Performance evaluation of the white cell precursor channel on the Sysmex XN hematology analyzer in HIV specimens

被引:7
|
作者
Schapkaitz, Elise [1 ]
Khoza, Phoyisile [2 ]
机构
[1] Univ Witwatersrand, Med Sch, Dept Mol Med & Haematol, Johannesburg, South Africa
[2] Natl Hlth Lab Serv, Dept Mol Med & Haematol, Johannesburg, South Africa
关键词
human immunodeficiency virus; morphology flags; Sysmex XN; white blood cell differentials; XE-5000; WPC;
D O I
10.1002/jcla.22335
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Automated hematology analyzers generate white blood cell (WBC) suspect flags to indicate smears requiring manual review. In a human immunodeficiency virus (HIV) patient population with a high frequency of abnormal specimens, the usefulness of these flags depends on their sensitivity and specificity. The white cell precursor (WPC) channel on the Sysmex XN-9000 analyzer (Sysmex Corporation, Kobe, Japan) has recently been introduced in order to distinguish immature cells such as blasts and abnormal lymphocytes. Methods: The efficiency of the flags generated by WPC reflex testing was compared to the white cell differential (WDF) on the Sysmex XN-9000 in 400 routine HIV samples. Results: A flag was generated by the WDF channel in 346 (86.50%) samples. Addition of the WPC to the WDF resulted in a 20.99% reduction in the smear review rate. WPC analysis correctly identified the initial WDF flag in 59 (60.82%) samples with 'blasts?' or 'abnormal lympho?' flags. WPC showed an excellent sensitivity of 100%. However, the specificity remained poor. On investigation, samples with chronic infection/ inflammation (35.00%) and lymphopenia (63.25%) were associated with a high false positive rate of 76.32% and 55.56%, respectively. WPC analysis removed 45 (51.72%) coexisting 'atypical lympho?' false positive WDF flags. The 'atypical lympho?' flag showed an increased sensitivity and specificity of 88.46% and 74.67%, respectively. Conclusion: White cell precursor reflex testing showed improved flagging efficiency for the 'atypical lympho?' flag, which was a good indicator of reactive lymphocytes during chronic infection/inflammation. Nevertheless, the false positive rate for 'abnormal lympho/blasts?' was high in the presence of lymphopenia and infection.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] AUTOMATED IMMATURE GRANULOCYTE COUNTS ON THE NEW SYSMEX XN AUTOMATED HEMATOLOGY ANALYZER
    Rosenthal, Nancy
    Connell, Barbara
    Brown, Bonnie
    Kruger, Julie
    Capper, Mary
    Blaine, Kimberly
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 130 - 130
  • [42] Evaluation of the Performance of the Sysmex XT-2000i Hematology Analyzer With Whole Blood Specimens Stored at Room Temperature
    Hill, Valerie L.
    Simpson, Virginia Z.
    Higgins, Jeanette M.
    Hu, Zonghui
    Stevens, Randy A.
    Metcalf, Julia A.
    Baseler, Michael
    LABMEDICINE, 2009, 40 (12): : 709 - 718
  • [43] AN EVALUATION OF THE SYSMEX NE-8000 HEMATOLOGY ANALYZER
    HALLAWELL, R
    OMALLEY, C
    HUSSEIN, S
    DAUER, RJ
    TANTI, M
    WOOTTON, AM
    MCGRATH, KM
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1991, 96 (05) : 594 - 601
  • [44] EVALUATION OF A NEW METHOD FOR THE ENUMERATION OF NUCLEATED RED BLOOD CELLS ON THE NEW SYSMEX XN™ AUTOMATED HEMATOLOGY ANALYZER
    Rosenthal, Nancy
    Brown, Bonnie
    Connell, Barbara
    Kruger, Julie
    Capper, Mary
    Blaine, Kimberly
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 131 - 131
  • [45] DISCRIMINATION OF PLATELETS AND FRAGMENTED ERYTHROCYTES IN THE PLT-F CHANNEL OF XN-SERIES SYSMEX AUTOMATED HEMATOLOGY ANALYZER
    Wada, Atsushi
    Takagi, Yuri
    Kawauchi, Sawako
    Kono, Mari
    Morikawa, Takashi
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 91 - 91
  • [46] Spurious low WBC count in the WNR channel of Sysmex XN-9000 hematology analyzer in a case with leukocyte aggregation
    Dai, Qingkai
    Peng, Luyun
    Shi, Rui
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (07) : E131 - E133
  • [47] Clinical value of smear review of flagged samples analyzed with the Sysmex XN hematology analyzer
    Winther-Larsen, Anne
    Vestergaard, Else Marie
    Abildgaard, Anders
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025, 63 (03) : 636 - 644
  • [48] Performance evaluation of platelet counting of Abbott Alinity hq and Sysmex XN-9000 automated hematology analyzer compared with international reference method
    Kim, Hyun-Young
    Bang, Sung-Hwan
    Cho, Duck
    Kim, Hee-Jin
    Kim, Sun-Hee
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (03) : 387 - 394
  • [49] What is your diagnosis? Additional cluster on the white blood cell differential scattergram using the Sysmex XN-V hematology analyzer in the blood of a dog
    Chenal, Theo
    Guerlin, Maud
    Jurrus, Maxime
    Tante, Rachel
    Granat, Fanny
    Trumel, Catherine
    VETERINARY CLINICAL PATHOLOGY, 2024, 53 : 122 - 124
  • [50] EVALUATION OF THE AUTOMATED HEMATOLOGY ANALYZER SYSMEX NE-8000
    DEVREESE, K
    DELOGI, E
    FRANCART, C
    HEYNDRICKX, B
    PHILIPPE, J
    LEROUXROELS, G
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1991, 29 (05): : 339 - 345