Optimization of derivative kernels for exact cone-beam ROI reconstruction in spiral computed tomography

被引:0
|
作者
Lauritsch, G [1 ]
Sourbelle, K
Tam, KC
机构
[1] Seimens Med Solut, D-91050 Erlangen, Germany
[2] Univ Erlangen Nurnberg, IMP, D-91054 Erlangen, Germany
[3] Siemens Corp Res Inc, Princeton, NJ 08540 USA
关键词
cone-beam computed tomography (CT); cone-beam filtered backprojection (FBP); algorithms; filter design; finite differences; linogram; local region-of-interest (ROI) algorithm; numerical derivative; ringing artifacts; spiral CT; three-dimensional (3-D) radon inversion; windowed filter kernels;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A recent comparison study showed that exact filtered backprojection (FBP)-based solutions of the long object problem are numerically stable and that their image quality is mostly affected by filter design. In this paper, the impact of the kernel design on image. quality is demonstrated systematically in the case of the local region-of-interest (ROI) algorithm. First of all, the implementation is improved by the use of the Linogram technique to eliminate interpolations in the filter step and to speed up computation. In the case of the local ROI method, favorable kernels for one-dimensional (1-D) partial derivatives have to be designed. Finite differences are fast to compute but cause ringing artifacts. More sophisticated kernels can be designed in a straightforward manner in Fourier space, identifying the window function most appropriate to a given application. No ringing artifacts are observed for the windowed kernels under investigation. In addition, windowed kernels are superior to the finite differences regarding the spatial resolution versus noise properties. For partial derivatives in directions with data truncation the projection data have to be extrapolated before convolution. Alternatively, very short derivative kernels can be used, generally at a decreased image quality. Kernels of different design and length can be combined.
引用
收藏
页码:728 / 732
页数:5
相关论文
共 50 条
  • [21] Characteristics of noise and resolution on image reconstruction in cone-beam computed tomography
    Lee, Seung-Wan
    Kim, Hee-Joung
    Lee, Chang-Lae
    Cho, Hyo-Min
    Park, Hye-Suk
    Kim, Dae-Hong
    Choi, Yu-No
    Ryu, Hyun-Ju
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [22] Comparison of radiation dose for implant imaging using conventional spiral tomography, computed tomography, and cone-beam computed tomography
    Chau, Anson C. M.
    Fung, Karl
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 107 (04): : 559 - 565
  • [23] Exact Interior Reconstruction with Cone-Beam CT
    Ye, Yangbo
    Yu, Hengyong
    Wang, Ge
    INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2007, 2007
  • [24] Approximate and exact cone-beam reconstruction with standard and non-standard spiral scanning
    Wang, Ge
    Ye, Yangbo
    Yu, Hengyong
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (06): : R1 - R13
  • [25] Approximate and exact cone-beam reconstruction with standard and non-standard spiral scanning
    Wang, Ge
    Ye, Yangbo
    Yu, Hengyong
    DEVELOPMENTS IN X-RAY TOMOGRAPHY V, 2006, 6318
  • [26] Determination of exact reconstruction regions in composite-circling cone-beam tomography
    Ye, Lena
    Yu, Hengyong
    Wang, Ge
    MEDICAL PHYSICS, 2009, 36 (08) : 3448 - 3454
  • [28] Optimization Reconstruction of Spiral Cone-beam CT with Displaced Detector and Increased Pitch
    Zou, Xiao-Bing
    Li, Zong-Jian
    PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON MEDICINE AND BIOPHARMACEUTICALS, 2016, : 427 - 435
  • [29] Weighted FDK Algorithm from Spiral Cone-Beam Computed Tomography with Displaced Detector
    Zou, Xiaobing
    Zeng, Li
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2015, 5 (02) : 290 - 295
  • [30] Evaluating the mandible with cone-beam computed tomography
    Stratemann, Scott A.
    Huang, John C.
    Maki, Koutaro
    Hatcher, David C.
    Miller, Arthur J.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S58 - S70