Automatic anatomy recognition in whole-body PET/CT images

被引:18
|
作者
Wang, Huiqian [1 ,2 ]
Udupa, Jayaram K. [2 ]
Odhner, Dewey [2 ]
Tong, Yubing [2 ]
Zhao, Liming [2 ,3 ]
Torigian, Drew A. [2 ]
机构
[1] Chongqing Univ, Coll Optoelect Engn, Chongqing 400044, Peoples R China
[2] Univ Penn, Dept Radiol, Med Image Proc Grp, Philadelphia, PA 19104 USA
[3] Chongqing Univ Posts & Telecommun, Res Ctr Intelligent Syst & Robot, Chongqing 400065, Peoples R China
关键词
image segmentation; PET/CT imaging; object recognition; texture; fuzzy models; POSITRON-EMISSION-TOMOGRAPHY; 3D CT IMAGES; THRESHOLD SEGMENTATION; REGRESSION FORESTS; NECK-CANCER; LOCALIZATION; HEAD; CLASSIFICATION; DELINEATION; SYSTEM;
D O I
10.1118/1.4939127
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Whole-body positron emission tomography/computed tomography (PET/CT) has become a standard method of imaging patients with various disease conditions, especially cancer. Body-wide accurate quantification of disease burden in PET/CT images is important for characterizing lesions, staging disease, prognosticating patient outcome, planning treatment, and evaluating disease response to therapeutic interventions. However, body-wide anatomy recognition in PET/CT is a critical first step for accurately and automatically quantifying disease body-wide, body-region-wise, and organwise. This latter process, however, has remained a challenge due to the lower quality of the anatomic information portrayed in the CT component of this imaging modality and the paucity of anatomic details in the PET component. In this paper, the authors demonstrate the adaptation of a recently developed automatic anatomy recognition (AAR) methodology [Udupa et al., "Body-wide hierarchical fuzzy modeling, recognition, and delineation of anatomy in medical images," Med. Image Anal. 18, 752-771 (2014)] to PET/CT images. Their goal was to test what level of object localization accuracy can be achieved on PET/CT compared to that achieved on diagnostic CT images. Methods: The authors advance the AAR approach in this work in three fronts: (i) from body-region-wise treatment in the work of Udupa et al. to whole body; (ii) from the use of image intensity in optimal object recognition in the work of Udupa et al. to intensity plus object-specific texture properties, and (iii) from the intramodality model-building-recognition strategy to the intermodality approach. The whole-body approach allows consideration of relationships among objects in different body regions, which was previously not possible. Consideration of object texture allows generalizing the previous optimal threshold-based fuzzy model recognition method from intensity images to any derived fuzzy membership image, and in the process, to bring performance to the level achieved on diagnostic CT and MR images in body-region-wise approaches. The intermodality approach fosters the use of already existing fuzzy models, previously created from diagnostic CT images, on PET/CT and other derived images, thus truly separating the modality-independent object assembly anatomy from modality-specific tissue property portrayal in the image. Results: Key ways of combining the above three basic ideas lead them to 15 different strategies for recognizing objects in PET/CT images. Utilizing 50 diagnostic CT image data sets from the thoracic and abdominal body regions and 16 whole-body PET/CT image data sets, the authors compare the recognition performance among these 15 strategies on 18 objects from the thorax, abdomen, and pelvis in object localization error and size estimation error. Particularly on texture membership images, object localization is within three voxels on whole-body low-dose CT images and 2 voxels on body-region-wise low-dose images of known true locations. Surprisingly, even on direct body-region-wise PET images, localization error within 3 voxels seems possible. Conclusions: The previous body-region-wise approach can be extended to whole-body torso with similar object localization performance. Combined use of image texture and intensity property yields the best object localization accuracy. In both body-region-wise and whole-body approaches, recognition performance on low-dose CT images reaches levels previously achieved on diagnostic CT images. The best object recognition strategy varies among objects; the proposed framework however allows employing a strategy that is optimal for each object. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:613 / 629
页数:17
相关论文
共 50 条
  • [21] Synthetic CT generation from non-attenuation corrected PET images for whole-body PET imaging
    Dong, Xue
    Wang, Tonghe
    Lei, Yang
    Higgins, Kristin
    Liu, Tian
    Curran, Walter J.
    Mao, Hui
    Nye, Jonathon A.
    Yang, Xiaofeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (21):
  • [22] CT Radiation Dose Optimization in Whole-Body PET/CT Examination
    Tonkopi, E.
    Ross, A.
    MacDonald, A.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [23] Body region localization in whole-body low-dose CT images of PET/CT scans using virtual landmarks
    Bai, Peirui
    Udupa, Jayaram K.
    Tong, Yubing
    Xie, ShiPeng
    Torigian, Drew A.
    MEDICAL PHYSICS, 2019, 46 (03) : 1286 - 1299
  • [24] Is Automatic Tumor Segmentation on Whole-Body 18F-FDG PET Images a Clinical Reality?
    Sundar, Lalith Kumar Shiyam
    Beyer, Thomas
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65 (07) : 995 - 997
  • [25] Comparison Between True Whole-Body and Limited Whole-Body FDG PET/CT Studies in Oncology Patients
    Sebro, R.
    Aparici, Mari C.
    Pampaloni, Hernandez M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [26] Effect of patient arm motion in whole-body PET/CT
    Lodge, Martin
    Mhlanga, Joyce
    Wahl, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [27] Role of Limited Whole-Body PET/CT in Pediatric Lymphoma
    Sammer, Marla B. K.
    Shulkin, Barry L.
    Alessio, Adam
    Parisi, Marguerite T.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (05) : 1047 - 1055
  • [28] Comparison of whole-body MRI with whole-body dual-modality PET/CT for tumor staging in oncology
    Antoch, G
    Vogt, FM
    Lauenstein, TC
    Mueller, SP
    Ruehm, SG
    Debatin, JF
    RADIOLOGY, 2002, 225 : 659 - 659
  • [29] Effect of Patient Arm Motion in Whole-Body PET/CT
    Lodge, Martin A.
    Mhlanga, Joyce C.
    Cho, Steve Y.
    Wahl, Richard L.
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (12) : 1891 - 1897
  • [30] Whole-body tumor burden in PET/CT expert review
    Santos, Dihego F.
    Takahashi, Maria Emilia
    Camacho, Mariana
    Lopes de Lima, Mariana da Cunha
    Amorim, Barbara Juarez
    Rohren, Eric M.
    Etchebehere, Elba
    CLINICAL AND TRANSLATIONAL IMAGING, 2023, 11 (01) : 5 - 22