Body Birdcage Coil using Cylindrical Rods

被引:0
|
作者
Sidhique, A. [1 ]
Rao, P. H. [1 ]
Harsh, Rajesh [2 ]
机构
[1] SAMEER Ctr Electromagnet, Chennai, Tamil Nadu, India
[2] SAMEER, Mumbai, Maharashtra, India
来源
2017 IEEE APPLIED ELECTROMAGNETICS CONFERENCE (AEMC) | 2017年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A high-pass Body Birdcage Coil using cylindrical rods designed for a 1.5 T MRI system is presented. The operating resonant mode is investigated for optimized magnetic field homogeneity. The shielded Body Birdcage Coil is energized in the circularly polarized feed configuration. The input ports of the Body Birdcage Coil are matched (S-11 = S-22 = -19.7 dB) and isolated (S-21 = S-12 = -26.4 dB) at 64.3 MHz. The same analysis has been extended for human model loaded Body Birdcage Coil. A slight shift in the resonant frequency from 64.3 MHz to 63.5 MHz is observed for unloaded and human model loaded Birdcage Coil. The B-1(+) field distribution of human model loaded and unloaded Body Birdcage Coils are presented for dominant mode.
引用
收藏
页数:2
相关论文
共 50 条
  • [41] Three-dimensional high-inductance birdcage coil for NQR applications
    Peshkovsky, A. S.
    Cerioni, L.
    Osan, T. M.
    Avdievich, N. I.
    Pusiol, D. J.
    SOLID STATE NUCLEAR MAGNETIC RESONANCE, 2006, 30 (02) : 75 - 80
  • [42] An Extended Theory and Method for Double-Tuned Conformal Birdcage Coil in MRI
    Xie, Junyao
    Guan, Jinglong
    Yao, Kaiwen
    Liu, Min
    You, Xueqiu
    Sun, Huijun
    Chen, Zhong
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2024, 73 : 1 - 11
  • [43] A novel method for coil efficiency estimation: Validation with a 13C birdcage
    Giovannetti, Giulio
    Frijia, Francesca
    Hartwig, Valentina
    Menichetti, Luca
    De Marchi, Daniele
    Positano, Vincenzo
    Landini, Luigi
    Lombardi, Massimo
    Santarelli, Maria Filomena
    Ardenkjaer-Larsen, Jan Henrik
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2012, 41B (04) : 139 - 143
  • [44] Potential advantage of higher-order modes of birdcage coil for parallel imaging
    Wang, Chunsheng
    Qu, Peng
    Shen, Gary X.
    JOURNAL OF MAGNETIC RESONANCE, 2006, 182 (01) : 160 - 167
  • [45] Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T
    Valkovic, Ladislav
    Dragonu, Iulius
    Almujayyaz, Salam
    Batzakis, Alex
    Young, Liam A. J.
    Purvis, Lucian A. B.
    Clarke, William T.
    Wichmann, Tobias
    Lanz, Titus
    Neubauer, Stefan
    Robson, Matthew D.
    Klomp, Dennis W. J.
    Rodgers, Christopher T.
    PLOS ONE, 2017, 12 (10):
  • [46] The calculation of the cylindrical coil spring
    Bergstraber, M
    ZEITSCHRIFT DES VEREINES DEUTSCHER INGENIEURE, 1933, 77 : 198 - 198
  • [47] Magnetic field of a cylindrical coil
    Labinac, Velimir
    Erceg, Natasa
    Kotnik-Karuza, Dubravka
    AMERICAN JOURNAL OF PHYSICS, 2006, 74 (07) : 621 - 627
  • [48] The electromagnetic fields of a 64 MHz quadrature driven birdcage coil in ASTM phantom
    Kozlov, Mikhail
    Angelone, L. M.
    Kainz, W.
    2017 47TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2017, : 715 - 718
  • [49] SAR evaluation of pregnant woman models in 64 MHz MRI birdcage coil
    Shamsi, S.
    Wu, D. G.
    Chen, J.
    Liu, R.
    Kainz, W.
    2006 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM DIGEST, VOLS 1-5, 2006, : 225 - +
  • [50] Influence of reference tube location on the measured sodium concentrations in calf muscles using a birdcage coil at 3T
    Baron, Paul
    Potze, Jan Hendrik
    Sijens, Paul E.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 624 - 632