Accuracy of tibial component positioning in the robotic arm assisted versus conventional unicompartmental knee arthroplasty

被引:8
|
作者
Thilak, Jai [1 ]
Thadi, Mohan [1 ]
Mane, Prajwal P. [1 ]
Sharma, Anubhav [1 ]
Mohan, Vipin [1 ]
Babu, Balu C. [1 ]
机构
[1] Amrita Inst Med Sci, Kochi 682041, Kerala, India
关键词
Unicompartmental; Partial knee arthroplasty; Robotic-assisted surgery; Osteoarthritis; Haptic guidance; COMPUTER NAVIGATION; EARLY FAILURE; REPLACEMENT; ALIGNMENT; IMPLANTATION; REVISION; PATIENT;
D O I
10.1016/j.jor.2020.08.022
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Our study aims to determine the planned accuracy of the tibial component placement in robotic arm assisted unicompartmental knee arthroplasty (UKA) versus the conventional jig based UKA of the initial cases done in India for the first time with this particular robotic system. Materials & Methods: Study group 1 consisted of patients who underwent robotic arm (MAKO, Stryker, USA) assisted UKA. Group 2 consists of patients who underwent a standard conventional jig based (Oxford knee, Biomet, UK). Post-operative radiographs were taken to determine the Tibial Implant position and orientation which were compared to their preoperative plan respectively by two independent observers. The mean error value was obtained for both study groups respectively and compared to determine the accuracy of the postoperative tibial implant placement. Results: In the Robotic arm assisted UKA, the deviation of post-operative varus angle from preoperative planned angle was about 0.43 degrees and post-operative Tibial slope alignment differed from preoperative plan was 0.41 degrees. In the Conventional UKA group post-operative varus angle differed from preoperative planned angle by about 2.12 degrees and post-operative Tibial slope alignment deviation from preoperative plan was 2.47 degrees. Conclusions: Robotic arm assisted system was more accurate compared to the conventional jig-based technique in achieving the planned orientation and alignment of the tibial implant in the initial learning phase of this particular Robotic System used for the first time in India. Mesh terms: partial knee replacement, robotic assisted surgery.
引用
收藏
页码:367 / 371
页数:5
相关论文
共 50 条
  • [1] Tibial baseplate positioning in robotic-assisted and conventional unicompartmental knee arthroplasty
    MacCallum K.P.
    Danoff J.R.
    Geller J.A.
    European Journal of Orthopaedic Surgery & Traumatology, 2016, 26 (1) : 93 - 98
  • [2] Improved Accuracy of Component Positioning with Robotic-Assisted Unicompartmental Knee Arthroplasty
    Bell, Stuart W.
    Anthony, Iain
    Jones, Bryn
    MacLean, Angus
    Rowe, Philip
    Blyth, Mark
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2016, 98 (08): : 627 - 635
  • [3] Arthroscopic Robotic Arm-assisted Unicompartmental Knee Arthroplasty
    O'Loughlin, Padhraig F.
    Pearle, Andrew D.
    Fyhrie, David P.
    Long, Fanxin
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2009, 17 (04) : 266 - 267
  • [4] Robotic-arm-assisted conversion of unicompartmental knee arthroplasty to total knee arthroplasty
    Magruder, Matthew L.
    Mcclure, Tanner
    Marchand, Kevin
    Mont, Michael A.
    Marchand, Robert C.
    JOURNAL OF ORTHOPAEDICS, 2024, 52 : 119 - 123
  • [5] Robotic arm- assisted conversion of unicompartmental knee arthroplasty to total knee arthroplasty
    Mancino, F.
    Fontalis, A.
    Grandhi, T. S. P.
    Magan, A.
    Plastow, R.
    Kayani, B.
    Haddad, F. S.
    BONE & JOINT JOURNAL, 2024, 106B (07): : 680 - 687
  • [6] An assessment of early functional rehabilitation and hospital discharge in conventional versus robotic-arm assisted unicompartmental knee arthroplasty
    Kayani, B.
    Konan, S.
    Tahmassebi, J.
    Rowan, F. E.
    Haddad, F. S.
    BONE & JOINT JOURNAL, 2019, 101B (01): : 24 - 33
  • [7] Robotic arm-assisted versus conventional unicompartmental knee arthroplasty EXPLORATORY SECONDARY ANALYSIS OF A RANDOMISED CONTROLLED TRIAL
    Blyth, M. J. G.
    Anthony, I.
    Rowe, P.
    Banger, M. S.
    MacLean, A.
    Jones, B.
    BONE & JOINT RESEARCH, 2017, 6 (11): : 631 - 639
  • [8] Improved Component Placement Accuracy with Robotic-Arm Assisted Total Knee Arthroplasty
    Mahoney, Ormonde
    Kinsey, Tracey
    Sodhi, Nipun
    Mont, Michael A.
    Chen, Antonia F.
    Orozco, Fabio
    Hozack, William
    JOURNAL OF KNEE SURGERY, 2022, 35 (03) : 337 - 344
  • [9] Length of stay and discharge dispositions following robotic arm-assisted total knee arthroplasty and unicompartmental knee arthroplasty versus conventional technique and predictors of delayed discharge
    Fontalis, A.
    Raj, R. D.
    Haddad, I. C.
    Donovan, C.
    Plastow, R.
    Oussedik, S.
    Gabr, A.
    Haddad, F. S.
    BONE & JOINT OPEN, 2023, 4 (10): : 791 - 800
  • [10] High accuracy of component positioning and restoration of lower limb alignment using robotic medial unicompartmental knee arthroplasty
    Emanuele, Diquattro
    Lettner, Jonathan
    Adriani, Marco
    Robert, Prill
    Mikhail, Salzmann
    Roland, Becker