摘要
Due to the rising life expectancy, bone diseases (e g osteoporosis, osteoarthritis) and trauma (e.g. fracture) have become an important socio-economic burden Accurate visualization and quantification of the bone micro structure in vivo is seen as an important step to enhance diagnosis and treatment. Micro-computed tomography (microCT) has become the gold standard in three-dimensional (3D) imaging of trabecular bone structure. Yet, usage is limited to ex vivo analyses, hence, it cannot be used to evaluate bone and bone adaptive responses in a patient High-resolution peripheral computed tomography (HR-pQCT) is considered the best technique to measure the bone microarchitecture in vivo By design HR pQCT is limited to scanning extremities, such as the distal radius and distal tibia with a limited field of view and long scanning time (similar to 2a 3 mm for a stack of 0.9 cm). Cone-beam computed tomography (CBCT) is a promising alternative with a much larger field of view. Yet, CBCT is challenged by artefacts that reduce image contrast, such that it is currently being used for qualitative evaluation only. Therefore, the aims of this work were first to enhance image contrast and second to determine the accuracy of high-resolution CBCT for bone microarchitectural assessment. Trapezia of nineteen female arthritic patients were scanned twice ex vivo, once using CBCT (NewTom 5G, Cefla, Verona, Italy) at a nominal voxel size of 75 mu m and once using microCT (SkyScan 1172, Bruker, Kontich, Belgium) at a voxel size of 19 84 mu m. The CBCT-scans were reconstructed following 2 protocols: (1) using the commercial software delivered with the scanner and (2) using in-house developed software After reconstruction and image processing, the images were segmented using adaptive thresholding Bone morphometry parameters including bone volume (BV), total tissue volume (TV), bone volume fraction (BV/TV), bone surface density (BS/ TV), trabecular thickness (Tb.Th), trabecular separation (Tb Sp) and trabecular number (Tb N) were calculated Statistical evaluations were made at a significance level of 5%. Significant correlations were found between the CBCT based bone parameters and the microCT based parameters with R-2 > 0.68 The in house reconstructed software outperformed the commercial software. Smaller bias (overestimation of TbTh decreased from 114.24% to 59.96%) as well as higher correlations were observed for the in-house processed images. Still, a significant overestimation was observed for BV/TV and Tb. Th and an underestimation for Tb.N. We conclude that our CBCT image reconstruction improved image contrast which allowed for an accurate quantification of trabecular bone microarchitecture.
引用
收藏
页码:206 / 214
页数:9
相关论文
共 50 条
  • [21] Forensic imaging of projectiles using cone-beam computed tomography
    von See, Constantin
    Bormann, Kai-Hendrik
    Schumann, Paul
    Goetz, Friedrich
    Gellrich, Nils-Claudius
    Ruecker, Martin
    FORENSIC SCIENCE INTERNATIONAL, 2009, 190 (1-3) : 38 - 41
  • [22] Quantification of skeletal asymmetries in normal adolescents: Cone-beam computed tomography analysis
    Sanders D.A.
    Chandhoke T.K.
    Uribe F.A.
    Rigali P.H.
    Nanda R.
    Progress in Orthodontics, 15 (1) : 1 - 11
  • [23] Evaluation of mastoid pneumatization using cone-beam computed tomography
    Demirel, Oguzhan
    Kaya, Elif
    Ucok, Cemile Ozlem
    ORAL RADIOLOGY, 2014, 30 (01) : 92 - 97
  • [25] Evaluating the mandible with cone-beam computed tomography
    Stratemann, Scott A.
    Huang, John C.
    Maki, Koutaro
    Hatcher, David C.
    Miller, Arthur J.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S58 - S70
  • [26] Simple ROI Cone-Beam Computed Tomography
    Maass, Clemens
    Knaup, Michael
    Sawall, Stefan
    Kachelriess, Marc
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3161 - 3168
  • [27] Operational principles for cone-beam computed tomography
    Hatcher, David C.
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2010, 141 (10): : 3S - 6S
  • [28] Cone-Beam Computed Tomography: Beyond Diagnostics
    Perel, Morton L.
    IMPLANT DENTISTRY, 2015, 24 (04) : 367 - 367
  • [29] Assessment of the Trabecular Bone Microstructure Surrounding Impacted Maxillary Canines Using Fractal Analysis on Cone-Beam Computed Tomography Images
    Sunal Akturk, Ezgi
    Toktas, Ahsen Irem
    Can, Erkay
    Kosen, Ezgi
    Sarica, Irfan
    DIAGNOSTICS, 2024, 14 (19)
  • [30] Comparative analysis of bone density measurements by using multislice spiral and cone-beam computed tomography
    Bal, Cenkhan
    Unsal, Gokce Soganci
    Turkyilmaz, Ilser
    JOURNAL OF DENTAL SCIENCES, 2020, 15 (03) : 388 - 389