Magnetic field effects on particle beams and their implications for dose calculation in MR-guided particle therapy

被引:50
|
作者
Fuchs, Hermann [1 ,2 ]
Moser, Philipp [1 ,2 ,3 ]
Groeschl, Martin [3 ]
Georg, Dietmar [1 ,2 ,4 ]
机构
[1] Med Univ Vienna AKH, Dept Radiat Oncol, Vienna, Austria
[2] Med Univ Vienna, Christian Doppler Lab Med Radiat Res Radiat Oncol, Vienna, Austria
[3] Vienna Univ Technol, Inst Appl Phys, Vienna, Austria
[4] Med Univ Vienna AKH, Ctr Comprehens Canc, Vienna, Austria
关键词
carbon ion; characterization; ion beam therapy; magnetic field; MR; proton; PROTON THERAPY; FEASIBILITY; DEFLECTION; SYSTEMS;
D O I
10.1002/mp.12105
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate and model effects of magnetic fields on proton and carbon ion beams for dose calculation. Methods: In a first step, Monte Carlo simulations using Gate 7.1/Geant4.10.0.p03 were performed for proton and carbon ion beams in magnetic fields ranging from 0 to 3 T. Initial particle energies ranged from 60 to 250 MeV (protons) and 120 to 400 MeV/u (carbon ions), respectively. The resulting dose distributions were analyzed focusing on beam deflection, dose deformation, as well as the impact of material heterogeneities. In a second step, a numerical algorithm was developed to calculate the lateral beam position. Using the Runge-Kutta method, an iterative solution of the relativistic Lorentz equation, corrected for the changing particle energy during penetration, was performed. For comparison, a gamma-index analysis was utilized, using a criteria of 2%/2 mm of the local maximum. Results: A tilt in the dose distribution within the Bragg peak area was observed, leading to non-negligible dose distribution changes. The magnitude was found to depend on the magnetic field strength as well as on the initial beam energy. Comparison of the 3 T dose distribution with non-B field (nominal) dose distributions, resulted in a gamma(mean) (mean value of the gamma distribution) of 0.6, with 14.4% of the values above 1 and gamma(1)% (1% of all points have an equal or higher gamma value) of 1.8. The presented numerical algorithm calculated the lateral beam offset with maximum errors of less than 2% with calculation times of less than 5 mu s. The impact of tissue interfaces on the proton dose distributions was found to be less than 2% for a dose voxel size of 1 x 1 x 1 mm(3). Conclusion: Non-negligible dose deformations at the Bragg peak area were identified for high initial energies and strong magnetic fields. A fast numerical algorithm based on the solution of the energy-corrected relativistic Lorentz equation was able to describe the beam path, taking into account the particle energy, magnetic field, and material. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:1149 / 1156
页数:8
相关论文
共 50 条
  • [31] Dose calculation accuracy of using tailored synthetic CT for MR-guided online adaptive radiotherapy
    Cusumano, D.
    Placidi, L.
    Teodoli, S.
    Boldrini, L.
    Greco, F.
    Longo, S.
    Cellini, F.
    Dinapoli, N.
    Valentini, V.
    De Spirito, M.
    Azario, L.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1101 - S1102
  • [32] Monitor unit range check as an alternative for secondary dose calculation in online MR-guided RT
    Schuring, D.
    Kierkels, R.
    Hilgers, G.
    Potters, I.
    den Hartogh, M.
    Jeene, P.
    Surmann, K.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1477 - S1478
  • [33] Oncology Scan-Improvements in Dose Calculation, Deformable Registration, and MR-Guided Radiation Delivery
    Klein, Eric E.
    Das, Indra J.
    Dong, Lei
    Rivard, Mark J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (03): : 395 - 397
  • [34] Integration of an Independent Monitor Unit Check for High-Magnetic-Field MR-Guided Radiation Therapy System
    Yang, Jie
    Zhang, Pengpeng
    Tyagi, Neelam
    Scripes, Paola Godoy
    Subashi, Ergys
    Liang, Jiayi
    Lovelock, Dale
    Mechalakos, James
    Li, Anyi
    Lim, Seng B.
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [35] Impact of Magnetic Field on Dose Distribution in MR-Guided Radiotherapy of Head and Neck Cancer (vol 10, 1739, 2020)
    Xia, Wenlong
    Zhang, Ke
    Li, Minghui
    Tian, Yuan
    Men, Kuo
    Wang, Jingbo
    Yi, Junlin
    Li, Yexiong
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2020, 10
  • [36] CALCULATION OF THE FLUCTUATION FIELD OF A FINE PARTICLE MAGNETIC SYSTEM
    COVERDALE, GN
    CHANTRELL, RW
    OGRADY, K
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1990, 83 (1-3) : 442 - 444
  • [37] Fast calculation methods for the magnetic field of particle lattices
    Royo-Silvestre, Isaac
    Gandia, David
    Beato-Lopez, J. J.
    Garaio, Eneko
    Gomez-Polo, Cristina
    JOURNAL OF APPLIED PHYSICS, 2025, 137 (06)
  • [38] Dose accumulation for personalized stereotactic MR-guided adaptive radiation therapy in prostate cancer
    Bohoudi, Omar
    Bruynzeel, Anna M. E.
    Tetar, Shyama
    Slotman, Ben J.
    Palacios, Miguel A.
    Lagerwaard, Frank J.
    RADIOTHERAPY AND ONCOLOGY, 2021, 157 : 197 - 202
  • [39] Phantoms in particle therapy to verify Monte Carlo dose calculation
    Wohlfahrt, P.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S363 - S364
  • [40] Propagation of particle beams in solar coronal magnetic field structures
    Aurass, H
    PLANETARY RADIO EMISSIONS IV, 1997, : 407 - 407