Monte Carlo study of photon fields from a flattening filter-free clinical accelerator

被引:90
|
作者
Vassiliev, ON [1 ]
Titt, U [1 ]
Kry, SF [1 ]
Pönisch, F [1 ]
Gillin, MT [1 ]
Mohan, R [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Houston, TX 77030 USA
关键词
D O I
10.1118/1.2174720
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In conventional clinical linear accelerators, the flattening filter scatters and absorbs a large fraction of primary photons. Increasing the beam-on time, which also increases the out-of-field exposure to patients, compensates for the reduction in photon fluence. In recent years, intensity modulated radiation therapy has been introduced, yielding better dose distributions than conventional three-dimensional conformal therapy. The drawback of this method is the further increase in beam-on time. An accelerator with the flattening filter removed, which would increase photon fluence greatly, could deliver considerably higher dose rates. The objective of the present study is to investigate the dosimetric properties of 6 and 18 MV photon beams from an accelerator without a flattening filter. The dosimetric data were generated using the Monte Carlo programs BEAMnrc and DOSXYZnrc. The accelerator model was based on the Varian Clinac 2100 design. We compared depth doses, dose rates, lateral profiles, doses outside collimation, total and collimator scatter factors for an accelerator with and without a flatteneing filter. The study showed that removing the filter increased the dose rate on the central axis by a factor of 2.31 (6 MV) and 5.45 (18 MV) at a given target current. Because the flattening filter is a major source of head scatter photons, its removal from the beam line could reduce the out-of-field dose. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:820 / 827
页数:8
相关论文
共 50 条
  • [1] A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements
    Dalaryd, Marten
    Kragl, Gabriele
    Ceberg, Crister
    Georg, Dietmar
    McClean, Brendan
    af Wetterstedt, Sacha
    Wieslander, Elinore
    Knoos, Tommy
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23): : 7333 - 7344
  • [2] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Mesbahi, Asghar
    Nejad, Farshad Seyed
    RADIATION MEDICINE, 2008, 26 (06): : 331 - 336
  • [3] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Asghar Mesbahi
    Farshad Seyed Nejad
    Radiation Medicine, 2008, 26 : 331 - 336
  • [4] Monte Carlo study of backscatter in a flattening filter free clinical accelerator
    Titt, U.
    Vassiliev, O. N.
    Ponisch, F.
    Kry, S. F.
    Mohan, R.
    MEDICAL PHYSICS, 2006, 33 (09) : 3270 - 3273
  • [5] Dosimetric properties of a flattening filter-free 6-MV photon beam: A Monte Carlo study
    Mesbahi A.
    Mehnati P.
    Keshtkar A.
    Farajollahi A.
    Radiation Medicine, 2007, 25 (7): : 315 - 324
  • [6] DNA strand breaks based on Monte Carlo simulation in and around the Lipiodol with flattening filter and flattening filter-free photon beams
    Kawahara, Daisuke
    Saito, Akito
    Nakano, Hisashi
    Nagata, Yasushi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (03) : 392 - 400
  • [7] Monte Carlo study of a flattening filter-free 6 MV photon beam using the BEAMnrc code.
    Kajaria, Ankit
    Sharma, Neeraj
    Sharma, Shiru
    Pradhan, Satyajit
    Mandal, Abhijit
    Aggarwal M, Lalit
    BIOMEDICAL RESEARCH-INDIA, 2017, 28 (04): : 1566 - 1573
  • [8] Dose evaluation of Grid Therapy using a 6 MV flattening filter-free (FFF) photon beam: A Monte Carlo study
    Martinez-Rovira, Immaculada
    Puxeu-Vaque, Josep
    Prezado, Yolanda
    MEDICAL PHYSICS, 2017, 44 (10) : 5378 - 5383
  • [9] Monte Carlo evaluation of the potential benefits of flattening filter free beams from the Oncor® clinical linear accelerator
    Asadi, Amin
    Razavi-Ratki, Seid Kazem
    Jabbari, Keyvan
    Najafzadeh, Milad
    Nickfarjam, Abolfazl
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (02) : 281 - 302
  • [10] A flattening filter free photon treatment concept evaluation with Monte Carlo
    Titt, U.
    Vassiliev, O. N.
    Poenisch, F.
    Dong, L.
    Liu, H.
    Mohan, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 1595 - 1602