automatically detect bleeding rates. A reservoir retaining bracket was instrumented with a load cell. The signal was digitized by a microcontroller and then processed and displayed on customized software written in LabView. In cases where bleeding rates reach critical levels, the device will automatically sound an alarm. Additionally, the bleeding rate is displayed on the screen with the status of the alarm, as well as the fluid level of the reservoir. These data are all logged to a file. The measurement system has been validated for gain stability and drift, as well as for sensor accuracy, with different in vitro examinations. Additionally, performance of the measurement device was tested in a clinical pilot study on patients recovering from cardiac surgical procedures. The in vitro investigation showed that the monitoring device had excellent gain and drift stability, as well as sensor accuracy, with a resolution of 2.6mL/h for the bleeding rate. During the clinical examination, bleeding rates of all patients were correctly measured. Continuously recording drainage volume using the developed system was comparable to manual measurements performed every 30min by a nurse. Implementation of continuous digital measurements could improve patient safety and reduce the workload of medical professionals working in intensive care units.
引用
收藏
页码:611 / 617
页数:7
相关论文
共 50 条
  • [21] Validation of a new method for measuring and continuously monitoring the efficiency of industrial flares
    Zeng, Yousheng
    Morris, Jon
    Dombrowski, Mark
    JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 2016, 66 (01) : 76 - 86
  • [22] VALIDATION OF ENVIRONMENTAL MONITORING METHODS AND LEVELS
    NOBLE, NH
    JOURNAL OF PARENTERAL SCIENCE AND TECHNOLOGY, 1993, 47 (01): : 26 - 28
  • [23] Development of a Noninvasive Continuous Blood Pressure Measurement and Monitoring System
    Islam, Md. Manirul
    Rafi, Fida Hasan Md.
    Mitul, Abu Farzan
    Ahmad, Mohiuddin
    Rashid, M. A.
    bin Abd Malek, Mohd Fareq
    2012 INTERNATIONAL CONFERENCE ON INFORMATICS, ELECTRONICS & VISION (ICIEV), 2012, : 1085 - 1090
  • [24] Design and Development of a Nonintrusive Pressure Measurement System for Pipeline Monitoring
    Sadeghioon, A. M.
    Walton, Robert
    Chapman, David
    Metje, Nicole
    Anthony, Carl
    Ward, Mike
    JOURNAL OF PIPELINE SYSTEMS ENGINEERING AND PRACTICE, 2014, 5 (03)
  • [25] Development of an innovative measurement system for audible noise monitoring of OHL
    Schichler, U.
    Troppauer, W.
    Fischer, B.
    Heine, T.
    Reich, K.
    Leonhardsberger, M.
    Oberzaucher, O.
    ELEKTROTECHNIK UND INFORMATIONSTECHNIK, 2018, 135 (08): : 556 - 562
  • [26] DETECTION OF BACTEREMIA BY THE CONTINUOUSLY MONITORING BACT/ALERT SYSTEM
    KENNEDY, GT
    BARR, JG
    GOLDSMITH, C
    JOURNAL OF CLINICAL PATHOLOGY, 1995, 48 (10) : 912 - 914
  • [27] Development and validation of a strain-based Structural Health Monitoring system
    Katsikeros, Ch. E.
    Labeas, G. N.
    MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2009, 23 (02) : 372 - 383
  • [28] Development and validation of a coordinate mapping system for lesion monitoring in hidradenitis suppurativa
    Frew, John
    Vekic, Dunja
    Woods, Jane
    Cains, Geoff
    JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2018, 79 (03) : AB106 - AB106
  • [29] OCCUPATIONAL WRIST POSTURAL ASSESSMENT AND MONITORING SYSTEM: DEVELOPMENT AND INITIAL VALIDATION
    Umar, Radin Zaid Radin
    Ling, Chai Fong
    Abdullasim, Nazreen
    Ahmad, Nadiah
    Halim, Isa
    Hamid, Malek
    JOURNAL OF ENGINEERING SCIENCE AND TECHNOLOGY, 2019, 14 (06): : 3421 - 3436
  • [30] Development and Validation of a Pilot Viscosity Monitoring System for Secondary Battery Slurries
    You H.D.
    Choi H.W.
    Lee B.M.
    Kim H.S.
    Hwang W.R.
    Transactions of the Korean Society of Mechanical Engineers, B, 2024, 48 (03) : 203 - 210