Setup error and motion during deep inspiration breath-hold breast radiotherapy measured with continuous portal imaging

被引:42
|
作者
Lutz, Christina Maria [1 ]
Poulsen, Per Rugaard [2 ,3 ]
Fledelius, Walther [3 ]
Offersen, Birgitte Vrou [2 ,3 ]
Thomsen, Mette Skovhus [1 ]
机构
[1] Aarhus Univ Hosp, Dept Med Phys, DK-8000 Aarhus, Denmark
[2] Aarhus Univ, Inst Clin Med, Aarhus, Denmark
[3] Aarhus Univ Hosp, Dept Oncol, DK-8000 Aarhus, Denmark
关键词
CANCER PATIENTS; RANDOMIZED-TRIALS; RADIATION-THERAPY; HEART-DISEASE; WOMEN; RISK; REPRODUCIBILITY; IRRADIATION; SURVIVORS; GUIDANCE;
D O I
10.3109/0284186X.2015.1045625
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. The position and residual motion of the chest wall of breast cancer patients during treatment in deep inspiration breath-hold (DIBH) were investigated. Material and methods. The study included 58 left-sided breast cancer patients treated with DIBH three-dimensional (3D) conformal radiotherapy in 15 or 25 fractions. The DIBH levels were monitored using an external marker block placed on the chest, either shifted 5 cm to the right at the level of the xiphoid process (Group 1, 27 consecutive patients) or placed medially on the inferior part of the sternum (Group 2, 31 consecutive patients). At every third treatment fraction, continuous portal images were acquired. The time-resolved chest wall position during treatment was compared with the planned position to determine the inter-fraction setup errors and the intra-fraction motion of the chest wall. Results. The DIBH compliance was 95% during both recruitment periods. A tendency of smaller inter-fraction setup errors and intra-fraction motion was observed for group 2 (medial marker block position). However, apart from a significantly reduced inter-field random shift (sigma = 1.7 mm vs. sigma = 0.9 mm, p = 0.005), no statistically significant differences between the groups were found. In a combined analysis, the group mean inter-fraction setup error was M = - 0.1 mm, with random and systematic errors of sigma = 1.7 mm and sigma = 1.4 mm. The group mean inter-field shift was M = 0.0 (sigma = 1.3 mm and sigma = 1.1 mm) and the group mean standard deviation of the intra-field motion was 0.5 mm. The absolute setup error had a maximum of 16.3 mm, exceeding 5 mm in 2.2% of the imaged fields. Conclusion. Compared to free breathing treatments, the primary benefit of the DIBH technique was the separation of the heart from the target rather than more accurate targeting. Despite a small gating window, occasional large errors in the chest wall position were observed for some patients, illustrating limitations of the external marker block as surrogate in a broad patient population.
引用
收藏
页码:193 / 200
页数:8
相关论文
共 50 条
  • [21] Selection of patients with left breast cancer for Deep-Inspiration Breath-Hold Radiotherapy Technique
    Czeremszynska, B.
    Drozda, S.
    Gorzynski, M.
    Kepka, L.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S555 - S555
  • [22] Setup margins based on the inter- and intrafractional setup error of left-sided breast cancer radiotherapy using deep inspiration breath-hold technique (DIBH) and surface guided radiotherapy (SGRT)
    Rudat, Volker
    Shi, Yanyan
    Zhao, Ruping
    Yu, Wei
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (06):
  • [23] Evaluating intra-fractional tumor motion in lung stereotactic radiotherapy with deep inspiration breath-hold
    Fu, Weihua
    Zhang, Yongqian
    Mehta, Kiran
    Chen, Alex
    Musunuru, Hima Bindu
    Pucci, Pietro
    Kubis, Jason
    Huq, M. Saiful
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (08):
  • [24] Interfraction heart position changes during voluntary moderate deep inspiration breath-hold radiotherapy
    Claassen-Janssen, F.
    Van Haaren, P.
    Van de Sande, I.
    Boersma, L.
    Van der Sangen, M.
    Hurkmans, C.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S21 - S21
  • [25] Prolonging deep inspiration breath-hold time to 3 min during radiotherapy, a simple solution
    Vakaet, Vincent
    Van Hulle, Hans
    Schoepen, Max
    Van Caelenberg, Els
    Van Greveling, Annick
    Holvoet, Jeroen
    Monten, Chris
    De Baerdemaeker, Luc
    De Neve, Wilfried
    Coppens, Marc
    Veldeman, Liv
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2021, 28 : 10 - 16
  • [26] Review of deep inspiration breath-hold techniques for the treatment of breast cancer
    Latty, Drew
    Stuart, Kirsty E.
    Wang, Wei
    Ahern, Verity
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (01) : 74 - 81
  • [27] Appropriate margin for planning target volume for breast radiotherapy during deep inspiration breath-hold by variance component analysis
    Yuka Ono
    Michio Yoshimura
    Tomohiro Ono
    Takahiro Fujimoto
    Yuki Miyabe
    Yukinori Matsuo
    Takashi Mizowaki
    Radiation Oncology, 16
  • [28] Appropriate margin for planning target volume for breast radiotherapy during deep inspiration breath-hold by variance component analysis
    Ono, Yuka
    Yoshimura, Michio
    Ono, Tomohiro
    Fujimoto, Takahiro
    Miyabe, Yuki
    Matsuo, Yukinori
    Mizowaki, Takashi
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [29] Fast 3D Imaging via Deep learning for Deep Inspiration Breath-hold Lung Radiotherapy
    Lei, Yang
    Tian, Zhen
    Wang, Tonghe
    Axente, Marian
    Roper, Justin
    Higgins, Kristin
    Bradley, Jeffrey D.
    Liu, Tian
    Yang, Xiaofeng
    MEDICAL IMAGING 2022: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2022, 12034
  • [30] Deep Inspiration Breath-Hold versus free breathing radiotherapy in mediastinal lymphoma
    Iversen, A.
    Mortensen, H. R.
    Rose, Hanne Krogh
    Hansen, Jolanta
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S696 - S696