Magnetic field effects on particle beams and their implications for dose calculation in MR-guided particle therapy

被引:50
|
作者
Fuchs, Hermann [1 ,2 ]
Moser, Philipp [1 ,2 ,3 ]
Groeschl, Martin [3 ]
Georg, Dietmar [1 ,2 ,4 ]
机构
[1] Med Univ Vienna AKH, Dept Radiat Oncol, Vienna, Austria
[2] Med Univ Vienna, Christian Doppler Lab Med Radiat Res Radiat Oncol, Vienna, Austria
[3] Vienna Univ Technol, Inst Appl Phys, Vienna, Austria
[4] Med Univ Vienna AKH, Ctr Comprehens Canc, Vienna, Austria
关键词
carbon ion; characterization; ion beam therapy; magnetic field; MR; proton; PROTON THERAPY; FEASIBILITY; DEFLECTION; SYSTEMS;
D O I
10.1002/mp.12105
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate and model effects of magnetic fields on proton and carbon ion beams for dose calculation. Methods: In a first step, Monte Carlo simulations using Gate 7.1/Geant4.10.0.p03 were performed for proton and carbon ion beams in magnetic fields ranging from 0 to 3 T. Initial particle energies ranged from 60 to 250 MeV (protons) and 120 to 400 MeV/u (carbon ions), respectively. The resulting dose distributions were analyzed focusing on beam deflection, dose deformation, as well as the impact of material heterogeneities. In a second step, a numerical algorithm was developed to calculate the lateral beam position. Using the Runge-Kutta method, an iterative solution of the relativistic Lorentz equation, corrected for the changing particle energy during penetration, was performed. For comparison, a gamma-index analysis was utilized, using a criteria of 2%/2 mm of the local maximum. Results: A tilt in the dose distribution within the Bragg peak area was observed, leading to non-negligible dose distribution changes. The magnitude was found to depend on the magnetic field strength as well as on the initial beam energy. Comparison of the 3 T dose distribution with non-B field (nominal) dose distributions, resulted in a gamma(mean) (mean value of the gamma distribution) of 0.6, with 14.4% of the values above 1 and gamma(1)% (1% of all points have an equal or higher gamma value) of 1.8. The presented numerical algorithm calculated the lateral beam offset with maximum errors of less than 2% with calculation times of less than 5 mu s. The impact of tissue interfaces on the proton dose distributions was found to be less than 2% for a dose voxel size of 1 x 1 x 1 mm(3). Conclusion: Non-negligible dose deformations at the Bragg peak area were identified for high initial energies and strong magnetic fields. A fast numerical algorithm based on the solution of the energy-corrected relativistic Lorentz equation was able to describe the beam path, taking into account the particle energy, magnetic field, and material. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:1149 / 1156
页数:8
相关论文
共 50 条
  • [21] A Fast Magnetic Field Mapping Method for An MR-Guided Radiotherapy Environment
    Stanescu, T.
    Constantin, D.
    Dayarian, I.
    Jaffray, D.
    MEDICAL PHYSICS, 2017, 44 (06) : 3142 - 3142
  • [22] MR-guided proton therapy: Impact of magnetic fields on the detector response
    Fuchs, Hermann
    Padilla-Cabal, Fatima
    Zimmermann, Lukas
    Palmans, Hugo
    Georg, Dietmar
    MEDICAL PHYSICS, 2021, 48 (05) : 2572 - 2579
  • [23] Cross Dose Engine Transformation Based Fast Dose Calculation for Online MR-Guided Adaptive Radiotherapy
    Song, T.
    Zhou, L.
    Li, Y.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [24] Needle localization in MR-guided biopsy and aspiration: Effects of field strength, sequence design, and magnetic field orientation
    Lewin, JS
    Duerk, JL
    Jain, VR
    Petersilge, CA
    Chao, CP
    Haaga, JR
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (06) : 1337 - 1345
  • [25] Development of a Dose Accumulation Pipeline to Support MR-Guided Adaptive Radiation Therapy
    Rusu, D.
    Morris, E.
    Glide-Hurst, C.
    MEDICAL PHYSICS, 2020, 47 (06) : E602 - E602
  • [26] Dose-guided particle therapy with PET imaging
    Parodi, K.
    Paganetti, H.
    Shih, H.
    Michaud, S.
    Bortfeld, T.
    MEDICAL PHYSICS, 2006, 33 (06) : 2203 - 2204
  • [27] FOCUSING OF RELATIVISTIC PARTICLE BEAMS IN MAGNETIC WIGGLER FIELD
    SMOLYAKOV, NV
    ZHURNAL TEKHNICHESKOI FIZIKI, 1992, 62 (03): : 137 - 145
  • [28] THE CALCULATION OF FORCE ON A WEAKLY MAGNETIC PARTICLE IN A MAGNETIC-FIELD
    BAGSTER, DF
    INTERNATIONAL JOURNAL OF MINERAL PROCESSING, 1987, 20 (1-2) : 1 - 15
  • [29] Feasibility of MRI guided proton therapy: magnetic field dose effects
    Raaymakers, B. W.
    Raaijmakers, A. J. E.
    Lagendijk, J. J. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (20): : 5615 - 5622
  • [30] Feasibility of MRI Guided Proton Therapy: Magnetic Field Dose Effects
    Raaymakers, B. W.
    Raaijmakers, A. J. E.
    Lagendijk, J. J. W.
    MEDICAL PHYSICS, 2008, 35 (06)