Comparison of optimal performance at 300 keV of three direct electron detectors for use in low dose electron microscopy

被引:232
|
作者
McMullan, G. [1 ]
Faruqi, A. R. [1 ]
Clare, D. [2 ,3 ]
Henderson, R. [1 ]
机构
[1] MRC, Mol Biol Lab, Cambridge CB2 0QH, England
[2] Univ London, Birkbeck Coll, London WC1E 7HX, England
[3] Univ London, Birkbeck Coll, Inst Struct & Mol Biol, London WC1E 7HX, England
基金
英国医学研究理事会; 英国惠康基金;
关键词
DQE; MTF; CMOS; NOISE TRANSFER; CCD CAMERAS; CRYO-EM; SIGNAL;
D O I
10.1016/j.ultramic.2014.08.002
中图分类号
TH742 [显微镜];
学科分类号
摘要
Low dose electron imaging applications such as electron cryo-microscopy are now benefit:Ling from the improved performance and flexibility of recently introduced electron imaging detectors in which electrons are directly incident on backthinned CMOS sensors. There are currently three commercially available detectors of this type: the Direct Electron DE-20, the FEI Falcon II and the Gatan K2 Summit. These have different characteristics and so it is important to compare their imaging properties carefully with a view to optimise how each is used. Results at 300 keV for both the modulation transfer function (MTF) and the detective quantum efficiency (DQE) are presented. Of these, the DQE is the most important in the study of radiation sensitive samples where detector performance is crucial. We find that all three defectors have a better DQE than film. The K2 Summit has the best DQE at low spatial frequencies but with increasing spatial frequency its DQE falls below that of the Falcon II. (C) 2014 The Authors. Published by Elsevier B.V.
引用
收藏
页码:156 / 163
页数:8
相关论文
共 50 条
  • [1] Direct electron detectors for electron microscopy
    Faruqi, A. R.
    ADVANCES IN IMAGING AND ELECTRON PHYSICS, VOL 145, 2007, 145 : 55 - 93
  • [2] Direct Detectors for Electron Microscopy
    Clough, R. N.
    Moldovan, G.
    Kirkland, A. I.
    ELECTRON MICROSCOPY AND ANALYSIS GROUP CONFERENCE 2013 (EMAG2013), 2014, 522
  • [3] Direct imaging detectors for electron microscopy
    Faruqi, A. R.
    McMullan, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 878 : 180 - 190
  • [4] Low-dose electron energy-loss spectroscopy using electron counting direct detectors
    Maigne, Alan
    Wolf, Matthias
    MICROSCOPY, 2018, 67 : i86 - i97
  • [6] Direct detectors and their applications in electron microscopy for materials science
    Levin, Barnaby D. A.
    JOURNAL OF PHYSICS-MATERIALS, 2021, 4 (04):
  • [7] Effect of electron irradiation dose on the performance of avalanche photodiode electron detectors
    Kawauchi, Taizo
    Wilde, Markus
    Fukutani, Katsuyuki
    Okano, Tatsuo
    Kishimoto, Shunji
    JOURNAL OF APPLIED PHYSICS, 2009, 105 (01)
  • [8] Noise coefficients of backscattered electron detectors for low voltage scanning electron microscopy
    Hejna, J.
    JOURNAL OF MICROSCOPY, 2013, 252 (01) : 35 - 48
  • [9] LOW-DOSE ELECTRON-MICROSCOPY
    MATSUO, T
    HOSOI, J
    ISHIDA, Y
    HARADA, Y
    JOURNAL OF ELECTRON MICROSCOPY, 1980, 29 (03): : 303 - 303
  • [10] Enhanced imaging in low dose electron microscopy using electron counting
    McMullan, G.
    Clark, A. T.
    Turchetta, R.
    Faruqi, A. R.
    ULTRAMICROSCOPY, 2009, 109 (12) : 1411 - 1416