A Comparison of the Mechanical Measures Used for Assessing Orthodontic Mini-Implant Stability

被引:9
|
作者
Hosein, Yara K. [1 ]
Dixon, S. Jeffrey [2 ]
Rizkalla, Amin S. [3 ]
Tassi, Ali [4 ]
机构
[1] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Div Grad Orthodont, London, ON, Canada
[2] Western Univ, Bone & Joint Inst, Schulich Sch Med & Dent, Dept Physiol & Pharmacol,Div Grad Orthodont, London, ON, Canada
[3] Western Univ, Schulich Sch Med Dent, Dept Chem & Biochem Engn, Div Grad Orthodont,Bone & Joint Inst,Fac Engn, London, ON, Canada
[4] Western Univ, Div Grad Orthodont, Schulich Sch Med & Dent, London, ON, Canada
关键词
torque; pull-out load; implant loosening; Periotest; mechanical retention; PULL-OUT STRENGTH; INSERTION TORQUE; SURFACE-TREATMENT; DAMPING CAPACITY; TAPER SHAPE; MINISCREWS; LENGTH; THREAD; IMPACT; DIAMETER;
D O I
10.1097/ID.0000000000000514
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose:Mechanical loosening remains a common complication associated with mini-implant failure. The purpose of this study was to compare common mechanical measures of mini-implant stability to determine their association and reliability.Materials and Methods:Ninety self-drilling orthodontic mini-implants from 6 manufacturers were inserted into artificial bone blocks. Insertion torques (ITs) and Periotest values (PVs) were measured. Subsequently, mini-implants underwent pull-out testing for measures of pull-out load (POL) and screw displacement (ScrD). Stability measurements were compared using one-way ANOVA, associations among them were assessed using correlation analyses, and reliability was evaluated using coefficients of variation (COVs).Results:Variations in stability of mini-implants were found, specific to the mechanical measure used for assessment (P < 0.05). The strongest correlations were found between IT and PV (r = -0.68) and between IT and POL (r = 0.66). Overall, PV showed the greatest variability (COV: 11%-100%) compared with IT (11%), POL (4%), and ScrD (19%).Conclusions:IT, PV, and POLs only agreed moderately in their assessment of mini-implant stability, and Periotest showed the least reliability in predicting mini-implant stability. As such, independent and interchangeable use of these stability measures should be avoided.
引用
收藏
页码:225 / 231
页数:7
相关论文
共 50 条
  • [31] Bivariate optimization of orthodontic mini-implant thread height and pitch
    Shuning Shen
    Yingying Sun
    Chen Zhang
    Yongjin Yang
    Zhiren Li
    Xingwei Cai
    Yinzhong Duan
    Tao Li
    International Journal of Computer Assisted Radiology and Surgery, 2015, 10 : 109 - 116
  • [32] Primary migration of a mini-implant under a functional orthodontic loading
    Joseph W. Pittman
    Anand Navalgund
    Steve H. Byun
    Hechang Huang
    Albert H. Kim
    Do-Gyoon Kim
    Clinical Oral Investigations, 2014, 18 : 721 - 728
  • [33] A prospective study of factors associated with orthodontic mini-implant survival
    Tsai, Chia-Chun
    Chang, Hong-Po
    Pan, Chin-Yun
    Chou, Szu-Ting
    Tseng, Yu-Chuan
    JOURNAL OF ORAL SCIENCE, 2016, 58 (04) : 515 - 521
  • [34] Systematic review of mini-implant displacement under orthodontic loading
    Manuel Nienkemper
    Jrg Handschel
    Dieter Drescher
    International Journal of Oral Science, 2014, 6 (01) : 1 - 6
  • [35] Systematic review of mini-implant displacement under orthodontic loading
    Manuel Nienkemper
    Jrg Handschel
    Dieter Drescher
    International Journal of Oral Science, 2014, (01) : 1 - 6
  • [36] Bone thickness of the palate for orthodontic mini-implant anchorage in adults
    Kang, Sungmin
    Lee, Shin-Jae
    Ahn, Sug-Joon
    Heo, Min-Suk
    Kim, Tae-Woo
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 131 (04) : S74 - S81
  • [37] The impact of healing time before loading on orthodontic mini-implant stability: A histomorphometric study in minipigs
    Oltramari-Navarro, Paula V. P.
    Navarro, Ricardo L.
    Henriques, Jose Fernando C.
    Cestari, Tania M.
    Francischone, Carlos E.
    Taga, Rumio
    McNamara, James A., Jr.
    ARCHIVES OF ORAL BIOLOGY, 2013, 58 (07) : 806 - 812
  • [38] Influence of bone density, screw size and surgical procedure on orthodontic mini-implant placement - part B: implant stability
    Moehlhenrich, S. C.
    Heussen, N.
    Modabber, A.
    Bock, A.
    Hoelzle, F.
    Wilmes, B.
    Danesh, G.
    Szalma, J.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2021, 50 (04) : 565 - 572
  • [40] Evaluation of orthodontic mini-implant anchorage in premolar extraction therapy in adolescents
    Garfinkle, Judah S.
    Cunningham, Larry L., Jr.
    Beeman, Cynthia S.
    Kluemper, G. Thomas
    Hicks, E. Preston
    Kim, Mi-Ok
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (05) : 642 - 653