Inappropriate implantable cardioverter-defibrillator therapy due to the detection of premature ventricular complexes

被引:5
|
作者
Goldberger, JJ [1 ]
Horvath, G [1 ]
Challapalli, R [1 ]
Kadish, AH [1 ]
机构
[1] Northwestern Univ, Dept Med, Div Cardiol, Chicago, IL 60611 USA
来源
关键词
implantable cardioverter defibrillator; inappropriate shocks;
D O I
10.1111/j.1540-8159.1999.tb00553.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inappropriate shocks delivered by implantable cardioverter-defibrillators may occur in 25%-40% of patients treated with these devices. This article describes a rare cause for inappropriate shocks. Two patients received inappropriate shocks due to the presence of premature ventricular complexes during attempted reconfirmation of ventricular fibrillation. Knowledge of device algorithms for detection and reconfirmation of ventricular fibrillation, designed to have a high degree of sensitivity and therefore lower specificity, allows for reprogramming to avoid fur ther inappropriate shocks.
引用
收藏
页码:825 / 828
页数:4
相关论文
共 50 条
  • [21] Inappropriate implantable cardioverter-defibrillator shocks in MADIT II
    Daubert, James P.
    Zareba, Wojciech
    Cannom, David S.
    McNitt, Scott
    Rosero, Spencer Z.
    Wang, Paul
    Schuger, Claudio
    Steinberg, Jonathan S.
    Higgins, Steven L.
    Wilber, David J.
    Klein, Helmut
    Andrews, Mark L.
    Hall, W. Jackson
    Moss, Arthur J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (14) : 1357 - 1365
  • [22] Pacemaker/implantable cardioverter-defibrillator problem: Right ventricular lead missensing in a biventricular implantable cardioverter-defibrillator
    Otten, Richard F.
    Foreman, Lynne D.
    Groh, William J.
    HEART RHYTHM, 2008, 5 (01) : 158 - 159
  • [23] Methods of minimizing inappropriate implantable cardioverter-defibrillator shocks
    Schaer B.
    Osswald S.
    Current Cardiology Reports, 2000, 2 (4) : 346 - 352
  • [24] Inappropriate implantable cardioverter-defibrillator shock in unusual circumstances
    Suran, Maria-Claudia-Berenice
    Margulescu, Andrei-Dumitru
    Siliste, Calin
    Vinereanu, Dragos
    ACTA CARDIOLOGICA, 2016, 71 (03) : 369 - 370
  • [25] ADVANCES IN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR THERAPY
    RAITT, MH
    BARDY, GH
    CURRENT OPINION IN CARDIOLOGY, 1994, 9 (01) : 23 - 29
  • [26] Inappropriate shock delivery due to ventricular double detection with a biventricular pacing implantable cardioverter defibrillator
    Schreieck, J
    Zrenner, B
    Kolb, C
    Ndrepepa, G
    Schmitt, C
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (07): : 1154 - 1157
  • [27] Inappropriate implantable cardioverter-defibrillator therapy: Many causes and many possible solutions
    Zareba, Wojciech
    CARDIOLOGY JOURNAL, 2009, 16 (05) : 391 - 393
  • [28] Risk prediction of inappropriate implantable cardioverter-defibrillator therapy using machine learning
    Tateishi, R.
    Suzuki, M.
    Shimizu, M.
    Shimada, H.
    Tsunoda, T.
    Miyazaki, H.
    Misu, Y.
    Yamakami, Y.
    Yamaguchi, M.
    Kato, N.
    Isshiki, A.
    Kimura, S.
    Fujii, H.
    Nishizaki, M.
    Sasano, T.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [29] Risk prediction of inappropriate implantable cardioverter-defibrillator therapy using machine learning
    Tateishi, Ryo
    Suzuki, Makoto
    Shimizu, Masato
    Shimada, Hiroshi
    Tsunoda, Takahiro
    Miyazaki, Hiroko
    Misu, Yoshiki
    Yamakami, Yosuke
    Yamaguchi, Masao
    Kato, Nobutaka
    Isshiki, Ami
    Kimura, Shigeki
    Fujii, Hiroyuki
    Nishizaki, Mitsuhiro
    Sasano, Tetsuo
    SCIENTIFIC REPORTS, 2023, 13 (01):
  • [30] Inappropriate Implantable Cardioverter-Defibrillator Therapy During Surgery: An Important and Preventable Complication
    Rozner, Marc A.
    Kahl, Edward A.
    Schulman, Peter M.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2017, 31 (03) : 1037 - 1041