Performance Evaluation of a PET/MR Detector Based on the ClearPEM Technology

被引:0
|
作者
Neves, Jorge A. [1 ,2 ]
Bugalho, Ricardo [1 ]
Ortigao, Catarina [1 ]
Zorraquino, Carlos [1 ]
Silva, Rui [1 ]
Silva, Jose C. [1 ]
Magill, Arthur W. [2 ]
Tavernier, Stefaan [3 ]
Gruetter, Rolf [2 ]
Varela, Joao [1 ]
机构
[1] LIP Lab Instrumentat & Expt Particle Phys, Lisbon, Portugal
[2] Ecole Polytech Fed Lausanne, Lausanne, Switzerland
[3] Vrije Univ Brussel, Brussels, Belgium
关键词
Electromagnetic Interference (EMI); Magnetic Resonance Imaging (MRI); Positron Emission Tomography (PET); Radio-frequency (RF); RF Coil; Ultra-High Field (UHF);
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
We present performance results for a PET detector prototype in simultaneous operation within a 7 T MR system. The detectors are based on ClearPEM technology, a dedicated PET scanner for high-resolution breast imaging, and suitable for integration into combined PET/MR systems at ultra-high magnetic fields. To demonstrate simultaneous PET/MR detection with minimal electromagnetic interference between the two systems, the PET detectors and front-end electronics were carefully shielded and a new dedicated RF volume coil was constructed to mitigate electromagnetic coupling with the shielded electronics. Noise measurements at the level of the PET front-end electronics and the MR RF receiver have shown the mutual interference of RF power and digital noise, and demonstrated the excellent shielding effectiveness achieved. Simultaneous data acquisitions with a Ge-68 source and a saline phantom have shown no distortions or artifacts on MR images caused by the detector modules, and no degradation on the detector performance and energy resolution. Therefore we conclude that both systems can be operated simultaneously with negligible or no electromagnetic interference.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Evaluation of Performance and Stability of an MR Compatible PET Detector
    Shams, Ehsan
    Thiessen, Jonathan D.
    Bishop, Daryl
    Kozlowski, Piotr
    Retiere, Fabrice
    Khan, Muhammad Salman
    Sossi, Vesna
    Stortz, Greg
    Thompson, Christopher J.
    Goertzen, Andrew L.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [2] Evaluation of PET performance and MR compatibility of a preclinical PET/MR insert with digital silicon photomultiplier technology
    Patrick Hallen
    David Schug
    Jakob Wehner
    Bjorn Weissler
    Pierre Gebhardt
    Benjamin Goldschmidt
    Andre Salomon
    Peter Duppenbecker
    Fabian Kiessling
    Volkmar Schultz
    EJNMMI Physics, 2 (Suppl 1)
  • [3] Monte Carlo based study on influence of SiPM specification on detector performance for PET/MR
    Pulko, Jozef
    Hohberg, Melanie
    Velroyen, Astrid
    Ziegler, Sibylle
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [4] Design and Detector Performance of the PET Component of the TRIMAGE PET/MR/EEG Scanner
    Belcari, Nicola
    Bisogni, Maria Giuseppina
    Camarlinghi, Niccolo
    Carra, Pietro
    Cerello, Piergiorgio
    Morrocchi, Matteo
    Patera, Alessandra
    Sportelli, Giancarlo
    Del Guerra, Alberto
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (03) : 292 - 301
  • [5] High performance detector head for PET and PET/MR with continuous crystals and SiPMs
    Llosa, G.
    Barrillon, P.
    Barrio, J.
    Bisogni, M. G.
    Cabello, J.
    Del Guerra, A.
    Etxebeste, A.
    Gillam, J. E.
    Lacasta, C.
    Oliver, J. F.
    Rafecas, M.
    Solaz, C.
    Stankova, V.
    de La Taille, C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 702 : 3 - 5
  • [6] APD-based PET detector for simultaneous PET/MR imaging
    Grazioso, Ronald
    Zhang, Nan
    Corbeil, James
    Schmand, Matthias
    Ladebeck, Ralf
    Vester, Markus
    Schnur, Guenter
    Renz, Wolfgang
    Fischer, Hubertus
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02): : 301 - 305
  • [7] MR Performance Comparison of a PET/MR System Before and After SiPM-Based Time-of-Flight PET Detector Insertion
    Khalighi, Mohammad Mehdi
    Delso, Gaspar
    Maramraju, Sri Harsha
    Deller, Timothy W.
    Levin, Craig S.
    Glover, Gary H.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (05) : 2419 - 2423
  • [8] PET Performance Evaluation of an MR-Compatible PET Insert
    Wu, Yibao
    Catana, Ciprian
    Farrell, Richard
    Dokhale, Purushottam A.
    Shah, Kanai S.
    Qi, Jinyi
    Cherry, Simon R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 574 - 580
  • [9] Evaluation of silicon photomultipliers: A promising new detector for MR compatible PET
    McElroy, David P.
    Saveliev, Valeri
    Reznik, Alla
    Rowlands, John A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 106 - 109
  • [10] ToF Performance Evaluation of PET Modules With Digital Silicon Photomultiplier Technology During MR Operation
    Schug, David
    Wehner, Jakob
    Dueppenbecker, Peter Michael
    Weissler, Bjoern
    Gebhardt, Pierre
    Goldschmidt, Benjamin
    Solf, Torsten
    Kiessling, Fabian
    Schulz, Volkmar
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (03) : 658 - 663