Inappropriate implantable cardioverter-defibrillator therapy due to the detection of premature ventricular complexes

被引:5
|
作者
Goldberger, JJ [1 ]
Horvath, G [1 ]
Challapalli, R [1 ]
Kadish, AH [1 ]
机构
[1] Northwestern Univ, Dept Med, Div Cardiol, Chicago, IL 60611 USA
来源
关键词
implantable cardioverter defibrillator; inappropriate shocks;
D O I
10.1111/j.1540-8159.1999.tb00553.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inappropriate shocks delivered by implantable cardioverter-defibrillators may occur in 25%-40% of patients treated with these devices. This article describes a rare cause for inappropriate shocks. Two patients received inappropriate shocks due to the presence of premature ventricular complexes during attempted reconfirmation of ventricular fibrillation. Knowledge of device algorithms for detection and reconfirmation of ventricular fibrillation, designed to have a high degree of sensitivity and therefore lower specificity, allows for reprogramming to avoid fur ther inappropriate shocks.
引用
收藏
页码:825 / 828
页数:4
相关论文
共 50 条
  • [1] Inappropriate therapy due to triple counting of the ventricular electrogram in a patient with implantable cardioverter-defibrillator
    Alasti, M
    Haghjoo, M
    Alizadeh, A
    Fazelifar, AF
    Sadr-Ameli, MA
    HEART RHYTHM, 2005, 2 (11) : 1253 - 1255
  • [2] Impact of Carvedilol and Metoprolol on Inappropriate Implantable Cardioverter-Defibrillator Therapy
    Ruwald, Martin H.
    Abu-Zeitone, Abeer
    Jons, Christian
    Ruwald, Anne-Christine
    McNitt, Scott
    Kutyifa, Valentina
    Zareba, Wojciech
    Moss, Arthur J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (15) : 1343 - 1350
  • [3] Predictors and Proarrhythmic Consequences of Inappropriate Implantable Cardioverter-Defibrillator Therapy
    Tenma, Taro
    Yokoshiki, Hisashi
    Mizukami, Kazuya
    Mitsuyama, Hirofumi
    Watanabe, Masaya
    Sasaki, Ryo
    Maeno, Motoki
    Matsui, Yoshiro
    Tsutsui, Hiroyuki
    CIRCULATION JOURNAL, 2015, 79 (09) : 1920 - U242
  • [4] A case of bad timing: Inappropriate implantable cardioverter defibrillator therapy due to a critically placed premature ventricular contraction
    Healy, E
    Ngarmukos, T
    Rosenthal, L
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (09): : 1403 - 1405
  • [5] Inappropriate implantable cardioverter-defibrillator shocks due to temporary epicardial pacing
    Gelissen, Marloes
    Debie, Luuk
    van Opstal, Jurren
    EUROPACE, 2010, 12 (11): : 1648 - 1648
  • [6] Inappropriate subcutaneous implantable cardioverter-defibrillator shock due to electromagnetic interference
    Kohli, Utkarsh
    Hassan, Sohail
    EUROPACE, 2020, 22 (06): : 971 - 971
  • [7] PREDICTORS OF INAPPROPRIATE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR DISCHARGES
    Catanzaro, John N.
    Makaryus, Amgad
    Makaryus, John
    Sison, Cristina
    Vavasis, Christos
    Donaldson, David
    Altman, Erik
    Beldner, Stuart
    Jadonath, Ram
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E172 - E172
  • [8] Implantable Cardioverter-Defibrillator Shock: Appropriate or Inappropriate?
    Nault, Michael A.
    McIntyre, William F.
    Simpson, Christopher S.
    Redfearn, Damian P.
    Abdollah, Hoshiar
    Brennan, F. James
    Baranchuk, Adrian
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2010, 15 (02) : 181 - 183
  • [9] Inappropriate detection and therapy from an implantable cardioverter defibrillator due to atrial undersensing
    Paraskevaidis, S
    Polymeropoulos, K
    Vassilikos, V
    Louridas, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (09): : 1002 - 1004
  • [10] Sensing defect due to polarization in a ventricular implantable cardioverter-defibrillator
    Blommaert, D
    Mucumbitsi, J
    De Roy, L
    HEARTWEB, 1999, 4 (10): : U15 - U18