Quality assurance and system stability of a clinical MRI-guided focused ultrasound system: Four-year experience

被引:33
|
作者
McDannold, Nathan [1 ]
Hynynen, Kullervo [1 ]
机构
[1] Harvard Univ, Sch Med, Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
关键词
magnetic resonance; focused ultrasound; water proton chemical shift; temperature mapping;
D O I
10.1118/1.2352853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To retrospectively evaluate the four-year experience of a quality assurance method for a MRI-guided focused ultrasound system that uses temperature maps acquired during heating in an ultrasound/MRI phantom. This quality assurance method was performed before 148 clinical uterine fibroid thermal ablation treatments. The stability of the peak temperature rise, the targeting accuracy, the shape of the heated zone, and the noise level in the imaging was evaluated. The peak temperature rise was mostly stable for the first three years. An increase in heating was observed when the system was replaced after year three. Detection of this increase was taken into account in the subsequent clinical treatments. A small secondary hotspot was detected by the temperature maps and was seen to be resolved after system calibration. The average standard deviation in unheated regions of the phantom in the temperature maps was 0.5 +/- 0.2 degrees C; it was less than 1 degrees C in all but one procedure. The average initial targeting error was 2.8 +/- 1.8 and 2.8 +/- 2.1 mm in two radial directions and 7.7 +/- 2.9 mm along the ultrasound beam direction. The width of the heating profile was consistent over the four years. This simple method to evaluate the performance appeared to be sensitive to small changes in system performance, which was adequately stable over a four-year time period. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4307 / 4313
页数:7
相关论文
共 50 条
  • [31] Four-year clinical evaluation of a universal composite resin system.
    Gordan, V. V.
    Watson, R.
    Shen, C.
    Mjor, I. A.
    JOURNAL OF DENTAL RESEARCH, 2003, 82 : B197 - B197
  • [32] Multi-Channel RF System for MRI-Guided Transurethral Ultrasound Thermal Therapy
    Yak, Nicolas
    Asselin, Matthew
    Chopra, Rajiv
    Bronskill, Michael
    8TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2009, 1113 : 382 - 386
  • [33] Is there still a role for lesioning in functional neurosurgery? Preliminary Italian (and world-first) experience with a transcranial MRI-guided focused ultrasound surgery system operating at 1.5 Tesla
    Iacopino, Domenico G.
    Giugno, Antonella
    Maugeri, Rosario
    Gagliardo, Cesare
    Franzini, Angelo
    Catalano, Carlo
    Midiri, Massimo
    JOURNAL OF NEUROSURGICAL SCIENCES, 2017, 61 (06) : 681 - 683
  • [34] MR guided focused ultrasound: technical acceptance measures for a clinical system
    Gorny, K. R.
    Hangiandreou, N. J.
    Hesley, G. K.
    Gostout, B. S.
    McGee, K. P.
    Felmlee, J. P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (12): : 3155 - 3173
  • [35] Four-year follow-up results of magnetic resonance-guided focused ultrasound thalamotomy for essential tremor
    Park, Yong-Sook
    Jung, Na Young
    Na, Young Cheol
    Chang, Jin Woo
    MOVEMENT DISORDERS, 2019, 34 (05) : 727 - 734
  • [36] Pre-clinical testing of a phased array ultrasound system for MRI-guided noninvasive surgery of the brain - A primate study
    Hynynen, Kullervo
    McDannold, Nathan
    Clement, Greg
    Jolesz, Ferenc A.
    Zadicario, Eyal
    Killiany, Ron
    Moore, Tara
    Rosen, Douglas
    EUROPEAN JOURNAL OF RADIOLOGY, 2006, 59 (02) : 149 - 156
  • [37] Quality Assurance and Clinical Experience of MASEP Rotating Gamma System
    Pokhrel, S.
    Ballo, M.
    Dash, G.
    Gray, J.
    Kulkarni, R.
    Patra, P.
    MEDICAL PHYSICS, 2020, 47 (06) : E835 - E836
  • [38] Implementation of a Musculoskeletal Ultrasound (MSUS) Curriculum in an Academic Rheumatology Fellowship Within an Integrated Health System: A Four-Year Experience
    Bulbin, David
    Oppermann, Brian
    Cote, Jonida
    Pugliese, David
    Denio, Alfred
    ARTHRITIS & RHEUMATOLOGY, 2020, 72
  • [39] MRI-guided focused ultrasound ablation (MRgFUS) of uterine fibroids: Clinical predictors of successful outcome at three years.
    Doyle, J. O.
    Betjes, H.
    Missmer, S. A.
    Fennessy, F. M.
    Tempany, C. M. C.
    Stewart, E. A.
    FERTILITY AND STERILITY, 2007, 88 : S82 - S82
  • [40] Correction of the skull density ratio for transcranial MRI-guided focused ultrasound thalamotomy: clinical significance of predicting therapeutic temperature
    Hori, Hiroki
    Iwamuro, Hirokazu
    Nakano, Masayuki
    Ouchi, Takahiro
    Kawahara, Takashi
    Taira, Takaomi
    Abe, Keiichi
    Iijima, Ken
    Yamaguchi, Toshio
    JOURNAL OF NEUROSURGERY, 2021, 135 (05) : 1436 - 1444