Ultrasound fusion image error correction using subject-specific liver motion model and automatic image registration

被引:7
|
作者
Yang, Minglei [1 ,3 ]
Ding, Hui [1 ,3 ]
Zhu, Lei [2 ,4 ]
Wang, Guangzhi [1 ,3 ]
机构
[1] Tsinghua Univ, Sch Med, Dept Biomed Engn, Beijing 100084, Peoples R China
[2] Beijing Shen Mindray Med Elect Technol Res Inst C, Beijing 100085, Peoples R China
[3] Tsinghua Univ, Sch Med, Room C249, Beijing 100084, Peoples R China
[4] 3F Bldg5,8 Chuangye Rd, Beijing 100085, Peoples R China
基金
中国国家自然科学基金;
关键词
Ultrasound fusion imaging; Respiration-induced fusion error; Subject-specific motion model; Automatic image registration; Liver;
D O I
10.1016/j.compbiomed.2016.10.008
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: Ultrasound fusion imaging is an emerging tool and benefits a variety of clinical applications, such as image-guided diagnosis and treatment of hepatocellular carcinoma and unresectable liver metastases. However, respiratory liver motion-induced misalignment of multimodal images (i.e., fusion error) compromises the effectiveness and practicability of this method. The purpose of this paper is to develop a subject-specific liver motion model and automatic registration-based method to correct the fusion error. Methods: An online-built subject-specific motion model and automatic image registration method for 2D ultrasound-3D magnetic resonance (MR) images were combined to compensate for the respiratory liver motion. The key steps included: 1) Build a subject-specific liver motion model for current subject online and perform the initial registration of pre-acquired 3D MR and intra-operative ultrasound images; 2) During fusion imaging, compensate for liver motion first using the motion model, and then using an automatic registration method to further correct the respiratory fusion error. Evaluation experiments were conducted on liver phantom and five subjects. Results: In the phantom study, the fusion error (superior-inferior axis) was reduced from 13.90 +/- 2.38 mm to 4.26 +/- 0.78 mm by using the motion model only. The fusion error further decreased to 0.63 +/- 0.53 mm by using the registration method. The registration method also decreased the rotation error from 7.06 +/- 0.21 degrees to 1.18 +/- 0.66. In the clinical study, the fusion error was reduced from 12.90 +/- 9.58 nun to 6.12 +/- 2.90 mm by using the motion model alone. Moreover, the fusion error decreased to 1.96 +/- 0.33 mm by using the registration method. Conclusions: The proposed method can effectively correct the respiration-induced fusion error to improve the fusion image quality. This method can also reduce the error correction dependency on the initial registration of ultrasound and MR images. Overall, the proposed method can improve the clinical practicability of ultrasound fusion imaging.
引用
收藏
页码:99 / 109
页数:11
相关论文
共 50 条
  • [21] Motion Artifact Correction in OCT Volume Scans Using Image Registration
    Kraus, M.
    Mayer, M. A.
    Bock, R.
    Potsaid, B.
    Manjunath, V.
    Duker, J. S.
    Hornegger, J.
    Fujimoto, J. G.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [22] Correction of motion artifact in cardiac optical mapping using image registration
    Rohde, GK
    Dawant, BM
    Lin, SF
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (02) : 338 - 341
  • [23] A spatiotemporal satellite image fusion model with autoregressive error correction (AREC)
    Wang, Jing
    Huang, Bo
    INTERNATIONAL JOURNAL OF REMOTE SENSING, 2018, 39 (20) : 6731 - 6756
  • [24] Experimental validation of a subject-specific finite element model of lumbar spine segment using digital image correlation
    Garavelli, Chiara
    Curreli, Cristina
    Palanca, Marco
    Aldieri, Alessandra
    Cristofolini, Luca
    Viceconti, Marco
    PLOS ONE, 2022, 17 (09):
  • [25] Automatic 3D ultrasound calibration for image guided therapy using intramodality image registration
    Schlosser, Jeffrey
    Kirmizibayrak, Can
    Shamdasani, Vijay
    Metz, Steve
    Hristov, Dimitre
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (21): : 7481 - 7496
  • [26] Image Registration Using Wavelet-Based Motion Model
    Yu-Te Wu
    Takeo Kanade
    Ching-Chung Li
    Jeffrey Cohn
    International Journal of Computer Vision, 2000, 38 : 129 - 152
  • [27] Image registration using wavelet-based motion model
    Wu, YT
    Kanade, T
    Li, CC
    Cohn, J
    INTERNATIONAL JOURNAL OF COMPUTER VISION, 2000, 38 (02) : 129 - 152
  • [28] Motion correction for myocardial T1 mapping using image registration with synthetic image estimation
    Xue, Hui
    Shah, Saurabh
    Greiser, Andreas
    Guetter, Christoph
    Littmann, Arne
    Jolly, Marie-Pierre
    Arai, Andrew E.
    Zuehlsdorff, Sven
    Guehring, Jens
    Kellman, Peter
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (06) : 1644 - 1655
  • [29] Automatic Thyroid Ultrasound Image Classification Using Feature Fusion Network
    Zhao, Xiaohui
    Shen, Xueqin
    Wan, Wenbo
    Lu, Yuanyuan
    Hu, Shidong
    Xiao, Ruoxiu
    Du, Xiaohui
    Li, Junlai
    IEEE ACCESS, 2022, 10 : 27917 - 27924
  • [30] Registration error of the liver CT using deformable image registration of MIM Maestro and Velocity AI
    Nobuyoshi Fukumitsu
    Kazunori Nitta
    Toshiyuki Terunuma
    Toshiyuki Okumura
    Haruko Numajiri
    Yoshiko Oshiro
    Kayoko Ohnishi
    Masashi Mizumoto
    Teruhito Aihara
    Hitoshi Ishikawa
    Koji Tsuboi
    Hideyuki Sakurai
    BMC Medical Imaging, 17