A case of bad timing: Inappropriate implantable cardioverter defibrillator therapy due to a critically placed premature ventricular contraction

被引:5
|
作者
Healy, E [1 ]
Ngarmukos, T [1 ]
Rosenthal, L [1 ]
机构
[1] UMass Mem Med Ctr, Dept Med, Div Cardiovasc Med, Sect Electrophysiol & Pacing, Worcester, MA 01581 USA
来源
关键词
ICD; inappropriate therapy; PVC;
D O I
10.1046/j.1460-9592.2002.01403.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 58-year-old man with an ischemic cardiomyopathy underwent ICD implantation after having an electrophysiological study positive for inducible, sustained monomorphic VT. Approximately 1 week later he experienced a shock. Inspection of stored electrograms revealed that a run of nonsustained VT had initiated his ICD to charge. The episode spontaneously terminated, yet the shock was still delivered during sinus rhythm. Further inspection revealed that a critically timed PVC during the "second look" interval resulted in delivered therapy. Thus, inappropriate ICD therapy can result from causes other than SVT, lead malfunction, or over sensing.
引用
收藏
页码:1403 / 1405
页数:3
相关论文
共 50 条
  • [1] Inappropriate implantable cardioverter-defibrillator therapy due to the detection of premature ventricular complexes
    Goldberger, JJ
    Horvath, G
    Challapalli, R
    Kadish, AH
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (05): : 825 - 828
  • [2] Inappropriate therapy due to triple counting of the ventricular electrogram in a patient with implantable cardioverter-defibrillator
    Alasti, M
    Haghjoo, M
    Alizadeh, A
    Fazelifar, AF
    Sadr-Ameli, MA
    HEART RHYTHM, 2005, 2 (11) : 1253 - 1255
  • [3] A Case of Exhaustion of Implantable Cardioverter Defibrillator Therapies Due To Inappropriate Programming
    Fragakis, Nikolaos K.
    Antoniadis, Antonios P.
    Geleris, Paraschos
    HELLENIC JOURNAL OF CARDIOLOGY, 2011, 52 (04) : 354 - 356
  • [4] Tachycardia Induction Due to Inappropriate Implantable Cardioverter Defibrillator Therapy: What Is the Mechanism?
    Tzeis, Stylianos
    Andrikopoulos, George
    Rassias, Ioannis
    Theodorakis, George
    HELLENIC JOURNAL OF CARDIOLOGY, 2010, 51 (06) : 558 - 562
  • [5] Inappropriate detection and therapy from an implantable cardioverter defibrillator due to atrial undersensing
    Paraskevaidis, S
    Polymeropoulos, K
    Vassilikos, V
    Louridas, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (09): : 1002 - 1004
  • [6] Inappropriate implantable cardiac defibrillator shock due to couplet premature ventricular contractions
    Wongcharoen, Wanwarang
    Benjanuwattra, Thanawat
    Pongprot, Yupada
    Rachakhom, Cheeranun
    Phrommintikul, Arintaya
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (05): : 609 - 612
  • [7] A case of inappropriate implantable cardioverter defibrillator therapy induced by T-wave oversensing due to hyperkalemia
    Aoki, Kosuke
    Okajima, Katsunori
    Kiuchi, Kunihiko
    Yokoi, Kiminobu
    Teranishi, Jin
    Shimane, Akira
    JOURNAL OF ARRHYTHMIA, 2015, 31 (06) : 395 - 397
  • [8] Decrease in amplitude of intracardiac ventricular electrogram and inappropriate therapy in patients with an implantable cardioverter defibrillator
    Watanabe, Hiroshi
    Chinushi, Masaomi
    Izumi, Daisuke
    Sato, Akinori
    Okada, Shinsuke
    Okamura, Kazuki
    Komura, Satoru
    Hosaka, Yukio
    Furushima, Hiroshi
    Washizuka, Takashi
    Aizawa, Yoshifusa
    INTERNATIONAL HEART JOURNAL, 2006, 47 (03) : 363 - 370
  • [9] Inappropriate shock delivery due to ventricular double detection with a biventricular pacing implantable cardioverter defibrillator
    Schreieck, J
    Zrenner, B
    Kolb, C
    Ndrepepa, G
    Schmitt, C
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (07): : 1154 - 1157
  • [10] Cardiac thromboembolic stroke due to inappropriate implantable cardioverter defibrillator shock
    Zizek, D.
    Antolic, B.
    Jan, M.
    Notar, L.
    EUROPEAN JOURNAL OF HEART FAILURE, 2018, 20 : 207 - 207