Accuracy of tibial component positioning in the robotic arm assisted versus conventional unicompartmental knee arthroplasty

被引:8
|
作者
Thilak, Jai [1 ]
Thadi, Mohan [1 ]
Mane, Prajwal P. [1 ]
Sharma, Anubhav [1 ]
Mohan, Vipin [1 ]
Babu, Balu C. [1 ]
机构
[1] Amrita Inst Med Sci, Kochi 682041, Kerala, India
关键词
Unicompartmental; Partial knee arthroplasty; Robotic-assisted surgery; Osteoarthritis; Haptic guidance; COMPUTER NAVIGATION; EARLY FAILURE; REPLACEMENT; ALIGNMENT; IMPLANTATION; REVISION; PATIENT;
D O I
10.1016/j.jor.2020.08.022
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Our study aims to determine the planned accuracy of the tibial component placement in robotic arm assisted unicompartmental knee arthroplasty (UKA) versus the conventional jig based UKA of the initial cases done in India for the first time with this particular robotic system. Materials & Methods: Study group 1 consisted of patients who underwent robotic arm (MAKO, Stryker, USA) assisted UKA. Group 2 consists of patients who underwent a standard conventional jig based (Oxford knee, Biomet, UK). Post-operative radiographs were taken to determine the Tibial Implant position and orientation which were compared to their preoperative plan respectively by two independent observers. The mean error value was obtained for both study groups respectively and compared to determine the accuracy of the postoperative tibial implant placement. Results: In the Robotic arm assisted UKA, the deviation of post-operative varus angle from preoperative planned angle was about 0.43 degrees and post-operative Tibial slope alignment differed from preoperative plan was 0.41 degrees. In the Conventional UKA group post-operative varus angle differed from preoperative planned angle by about 2.12 degrees and post-operative Tibial slope alignment deviation from preoperative plan was 2.47 degrees. Conclusions: Robotic arm assisted system was more accurate compared to the conventional jig-based technique in achieving the planned orientation and alignment of the tibial implant in the initial learning phase of this particular Robotic System used for the first time in India. Mesh terms: partial knee replacement, robotic assisted surgery.
引用
收藏
页码:367 / 371
页数:5
相关论文
共 50 条
  • [31] ROBOTIC-ARM-ASSISTED LATERAL UNICOMPARTMENTAL KNEE ARTHROPLASTY WITH A FIXED-BEARING IMPLANT
    Premkumar, Ajay
    Bayoumi, Tarik
    Pearle, Andrew D.
    JBJS ESSENTIAL SURGICAL TECHNIQUES, 2023, 13 (02):
  • [32] Clinical Outcome Scores Post Medial Unicompartmental Knee Arthroplasty: A Comparison of the MAKO Robotic Arm versus the Oxford Conventional Approach
    Tan, C. M. P.
    Shih, S. S. W.
    Ravichandran, V
    Quah, E. S. H.
    Kunnasegaran, R.
    MALAYSIAN ORTHOPAEDIC JOURNAL, 2025, 19 (01) : 3 - 10
  • [33] Minimally invasive versus conventional unicompartmental knee arthroplasty
    Heyse, Thomas J.
    Efe, Turgay
    Rumpf, Stefan
    Schofer, Markus D.
    Fuchs-Winkelmann, Susanne
    Schmitt, Jan
    Hauk, Carsten
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (09) : 1287 - 1290
  • [34] Does component placement affect short-term clinical outcome in robotic-arm assisted unicompartmental knee arthroplasty?
    Zambianchi, F.
    Franceschi, G.
    Rivi, E.
    Banchelli, F.
    Marcovigi, A.
    Nardacchione, R.
    Ensini, A.
    Catani, F.
    BONE & JOINT JOURNAL, 2019, 101B (04): : 435 - 442
  • [35] Computer assisted versus conventional cemented total knee prostheses alignment accuracy and micromotion of the tibial component
    van Strien, Thisbe
    van der Linden-van der Zwaag, Enrike
    Kaptein, Bart
    van Erkel, Arjan
    Valstar, Edward
    Nelissen, Rob
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (05) : 1255 - 1261
  • [36] Excellent 10-Year survivorship of robotic-arm-assisted unicompartmental knee arthroplasty
    Roche, Martin
    Vakharia, Rushabh M.
    Law, Tsun
    Rodriguez, Hugo C.
    Mekkawy, Kevin L.
    Sodhi, Nipun
    Dubin, Jeremy A.
    Hameed, Daniel
    Mont, Michael A.
    JOURNAL OF ORTHOPAEDICS, 2024, 55 : 32 - 37
  • [37] Short term outcomes following robotic arm-assisted lateral unicompartmental knee arthroplasty
    Wignadasan, Warran
    Chang, Justin
    Fontalis, Andreas
    Plastow, Ricci
    Haddad, Fares S.
    FRONTIERS IN SURGERY, 2023, 10
  • [38] Computer assisted versus conventional cemented total knee prostheses alignment accuracy and micromotion of the tibial component
    Thisbe van Strien
    Enrike van der Linden-van der Zwaag
    Bart Kaptein
    Arjan van Erkel
    Edward Valstar
    Rob Nelissen
    International Orthopaedics, 2009, 33
  • [39] Robotic arm-assisted conversion of unicompartmental knee arthroplasty to total knee arthroplasty (vol 106, pg 680, 2024)
    Mancino, F.
    Fontalis, A.
    Grandhi, T. S. P.
    Magan, A.
    Plastow, R.
    Kayani, B.
    Haddad, F. S.
    BONE & JOINT JOURNAL, 2024, 106B (09): : 1031 - 1031
  • [40] Are there functional biomechanical differences in robotic arm-assisted bi-unicompartmental knee arthroplasty compared with conventional total knee arthroplasty? A prospective, randomized controlled trial
    Banger, M. S.
    Doonan, J.
    Jones, B. G.
    MacLean, A. D.
    Rowe, P. J.
    Blyth, M. J. G.
    BONE & JOINT JOURNAL, 2022, 104B (04): : 433 - 443