Joint estimation of dynamic PET images and temporal basis functions using fully 4D ML-EM

被引:71
|
作者
Reader, Andrew J.
Sureau, Florent C.
Comtat, Claude
Trebossen, Regine
Buvat, Irene
机构
[1] Univ Manchester, Sch Chem Engn & Analyt Sci, Manchester M60 1QD, Lancs, England
[2] DRM, DSV, CEA, Serv Hosp Frederic Joliot, Orsay, France
[3] Siemens Med Solut, St Denis, France
[4] CHU Pitie Salpetriere, UPMC, INSERM, UMR 678, Paris, France
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2006年 / 51卷 / 21期
关键词
D O I
10.1088/0031-9155/51/21/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A fully 4D joint-estimation approach to reconstruction of temporal sequences of 3D positron emission tomography (PET) images is proposed. The method estimates both a set of temporal basis functions and the corresponding coefficient for each basis function at each spatial location within the image. The joint estimation is performed through a fully 4D version of the maximum likelihood expectation maximization (ML-EM) algorithm in conjunction with two different models of the mean of the Poisson measured data. The first model regards the coefficients of the temporal basis functions as the unknown parameters to be estimated and the second model regards the temporal basis functions themselves as the unknown parameters. The fully 4D methodology is compared to the conventional frame-by-frame independent reconstruction approach (3D ML-EM) for varying levels of both spatial and temporal post-reconstruction smoothing. It is found that using a set of temporally extensive basis functions (estimated from the data by 4D ML-EM) significantly reduces the spatial noise when compared to the independent method for a given level of image resolution. In addition to spatial image quality advantages, for smaller regions of interest (where statistical quality is often limited) the reconstructed time-activity curves show a lower level of bias and a lower level of noise compared to the independent reconstruction approach. Finally, the method is demonstrated on clinical 4D PET data.
引用
收藏
页码:5455 / 5474
页数:20
相关论文
共 50 条
  • [41] Direct 4D parametric imaging for linearized models of reversibly binding PET tracers using generalized AB-EM reconstruction
    Rahmim, Arman
    Zhou, Yun
    Tang, Jing
    Lu, Lijun
    Sossi, Vesna
    Wong, Dean F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (03): : 733 - 755
  • [42] Real-time dynamic display of registered 4D cardiac MR and ultrasound images using a GPU
    Zhang, Q.
    Huang, X.
    Eagleson, R.
    Guiraudon, G.
    Peters, T. M.
    MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [43] Whole-Body dynamic PET: Effect of temporal gaps on FDG Ki quantification from 3D and 4D reconstruction algorithms
    Chalampalakis, Zacharias
    Stute, Simon
    Filipovic, Marina
    Merlin, Thibaut
    Comtat, Claude
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [44] Cardiac-gated parametric images from 82Rb PET from dynamic frames and direct 4D reconstruction
    Germino, Mary
    Carson, Richard E.
    MEDICAL PHYSICS, 2018, 45 (02) : 639 - 654
  • [45] Automatic segmentation of 4D cardiac MR images for extraction of ventricular chambers using a spatio-temporal approach
    Atehortua, Angelica
    Zuluaga, Maria A.
    Ourselin, Sebastien
    Giraldo, Diana
    Romero, Eduardo
    MEDICAL IMAGING 2016: IMAGE PROCESSING, 2016, 9784
  • [46] Automatic Segmentation of Static and Moving Target Volumes Using Respiratory Ungated (3D) and Gated (4D) PET/CT Images
    Park, S.
    Ionascu, T.
    Killoran, J.
    Mamede, M.
    Gerbaudo, V.
    Chin, L.
    Berbeco, R.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [47] Estimation of full-dose 4D CT perfusion images from low-dose images using conditional generative adversarial networks
    Moghari, M. Dashtbani
    Zhou, L.
    Yu, B.
    Moore, K.
    Young, N.
    Fulton, R.
    Kyme, A.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [48] Evaluation of amplitude-based sorting algorithm to reduce lung tumor blurring in PET images using 4D NCAT phantom
    Wang, Jiali
    Byrne, James
    Franquiz, Juan
    McGoron, Anthony
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2007, 87 (02) : 112 - 122
  • [49] Preliminary clinical evaluation of a cardiac motion vector field estimation method from 4D cardiac-gated myocardial perfusion PET images
    Tsui, Benjamin
    Xu, Jingyan
    Lee, Taek-Soo
    Civelek, Ali
    Valenta-Schindler, Ines
    Schindler, Thomas
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [50] Estimating Dynamic Lung Images from High-Dimension Chest Surface Motion Using 4D Statistical Model
    He, Tiancheng
    Xue, Zhong
    Yu, Nam
    Nitsch, Paige L.
    Teh, Bin S.
    Wong, Stephen T.
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2014, PT II, 2014, 8674 : 138 - 145