A case of bad timing: Inappropriate implantable cardioverter defibrillator therapy due to a critically placed premature ventricular contraction

被引:5
|
作者
Healy, E [1 ]
Ngarmukos, T [1 ]
Rosenthal, L [1 ]
机构
[1] UMass Mem Med Ctr, Dept Med, Div Cardiovasc Med, Sect Electrophysiol & Pacing, Worcester, MA 01581 USA
来源
关键词
ICD; inappropriate therapy; PVC;
D O I
10.1046/j.1460-9592.2002.01403.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 58-year-old man with an ischemic cardiomyopathy underwent ICD implantation after having an electrophysiological study positive for inducible, sustained monomorphic VT. Approximately 1 week later he experienced a shock. Inspection of stored electrograms revealed that a run of nonsustained VT had initiated his ICD to charge. The episode spontaneously terminated, yet the shock was still delivered during sinus rhythm. Further inspection revealed that a critically timed PVC during the "second look" interval resulted in delivered therapy. Thus, inappropriate ICD therapy can result from causes other than SVT, lead malfunction, or over sensing.
引用
收藏
页码:1403 / 1405
页数:3
相关论文
共 50 条
  • [21] Inappropriate subcutaneous implantable cardioverter-defibrillator shock due to electromagnetic interference
    Kohli, Utkarsh
    Hassan, Sohail
    EUROPACE, 2020, 22 (06): : 971 - 971
  • [22] Implantable Cardioverter Defibrillator—An Unusual Case of Inappropriate Discharge During Showering
    Athanasios G. Manolis
    Apostolos G. Katsivas
    Charis V. Vassilopoulos
    Nikolaos E. Louvros
    Journal of Interventional Cardiac Electrophysiology, 2000, 4 : 265 - 268
  • [23] Implantable cardioverter defibrillator - An unusual case of inappropriate discharge during showering
    Manolis, AG
    Katsivas, AG
    Vassilopoulos, CV
    Louvros, NE
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2000, 4 (01) : 265 - 268
  • [24] Inappropriate subcutaneous implantable cardioverter defibrillator shock due to incomplete sealing of the seal plug: a case report
    Ajimi, Tsuneki
    Nozoe, Masatsugu
    Suematsu, Nobuhiro
    Kubota, Toru
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2022, 6 (09)
  • [25] Sensing defect due to polarization in a ventricular implantable cardioverter-defibrillator
    Blommaert, D
    Mucumbitsi, J
    De Roy, L
    HEARTWEB, 1999, 4 (10): : U15 - U18
  • [26] Appropriate Therapy But Not Inappropriate Shocks Predict Survival in Implantable Cardioverter Defibrillator Patients
    Dichtl, Wolfgang
    Wolber, Thomas
    Paoli, Ursula
    Bruellmann, Simon
    Stuehlinger, Markus
    Berger, Thomas
    Spuller, Karin
    Strasak, Alexander
    Pachinger, Otmar
    Haegeli, Laurent M.
    Duru, Firat
    Hintringer, Florian
    CLINICAL CARDIOLOGY, 2011, 34 (07) : 433 - 436
  • [27] Risk factors for the first and second inappropriate implantable cardioverter-defibrillator therapy
    Nishii, Nobuhiro
    Noda, Takashi
    Nitta, Takashi
    Aizawa, Yoshifusa
    Ohe, Tohru
    Kurita, Takashi
    IJC HEART & VASCULATURE, 2021, 34
  • [28] Inappropriate implantable cardioverter-defibrillator therapy during exercise: What is the mechanism?
    Vollmann, Dirk
    Luethje, Lars
    Zabel, Markus
    HEART RHYTHM, 2009, 6 (05) : 718 - 719
  • [29] Relationship Between Oral Amiodarone and Inappropriate Therapy From an Implantable Cardioverter Defibrillator
    Nagai, Takayuki
    Satomi, Kazuhiro
    Noda, Takashi
    Okamura, Hideo
    Yamada, Yuko
    Shimizu, Wataru
    Suyama, Kazuhiro
    Aihara, Naohiko
    Kamakura, Shiro
    Kurita, Takashi
    CIRCULATION JOURNAL, 2010, 74 (07) : 1302 - 1307
  • [30] Inappropriate Implantable Cardioverter- Defibrillator Therapy During Total Hip Arthroplasty
    DeLeon, Alexander M.
    Lu, Rachael
    Chen, Liting
    Asher, Yogen
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (05)