Rapid and direct analysis of statins in human plasma by column-switching liquid chromatography with restricted-access material

被引:24
|
作者
Fagundes, Vinicius Freire [1 ]
Leite, Camila Prado [1 ]
Pianetti, Gerson Antonio [1 ]
Fernandes, Christian [1 ]
机构
[1] Univ Fed Minas Gerais, Fac Farm, Dept Prod Farmaceut, BR-31270901 Belo Horizonte, MG, Brazil
关键词
Statins; Column-switching; Restricted access material; Online sample preparation; TANDEM MASS-SPECTROMETRY; LC-MS/MS; SIMVASTATIN; QUANTIFICATION; PRAVASTATIN; LOVASTATIN; HPLC; ACID; ROSUVASTATIN; PITAVASTATIN;
D O I
10.1016/j.jchromb.2013.12.002
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study presents the development of a column-switching liquid chromatographic method with direct injection of human plasma for simultaneous determination of four statins (lovastatin, pravastatin, rosuvastatin and simvastatin), the main class of drugs used in the treatment of hyperlipidemia. By using a C18 (30 mm x 4.6 mm, 15 mu m) a lab made bovine serum albumin restricted access material (RAM) column was prepared and compared with a commercial alquil-diol silica RAM column (C18, 25 mm x 4.0 mm, 25 mu m) in terms of their protein exclusion capacity and micromolecules retention. Foreflush and backflush modes were compared for both RAM columns to the number of theoretical plates, asymmetry, resolution and chromatographic run time. The developed method was validated in the range from 125 to 876 ng mL(-1) for lovastatin, rosuvastatin and simvastatin, and from 500 to 2000 ng mL(-1) for pravastatin, presenting selectivity, precision and accuracy intra and inter-run. Total analysis time (sample preparation and chromatographic separation) was only 16 min when the backflush mode was employed in the column-switching system. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:8 / 16
页数:9
相关论文
共 50 条
  • [21] DETERMINATION OF CEFODIZIME IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH COLUMN-SWITCHING
    BOMPADRE, S
    FERRANTE, L
    LEONE, L
    DEMARTINIS, M
    GINALDI, L
    QUAGLINO, D
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1995, 18 (14): : 2895 - 2909
  • [22] DETERMINATION OF ALFUZOSIN IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH COLUMN-SWITCHING
    CARLUCCI, G
    DIGIUSEPPE, E
    MAZZEO, P
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1994, 17 (18): : 3989 - 3997
  • [23] Advances in restricted-access media for direct injection analysis of biological fluids in high performance liquid chromatography
    Song, YX
    Liu, MC
    Zhu, PL
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 27 (02) : 230 - 236
  • [24] Influence of mobile phase conditions on the clean-up effect of restricted-access media precolumns for plasma samples injected in a column-switching system
    Z. Yu
    D. Westerlund
    Chromatographia, 1997, 44 : 589 - 594
  • [25] Capillary column switching restricted-access media-liquid chromatography-electrospray ionization-tandem mass Spectrometry system for simultaneous and direct analysis of drugs in biofluids
    Santos-Neto, Alvaro J.
    Markides, Karin E.
    Sjoberg, Per J. R.
    Bergquist, Jonas
    Lancas, Fernando M.
    ANALYTICAL CHEMISTRY, 2007, 79 (16) : 6359 - 6367
  • [26] Sensitive column-switching high-performance liquid chromatography method for determination of propiverine in human plasma
    Ban, EM
    Maeng, JE
    Woo, JS
    Kim, CK
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 831 (1-2): : 230 - 235
  • [27] Direct Determination of Benzamides in Serum by Column-Switching High-Performance Liquid Chromatography
    Ryoko Chiba
    Ayako Ogasawara
    Teppei Kubo
    Hiroyuki Yamazaki
    Masuo Umino
    Yoichi Ishizuka
    Analytical Sciences, 2003, 19 : 785 - 789
  • [28] Direct determination of benzamides in serum by column-switching high-performance liquid chromatography
    Chiba, R
    Ogasawara, A
    Kubo, T
    Yamazaki, H
    Umino, M
    Ishizuka, Y
    ANALYTICAL SCIENCES, 2003, 19 (05) : 785 - 789
  • [29] Column-switching high-performance liquid chromatographic analysis of azasetron in rat plasma by direct injection
    Um, S. Y.
    Jung, S. H.
    Jung, S. J.
    Kim, O. H.
    Choi, S. O.
    Kim, J. I.
    Chung, M. W.
    Oh, H. Y.
    DRUG METABOLISM REVIEWS, 2006, 38 : 146 - 146
  • [30] DIRECT INJECTION-ANALYSIS OF KETOPROFEN ENANTIOMERS IN PLASMA USING COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY SYSTEM
    YAGI, M
    SHIBUKAWA, A
    NAKAGAWA, T
    CHEMICAL & PHARMACEUTICAL BULLETIN, 1990, 38 (09) : 2513 - 2517