Intravascular ultrasound predictors of angiographic restenosis after sirolimus-eluting stent implantation

被引:215
|
作者
Hong, Myeong-Ki
Mintz, Gary S.
Lee, Cheol Whan
Park, Duk-Woo
Choi, Bong-Ryong
Park, Kyoung-Ha
Kim, Young-Hak
Cheong, Sang-Sig
Song, Jae-Kwan
Kim, Jae-Joong
Park, Seong-Wook
Park, Seung-Jung
机构
[1] Univ Ulsan, Coll Med, Dept Med, Asan Med Ctr, Seoul 138736, South Korea
[2] Cardiovasc Res Fdn, New York, NY USA
[3] Univ Ulsan, Coll Med, Dept Med, Asan Med Ctr, Kangnung, South Korea
关键词
ultrasonics; stent; coronary disease;
D O I
10.1093/eurheartj/ehi882
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims In many countries, drug-eluting stent implantation is the dominant interventional strategy. We evaluated the clinical, angiographic, procedural, and intravascular ultrasound (IVUS) predictors of angiographic restenosis after sirolimus-eluting stent (SES) implantation. Methods and results SES implantation was successfully performed in 550 patients with 670 native coronary lesions. Six-month follow-up angiography was performed in 449 patients (81.6%) with 543 lesions (81.1%). Clinical, angiographic, procedural, and IVUS predictors of restenosis were determined. Using multivariable logistic regression analysis, the only independent predictors of angiographic restenosis were post-procedural final minimum stent area by IVUS [odds ratio (OR) = 0.586, 95% confidence interval (Cl) 0.387-0.888, P = 0.012] and IVUS-measured stent length (OR = 1.029, 95% Cl 1.002-1.056, P = 0.035). Final minimum stent area by IVUS and IVUS-measured stent length that best separated restenosis from non-restenosis were 5.5 mm(2) and 40 mm, respectively. Lesions with final minimum stent area < 5.5 mm(2) and stent length > 40 mm had the highest rate of angiographic restenosis [17.7% (11/62)], P < 0.001 compared with other groups. Conclusion Independent predictors of angiographic restenosis after SES implantation were post-procedural final minimum stent area by IVUS and IVUS-measured stent length. The angiographic restenosis rate was highest in lesions with stent area < 5.5 mm(2) and stent length > 40 mm.
引用
收藏
页码:1305 / 1310
页数:6
相关论文
共 50 条
  • [21] Intravascular ultrasound assessment of stent thrombosis after successful implantation of sirolimus-eluting stents
    Fujii, K
    Carlier, SG
    Mintz, GS
    Yang, YM
    Moussa, S
    Costa, RA
    Costa, JD
    Kimura, M
    Sano, K
    Dangas, G
    Mehran, R
    Lansky, AJ
    Stone, GW
    Moses, JW
    Leon, MB
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (03) : 34A - 34A
  • [22] Angiographic and volumetric intravascular ultrasound comparison between direct sirolimus-eluting stent implantation versus predilation
    Muñoz, JS
    Abizaid, A
    Albertal, M
    Mintz, GS
    Graebin, R
    Feres, F
    Staico, R
    Mattos, LA
    Tanajura, LF
    Abizaid, AS
    Sousa, A
    Sousa, JE
    AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (12): : 1522 - 1525
  • [23] Optimal Strategy of Restenosis with Stent Fracture after Sirolimus-eluting Stent Implantation
    Otsuro, Suguru
    Kadota, Kazushige
    Mitsudo, Kazuaki
    Goto, Tsuyoshi
    Yamamoto, Hiyoyuki
    Fuku, Yasushi
    Hosoki, Shingo
    Tanaka, Hiroyuki
    Habara, Seiji
    Hasegawa, Daiji
    Imai, Masao
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 104 (6A): : 61D - 61D
  • [24] Incidence and predictors of angiographic restenosis after sirolimus-eluting stent for diffuse bare-metal in-stent restenosis.
    Lee, S. W.
    Suh, J.
    Kim, J. H.
    Lee, S. W.
    Park, S. W.
    Park, D. W.
    Hong, M. K.
    Lee, C. W.
    Kim, Y. H.
    Kim, J. J.
    Park, S. J.
    AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (08): : 48F - 48F
  • [25] Stent length as a predictor of restenosis after long sirolimus-eluting stent implantation
    Park, SJ
    Kim, YH
    Lee, CW
    Hong, MK
    Park, SW
    Gwon, HC
    Jang, YS
    Lee, MM
    Koo, BK
    Oh, DJ
    Seung, KB
    Tahk, SJ
    Yoon, J
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (03) : 64A - 64A
  • [26] Angiographic findings of stent thrombosis cases after sirolimus-eluting stent implantation
    Kozuma, K.
    Kimura, T.
    Honda, Y.
    Aizawa, T.
    Mitsudo, K.
    Miyazaki, S.
    Yamaguchi, T.
    Isshiki, T.
    EUROPEAN HEART JOURNAL, 2010, 31 : 660 - 660
  • [27] Incidence and predictors of restenosis after sirolimus-eluting stent implantation in high-risk patients
    Berenguer, A
    Mainar, V
    Bordes, P
    Valencia, J
    Gómez, S
    Lozano, T
    AMERICAN HEART JOURNAL, 2005, 150 (03) : 536 - 542
  • [28] Repeated Sirolimus-Eluting Stent Implantation to Treat Sirolimus-Eluting Stent and Bare-Metal Stent Restenosis
    Nishihira, Kensaku
    Shibata, Yoshisato
    Ishikawa, Tetsunori
    Nomura, Katsumasa
    Nakama, Tatsuya
    Mine, Daigo
    Inoue, Yohei
    Ashikaga, Keiichi
    Kuriyama, Nehiro
    Matsuyama, Akihiko
    Imamura, Takuroh
    Asada, Yujiro
    Kitamura, Kazuo
    CIRCULATION JOURNAL, 2010, 74 (11) : 2329 - 2333
  • [29] Predictors and Prognosis of Stent Fracture After Sirolimus-Eluting Stent Implantation
    Ino, Yasushi
    Toyoda, Yasuaki
    Tanaka, Atsushi
    Ishii, Shuuta
    Kusuyama, Yoshio
    Kubo, Takashi
    Takarada, Shigeho
    Kitabata, Hironori
    Tanimoto, Takashi
    Mizukoshi, Masato
    Imanishi, Toshio
    Akasaka, Takashi
    CIRCULATION JOURNAL, 2009, 73 (11) : 2036 - 2041
  • [30] Sirolimus-eluting stent implantation for unselected in-stent restenosis
    Cho, KH
    Jeong, YS
    Kim, SM
    Kim, DI
    Kim, DS
    Kim, KB
    Kim, KS
    Hur, SH
    Kim, YJ
    Park, JS
    Shim, BS
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (8A): : 44A - 44A