Disinfection and recontamination of rigid endoscopes. Improved safety using an immersion quiver system

被引:6
|
作者
Rohrmeier, C. [1 ]
Strutz, J. [1 ]
Schneider-Brachert, W. [2 ]
机构
[1] Univ Regensburg, Klin Hals Nasen Ohren Heilkunde, D-93053 Regensburg, Germany
[2] Univ Klinikum Regensburg, Inst Klin Mikrobiol & Hyg, Regensburg, Germany
关键词
Diagnostic techniques; respiratory system; Otorhinolaryngology; Infection control; Equipment contamination; Patient safety; FLEXIBLE FIBEROPTIC LARYNGOSCOPES; IN-VITRO CONTAMINATION; STAPHYLOCOCCUS-AUREUS; TRANSMISSION; RECOMMENDATIONS; GUIDELINES; INFECTION; GERMANY; VIRUS; RISK;
D O I
10.1007/s00106-014-2873-x
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Background. In Otorhinolaryngology, rigid endoscopes are used daily at a high frequency. There is no consensus for reprocessing these medical instruments. Often immersion disinfection procedures are used. The present study examined the possible risk of recontamination by this disinfection method and investigated the possibility of avoiding this risk by using a new immersion quiver system. Methods. Using coloured markers, a possible contact of the endoscope with the top edges of quivers of different diameters during endoscope removal was tested for. In addition, it was evaluated whether Staphylococcus aureus transfer is possible via this route. The same methodology was applied to a new immersion quiver system. Results. Whenever removing the rigid endoscopes from the conventional quiver, these touched the top of the quiver, regardless of its diameter. A transfer of Staphylococcus aureus from the quiver to the endoscope via this route could be detected in five out of eight attempts. During endoscope removal from the new immersion quiver system, no contact of the endoscope with the outer quiver occurred in 20 passes. In none of eight trials was a transfer of Staphylococcus aureus from previously contaminated immersion quivers to the endoscope shown; all immersion quivers were sterile after disinfection. Discussion. After conventional immersion disinfection, recontamination of rigid endoscopes by a contaminated quiver edge is possible. An immersion quiver system can resolve this risk of recontamination easily, by decontaminating not only the endoscope, but also the immersion quiver (inner quiver) itself in the disinfectant solution.
引用
收藏
页码:746 / 751
页数:6
相关论文
共 36 条
  • [31] An improved approach for association rule mining using a multi-criteria decision support system: a case study in road safety
    Ait-Mlouk, Addi
    Gharnati, Fatima
    Agouti, Tarik
    EUROPEAN TRANSPORT RESEARCH REVIEW, 2017, 9 (03)
  • [32] Improved Antishock Air-Gap Control Algorithm with Acceleration Feedforward Control for High-Numerical Aperture Near-Field Storage System Using Solid Immersion Lens
    Kim, Jung-Gon
    Shin, Won-Ho
    Hwang, Hyun-Woo
    Jeong, Jun
    Park, Kyoung-Su
    Park, No-Cheol
    Yang, Hyunseok
    Park, Young-Pil
    Park, Jin Moo
    Son, Do Hyeon
    Seo, Jeong Kyo
    Choi, In Ho
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2010, 49 (08)
  • [33] A Novel Vehicle Open Door Safety System Based on Cyclist Detection Using Fisheye Camera and Improved Deep Convolutional Generative Adversarial Nets
    Zhu, Miankuan
    Han, Lei
    Liang, Fujian
    Xi, Chaoxing
    Wu, Lei
    Zhang, Zutao
    2019 30TH IEEE INTELLIGENT VEHICLES SYMPOSIUM (IV19), 2019, : 2195 - 2201
  • [34] PATIENT SAFETY MAY BE IMPROVED USING A DOSE CALCULATING-AND-ROUNDING ELECTRONIC CHEMOTHERAPY ORDER ENTRY AND VERIFICATION SYSTEM WITH CLINICAL DECISION SUPPORT SYSTEM (CDSS) FEATURES IN THE PEDIATRIC SETTING
    Gilheeney, Stephen
    Ackerman, Jill
    De Luca, Jennifer
    Robinson, Donna
    Muller, Raymond
    Sklarin, Nancy
    NEURO-ONCOLOGY, 2016, 18 : 137 - 138
  • [35] Improved post-marketing safety surveillance of quadrivalent inactivated influenza vaccine in Mexico using a computerized, SMS-based follow-up system
    Betancourt-Cravioto, Miguel
    Cervantes-Powell, Patricia
    Tapia-Conyer, Roberto
    Ledlie, Shaleesa
    Gandhi-Banga, Sonja
    HUMAN VACCINES & IMMUNOTHERAPEUTICS, 2022, 18 (01)
  • [36] An Efficient Automatic Face Mask Detection System for Human Safety Based on Deep Learning using Novel YOLOv3 in Comparison of YOLO with Improved Accuracy
    Nithin, A.
    Jaisharma, K.
    Suresh, V.
    2022 14TH INTERNATIONAL CONFERENCE ON MATHEMATICS, ACTUARIAL SCIENCE, COMPUTER SCIENCE AND STATISTICS (MACS), 2022,