A 256 x 256 2-D Array Transducer with Row-Column Addressing for 3-D Rectilinear Imaging

被引:90
|
作者
Seo, Chi Hyung [1 ]
Yen, Jesse T. [2 ]
机构
[1] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[2] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
关键词
3-DIMENSIONAL ULTRASOUND; BREAST;
D O I
10.1109/TUFFC.2009.1107
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
We present simulation and experimental results from a 5-MHz, 256 x 256 2-D (65536 elements, 38.4 x 38.4 mm) 2-D array transducer with row-column addressing. The main benefits of this design are a reduced number of interconnects, a modified transmit/receive switching scheme with a simple diode circuit, and an ability to perform volumetric imaging of targets near the transducer with transmit beamforming in azimuth and receive beamforming in elevation. The final dimensions of the transducer were 38.4 mm x 38.4 rum x 300 mu m. After a row-column transducer was prototyped, the series resonance impedance was 104 Omega at 5.4 MHz. The measured -6 dB fractional bandwidth was 53% with a center frequency of 5.3 MHz. The SNR at the transmit focus was measured to be 30 dB. At 5 MHz, the average nearest neighbor crosstalk was -25 dB. In this paper, we present 3-D images of both 5 pairs of nylon wires embedded in a clear gelatin phantom and an 8 mm diameter cylindrical anechoic cyst phantom acquired from a 256 x 256 2-D array transducer made from a 1-3 composite. We display the azimuth and elevation B-scans as well as the C-scan for each image. The cross-section of the wires is visible in the azimuth B-scan, and the long axes can be seen in the elevation B-scan and C-scans. The pair of wires with 1-mm axial separation is discernible in the elevational B-scan. When a single wire from the wire target phantom was used, the measured lateral beamwidth was 0.68 mm and 0.70 mm at 30 mm depth in transmit beamforming and receive beamforming, respectively, compared with the simulated beamwidth of 0.55 mm. The cross-section of the cyst is visible in the azimuth B-scan whereas the long axes can be seen as a rectangle in the elevation B-scan and C-scans.
引用
收藏
页码:837 / 847
页数:11
相关论文
共 50 条
  • [31] 4-D Vector Doppler Imaging Using Row-Column Addressed Array
    Sun, Qiandong
    Fu, Yapeng
    Yan, Shaoyuan
    Xu, Kailiang
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2025, 72 (02) : 202 - 214
  • [32] Fast 3-D Velocity Estimation in 4-D Using a 62+62 Row-Column Addressed Array
    Schou, Mikkel
    Jorgensen, Lasse Thurmann
    Beers, Christopher
    Traberg, Marie Sand
    Tomov, Borislav Gueorguiev
    Bo Stuart, Matthias
    Jensen, Jorgen Arendt
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (03) : 608 - 623
  • [33] Experimental 3-D Vector Velocity Estimation with Row-Column Addressed Arrays
    Holbek, Simon
    Stuart, Matthias Bo
    Jensen, Jorgen Arendt
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [34] 3-D Imaging using Row-Column-Addressed 2-D Arrays with a Diverging Lens: Phantom Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [35] Volumetric 3-D Vector Flow Measurements using a 62+62 Row-Column Addressed Array
    Holbek, Simon
    Stuart, Matthias Bo
    Jensen, Jorgen Arendt
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [36] Theoretical assessment of a crossed electrode 2-D array for 3-D imaging
    Morton, CE
    Lockwood, GR
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 968 - 971
  • [37] Volumetric Ultrasound Imaging with Row-Column Addressed 2-D Arrays Using Spatial Matched Filter Beamforming
    Bouzari, Hamed
    Engholm, Mathias
    Christiansen, Thomas Lehrmann
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [38] 2-D Row-Column CMUT Arrays with an Open-Grid Support Structure
    Christiansen, Thomas Lehrmann
    Dahl-Petersen, Christian
    Jensen, Jorgen Arendt
    Thomsen, Erik V.
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 1712 - 1715
  • [39] Curvilinear 3-D Imaging Using Row-Column-Addressed 2-D Arrays With a Diverging Lens: Feasibility Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2017, 64 (06) : 978 - 988
  • [40] Curvilinear 3-D Imaging Using Row-Column-Addressed 2-D Arrays With a Diverging Lens: Phantom Study
    Bouzari, Hamed
    Engholm, Mathias
    Beers, Christopher
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (07) : 1182 - 1192